Trial Outcomes & Findings for Surgical Treatment of Aortic Stenosis With a Next Generation, Rapid Deployment Surgical Aortic Valve (NCT NCT01700439)

NCT ID: NCT01700439

Last Updated: 2023-05-19

Results Overview

Late patient years are calculated from 31 days post-implant to the date of the last follow-up visits (or contact) or adverse events. Late Patient year calculation:\[(Number of late adverse events/sum of late patient years) x 100\]

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

934 participants

Primary outcome timeframe

Events occurring ≥ 31 days and up through 8 years post-implant

Results posted on

2023-05-19

Participant Flow

Participant milestones

Participant milestones
Measure
Edwards TRANSFORM Trial Cohort
Subjects who were enrolled in the TRANSFORM trial and were eligible to receive an INTUITY study surgical aortic heart valve.
Overall Study
STARTED
934
Overall Study
Study Valve Group
885
Overall Study
Discharge
925
Overall Study
3 Month
862
Overall Study
1 Year
842
Overall Study
2 Year
795
Overall Study
3 Year
720
Overall Study
4 Year
667
Overall Study
5 Year
577
Overall Study
6 Year
357
Overall Study
7 Year
164
Overall Study
8 Year
40
Overall Study
COMPLETED
518
Overall Study
NOT COMPLETED
416

Reasons for withdrawal

Reasons for withdrawal
Measure
Edwards TRANSFORM Trial Cohort
Subjects who were enrolled in the TRANSFORM trial and were eligible to receive an INTUITY study surgical aortic heart valve.
Overall Study
Lost to Follow-up
47
Overall Study
Withdrawal by Subject
183
Overall Study
Death
186

Baseline Characteristics

Surgical Treatment of Aortic Stenosis With a Next Generation, Rapid Deployment Surgical Aortic Valve

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Edwards TRANSFORM Trial Cohort
n=934 Participants
Subjects who were enrolled in the TRANSFORM trial and were eligible to receive an Edwards INTUITY surgical aortic heart valve.
Age, Continuous
73.4 Years
STANDARD_DEVIATION 8.3 • n=5 Participants
Sex: Female, Male
Female
336 Participants
n=5 Participants
Sex: Female, Male
Male
598 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
4 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
865 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
65 Participants
n=5 Participants
Region of Enrollment
United States
934 participants
n=5 Participants

PRIMARY outcome

Timeframe: Events occurring ≥ 31 days and up through 8 years post-implant

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve.

Late patient years are calculated from 31 days post-implant to the date of the last follow-up visits (or contact) or adverse events. Late Patient year calculation:\[(Number of late adverse events/sum of late patient years) x 100\]

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
All cause mortality
3.9 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve related mortality
0.9 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Study valve reoperation
0.6 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Study valve explant
0.3 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Thromboembolism
1.8 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve thrombosis
0.0 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Endocarditis
0.1 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
All bleeding
4.7 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Major bleeding
2.4 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
All paravalvular leak
0.5 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Major paravalvular leak (PVL)
0.3 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Hemolysis
0.2 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Non-structural valve dysfunction other than PVL
0.1 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve migration/embolization
0.0 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve malposition
0.0 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve instability
0.0 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve dislodgement
0.0 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Valve stenosis
0.1 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Structural valve deterioration
0.4 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Conduction defects
0.4 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Conduction defects with permanent pacemaker
0.1 Percentage of events/late patient years
Number of Late Adverse Events Divided by Late Patient Years (Expressed as a Percentage) for the INTUTY Valve Implanted Cohort
Conduction defects without permanent pacemaker
0.3 Percentage of events/late patient years

SECONDARY outcome

Timeframe: Day of procedure

Population: Patients who were eligible to receive an Edwards INTUITY surgical aortic heart valve.

Device technical success is defined as the successful delivery and deployment of the aortic trial heart valve with maximum of two attempts and subject leaving the operating room (OR) with valve in place.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=934 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Percentage of Subjects With Edwards INTUITY Surgical Aortic Heart Valve Device Technical Success
94.8 Percentage of subjects

SECONDARY outcome

Timeframe: Day of procedure through discharge or 10 days post index procedure, whichever comes first.

Population: Patients who were eligible to receive an Edwards INTUITY surgical aortic heart valve.

Procedural success is defined as device technical success followed by the absence of adverse events resulting in device reoperation implant of permanent pacemaker (with baseline sinus rhythm and no other pre-existing conduction issues), or valve-related death within discharge or 10 days post index procedure, whichever comes first.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=934 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Percentage of Subjects With Edwards INTUITY Surgical Heart Valve Procedural Success
94.4 Percentage of subjects

SECONDARY outcome

Timeframe: Day of procedure

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve.

Surgical and hospitalization factors - Cardiopulmonary bypass time

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Average Amount of Time Subject Spent on Cardiopulmonary Bypass
86.6 Minutes
Standard Deviation 37.4

SECONDARY outcome

Timeframe: Day of procedure

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve. Data is not available for one subject.

Surgical and hospitalization factors - Cardiopulmonary cross clamp time

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=884 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Average Subject Time on Cardiopulmonary Cross Clamp
65.4 Minutes
Standard Deviation 29.7

SECONDARY outcome

Timeframe: Day of procedure through discharge from the hospital

Population: The group of subjects that received the Edwards INTUITY surgical aortic heart valve. Data is not available for 4 subjects.

Length of time surgical subjects were in the intensive care unit (ICU) after their heart valve replacement procedure.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=881 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Average Number of Days Subjects Were in the Intensive Care Unit (ICU)
3.1 Days
Standard Deviation 3.7

SECONDARY outcome

Timeframe: Baseline, 3-months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8 Years

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

The New York Heart Association (NYHA) functional classification system relates symptoms to everyday activities and the patient's quality of life. Class I. Patients with cardiac disease but without resulting limitation of physical activity. Class II. Patients with cardiac disease resulting in slight limitation of physical activity. They are comfortable at rest. Class III. Patients with cardiac disease resulting in marked limitation of physical activity. They are comfortable at rest. Class IV. Patients with cardiac disease resulting in inability to carry on any physical activity without discomfort. Symptoms of heart failure or the anginal syndrome may be present even at rest.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=885 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=885 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=885 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's New York Heart Association (NYHA) Classification Over Time
7 Years
109 Participants
31 Participants
4 Participants
0 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
8 Years
28 Participants
5 Participants
1 Participants
0 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
6 Years
251 Participants
63 Participants
14 Participants
0 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
5 Years
420 Participants
109 Participants
14 Participants
0 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
Baseline
135 Participants
468 Participants
264 Participants
15 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
3 months
671 Participants
148 Participants
20 Participants
0 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
1 Year
648 Participants
145 Participants
18 Participants
3 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
2 Years
606 Participants
133 Participants
22 Participants
3 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
3 Years
551 Participants
122 Participants
19 Participants
1 Participants
Subject's New York Heart Association (NYHA) Classification Over Time
4 Years
487 Participants
123 Participants
21 Participants
3 Participants

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Mean gradient is the average flow of blood through the aortic valve measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher value is considered worse, and a lower value is considered better but the value is dependent on the size and type of valve.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Mean Gradient Measurements Over Time
Discharge
17.7 mmHg
Standard Deviation 6.2
13.0 mmHg
Standard Deviation 4.1
12.3 mmHg
Standard Deviation 4.0
11.2 mmHg
Standard Deviation 3.7
9.7 mmHg
Standard Deviation 3.2
Subject's Average Mean Gradient Measurements Over Time
3 months
14.4 mmHg
Standard Deviation 4.4
11.1 mmHg
Standard Deviation 3.3
10.3 mmHg
Standard Deviation 3.5
9.7 mmHg
Standard Deviation 3.2
8.3 mmHg
Standard Deviation 2.5
Subject's Average Mean Gradient Measurements Over Time
8 Years
11.2 mmHg
Standard Deviation 0.0
13.2 mmHg
Standard Deviation 5.0
12.3 mmHg
Standard Deviation 8.3
12.9 mmHg
Standard Deviation 8.2
9.4 mmHg
Standard Deviation 2.2
Subject's Average Mean Gradient Measurements Over Time
1 Year
13.9 mmHg
Standard Deviation 4.0
11.3 mmHg
Standard Deviation 3.6
10.0 mmHg
Standard Deviation 3.2
9.3 mmHg
Standard Deviation 3.0
8.1 mmHg
Standard Deviation 3.3
Subject's Average Mean Gradient Measurements Over Time
2 Years
13.1 mmHg
Standard Deviation 4.5
11.1 mmHg
Standard Deviation 3.6
10.0 mmHg
Standard Deviation 3.7
9.3 mmHg
Standard Deviation 2.9
8.1 mmHg
Standard Deviation 3.6
Subject's Average Mean Gradient Measurements Over Time
3 Years
13.7 mmHg
Standard Deviation 4.6
11.0 mmHg
Standard Deviation 4.3
10.0 mmHg
Standard Deviation 3.4
9.5 mmHg
Standard Deviation 3.5
7.9 mmHg
Standard Deviation 2.9
Subject's Average Mean Gradient Measurements Over Time
4 Years
12.8 mmHg
Standard Deviation 3.8
10.8 mmHg
Standard Deviation 3.7
10.3 mmHg
Standard Deviation 6.7
9.5 mmHg
Standard Deviation 4.0
8.0 mmHg
Standard Deviation 3.7
Subject's Average Mean Gradient Measurements Over Time
5 Years
12.4 mmHg
Standard Deviation 3.5
10.7 mmHg
Standard Deviation 3.9
10.0 mmHg
Standard Deviation 5.6
9.8 mmHg
Standard Deviation 6.5
7.9 mmHg
Standard Deviation 4.4
Subject's Average Mean Gradient Measurements Over Time
6 Years
13.1 mmHg
Standard Deviation 3.9
10.5 mmHg
Standard Deviation 3.6
11.1 mmHg
Standard Deviation 6.4
9.3 mmHg
Standard Deviation 4.0
9.2 mmHg
Standard Deviation 6.5
Subject's Average Mean Gradient Measurements Over Time
7 Years
12.1 mmHg
Standard Deviation 4.0
10.8 mmHg
Standard Deviation 3.1
12.8 mmHg
Standard Deviation 7.4
9.8 mmHg
Standard Deviation 4.0
7.7 mmHg
Standard Deviation 1.8

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8-Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Peak gradient is the maximum value measured of flow of blood through the aortic valve as measured in millimeters of mercury. Gradients are evaluated by echocardiography over time. In general, a higher valve is considered worse, and a lower value is considered better, but the value is dependent on the size and type of valve.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Peak Gradient Measurements Over Time
5 Years
22.7 mmHg
Standard Deviation 6.3
19.5 mmHg
Standard Deviation 7.0
18.4 mmHg
Standard Deviation 10.0
17.9 mmHg
Standard Deviation 10.7
14.5 mmHg
Standard Deviation 7.6
Subject's Average Peak Gradient Measurements Over Time
6 Years
23.6 mmHg
Standard Deviation 6.1
19.1 mmHg
Standard Deviation 6.7
19.8 mmHg
Standard Deviation 9.4
17.1 mmHg
Standard Deviation 7.3
16.9 mmHg
Standard Deviation 10.9
Subject's Average Peak Gradient Measurements Over Time
Discharge
32.8 mmHg
Standard Deviation 11.9
24.1 mmHg
Standard Deviation 7.9
22.9 mmHg
Standard Deviation 7.8
20.8 mmHg
Standard Deviation 6.8
17.9 mmHg
Standard Deviation 5.8
Subject's Average Peak Gradient Measurements Over Time
3 months
26.0 mmHg
Standard Deviation 8.0
19.9 mmHg
Standard Deviation 6.4
19.0 mmHg
Standard Deviation 6.5
17.6 mmHg
Standard Deviation 5.8
15.0 mmHg
Standard Deviation 4.4
Subject's Average Peak Gradient Measurements Over Time
1 Year
25.0 mmHg
Standard Deviation 7.7
20.5 mmHg
Standard Deviation 6.5
18.2 mmHg
Standard Deviation 6.0
17.0 mmHg
Standard Deviation 5.7
14.6 mmHg
Standard Deviation 5.8
Subject's Average Peak Gradient Measurements Over Time
2 Years
23.4 mmHg
Standard Deviation 7.9
20.2 mmHg
Standard Deviation 6.6
18.4 mmHg
Standard Deviation 6.8
17.0 mmHg
Standard Deviation 5.3
14.6 mmHg
Standard Deviation 6.2
Subject's Average Peak Gradient Measurements Over Time
3 Years
24.7 mmHg
Standard Deviation 8.2
20.0 mmHg
Standard Deviation 7.8
18.5 mmHg
Standard Deviation 6.3
17.4 mmHg
Standard Deviation 6.3
14.5 mmHg
Standard Deviation 5.6
Subject's Average Peak Gradient Measurements Over Time
7 Years
22.1 mmHg
Standard Deviation 7.1
19.6 mmHg
Standard Deviation 5.9
23.5 mmHg
Standard Deviation 13.3
18.5 mmHg
Standard Deviation 7.7
14.6 mmHg
Standard Deviation 3.7
Subject's Average Peak Gradient Measurements Over Time
4 Years
23.0 mmHg
Standard Deviation 6.6
19.8 mmHg
Standard Deviation 6.8
18.9 mmHg
Standard Deviation 10.4
17.6 mmHg
Standard Deviation 7.2
14.6 mmHg
Standard Deviation 6.6
Subject's Average Peak Gradient Measurements Over Time
8 Years
21.3 mmHg
Standard Deviation 0.0
24.4 mmHg
Standard Deviation 8.1
21.0 mmHg
Standard Deviation 12.6
22.7 mmHg
Standard Deviation 14.7
17.2 mmHg
Standard Deviation 3.6

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Effective orifice area represents the cross-sectional area of the blood flow downstream of the aortic valve. Effective orifice area is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size and type of valve.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Effective Orifice Area (EOA) Measurement Over Time
4 Years
1.2 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.1 Centimeters squared
Standard Deviation 0.3
Subject's Effective Orifice Area (EOA) Measurement Over Time
5 Years
1.2 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.2
1.7 Centimeters squared
Standard Deviation 0.2
1.8 Centimeters squared
Standard Deviation 0.2
2.1 Centimeters squared
Standard Deviation 0.3
Subject's Effective Orifice Area (EOA) Measurement Over Time
6 Years
1.2 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.6 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.0 Centimeters squared
Standard Deviation 0.3
Subject's Effective Orifice Area (EOA) Measurement Over Time
2 Years
1.1 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.2 Centimeters squared
Standard Deviation 0.2
Subject's Effective Orifice Area (EOA) Measurement Over Time
3 Years
1.1 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.1 Centimeters squared
Standard Deviation 0.2
Subject's Effective Orifice Area (EOA) Measurement Over Time
Discharge
1.1 Centimeters squared
Standard Deviation 0.1
1.3 Centimeters squared
Standard Deviation 0.2
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.2 Centimeters squared
Standard Deviation 0.2
Subject's Effective Orifice Area (EOA) Measurement Over Time
3 months
1.1 Centimeters squared
Standard Deviation 0.1
1.3 Centimeters squared
Standard Deviation 0.1
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.2 Centimeters squared
Standard Deviation 0.2
Subject's Effective Orifice Area (EOA) Measurement Over Time
1 Year
1.1 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.7 Centimeters squared
Standard Deviation 0.2
1.9 Centimeters squared
Standard Deviation 0.2
2.2 Centimeters squared
Standard Deviation 0.2
Subject's Effective Orifice Area (EOA) Measurement Over Time
7 Years
1.2 Centimeters squared
Standard Deviation 0.1
1.4 Centimeters squared
Standard Deviation 0.1
1.6 Centimeters squared
Standard Deviation 0.3
1.8 Centimeters squared
Standard Deviation 0.3
2.1 Centimeters squared
Standard Deviation 0.1
Subject's Effective Orifice Area (EOA) Measurement Over Time
8 Years
1.1 Centimeters squared
Standard Deviation 0.0
1.4 Centimeters squared
Standard Deviation 0.1
1.6 Centimeters squared
Standard Deviation 0.3
1.7 Centimeters squared
Standard Deviation 0.3
2.1 Centimeters squared
Standard Deviation 0.1

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Effective orifice area index represents the minimal cross-sectional area of the blood flow downstream of the mitral valve divided by the person's body surface area. Effective orifice area index is evaluated by echocardiography over time. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
Discharge
0.6 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
3 months
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
1 Year
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
2 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
3 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
5 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
6 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.2
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
4 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.1
0.9 centimeters squared/meters squared
Standard Deviation 0.1
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
7 Years
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.2
0.9 centimeters squared/meters squared
Standard Deviation 0.2
1.0 centimeters squared/meters squared
Standard Deviation 0.1
Subject's Effective Orifice Area Index (EOAI) Measurement Over Time
8 Years
0.6 centimeters squared/meters squared
Standard Deviation 0.0
0.7 centimeters squared/meters squared
Standard Deviation 0.1
0.8 centimeters squared/meters squared
Standard Deviation 0.2
0.8 centimeters squared/meters squared
Standard Deviation 0.2
1.0 centimeters squared/meters squared
Standard Deviation 0.1

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Performance index is defined as the subject's effective orifice area (the cross-sectional area of the blood flow downstream of the aortic valve) divided by the subject's native orifice area. Performance Index is evaluated by echocardiography over time and are dependent on the size and type of the valve. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient and the size and type of valve.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Performance Index Measurement Over Time
Discharge
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
3 months
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
3 Years
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
1 Year
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
2 Years
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
4 Years
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
5 Years
1.0 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
Subject's Performance Index Measurement Over Time
7 Years
1.0 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.0
Subject's Performance Index Measurement Over Time
8 Years
0.9 Centimeters squared/meters squared
Standard Deviation 0.0
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.2
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.0
Subject's Performance Index Measurement Over Time
6 Years
1.0 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1
0.9 Centimeters squared/meters squared
Standard Deviation 0.1
0.8 Centimeters squared/meters squared
Standard Deviation 0.1

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Cardiac output is the amount of blood the heart pumps through the circulatory system in a minute. Cardiac output is evaluated by echocardiography over time. Normal cardiac output in healthy volunteers over 60 years of age was reported between 3.1 and 6.4 liters per minute. In general, a higher value is considered better, and a lower value is considered worse.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Cardiac Output Measurement at Over Time
8 Years
2.9 Liters/minutes
Standard Deviation 0.0
4.9 Liters/minutes
Standard Deviation 1.6
5.4 Liters/minutes
Standard Deviation 1.3
4.9 Liters/minutes
Standard Deviation 1.3
5.5 Liters/minutes
Standard Deviation 1.3
Subject's Cardiac Output Measurement at Over Time
Discharge
4.3 Liters/minutes
Standard Deviation 0.9
4.6 Liters/minutes
Standard Deviation 0.9
5.4 Liters/minutes
Standard Deviation 1.1
5.6 Liters/minutes
Standard Deviation 1.1
5.9 Liters/minutes
Standard Deviation 1.3
Subject's Cardiac Output Measurement at Over Time
3 months
4.3 Liters/minutes
Standard Deviation 0.7
4.4 Liters/minutes
Standard Deviation 0.8
5.0 Liters/minutes
Standard Deviation 0.9
5.2 Liters/minutes
Standard Deviation 1.0
5.4 Liters/minutes
Standard Deviation 1.1
Subject's Cardiac Output Measurement at Over Time
1 Year
4.3 Liters/minutes
Standard Deviation 0.9
4.4 Liters/minutes
Standard Deviation 1.0
4.9 Liters/minutes
Standard Deviation 0.9
5.1 Liters/minutes
Standard Deviation 1.0
5.5 Liters/minutes
Standard Deviation 1.2
Subject's Cardiac Output Measurement at Over Time
2 Years
4.1 Liters/minutes
Standard Deviation 0.8
4.3 Liters/minutes
Standard Deviation 0.8
4.9 Liters/minutes
Standard Deviation 0.9
5.0 Liters/minutes
Standard Deviation 1.0
5.3 Liters/minutes
Standard Deviation 1.0
Subject's Cardiac Output Measurement at Over Time
3 Years
4.0 Liters/minutes
Standard Deviation 0.7
4.3 Liters/minutes
Standard Deviation 0.9
4.9 Liters/minutes
Standard Deviation 0.9
5.1 Liters/minutes
Standard Deviation 1.0
5.2 Liters/minutes
Standard Deviation 1.1
Subject's Cardiac Output Measurement at Over Time
4 Years
4.1 Liters/minutes
Standard Deviation 0.8
4.1 Liters/minutes
Standard Deviation 0.8
4.9 Liters/minutes
Standard Deviation 1.0
5.1 Liters/minutes
Standard Deviation 1.0
5.2 Liters/minutes
Standard Deviation 1.1
Subject's Cardiac Output Measurement at Over Time
5 Years
4.2 Liters/minutes
Standard Deviation 0.7
4.3 Liters/minutes
Standard Deviation 0.9
4.8 Liters/minutes
Standard Deviation 1.0
5.0 Liters/minutes
Standard Deviation 1.1
5.2 Liters/minutes
Standard Deviation 0.9
Subject's Cardiac Output Measurement at Over Time
6 Years
4.3 Liters/minutes
Standard Deviation 0.7
4.2 Liters/minutes
Standard Deviation 0.9
4.9 Liters/minutes
Standard Deviation 1.1
5.0 Liters/minutes
Standard Deviation 1.1
5.3 Liters/minutes
Standard Deviation 1.2
Subject's Cardiac Output Measurement at Over Time
7 Years
3.9 Liters/minutes
Standard Deviation 0.6
4.2 Liters/minutes
Standard Deviation 0.8
5.0 Liters/minutes
Standard Deviation 1.0
5.1 Liters/minutes
Standard Deviation 1.1
5.0 Liters/minutes
Standard Deviation 0.9

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Hemodynamic performance - Cardiac index (CI) evaluated by echocardiography. Cardiac index is an assessment of the cardiac output value based on the patient's size. To obtain the cardiac index, divide the cardiac output by the person's body surface area (BSA). Cardiac Index is a useful marker of how well the heart is functioning as a pump by directly correlating the volume of blood pumped by the heart with an individual's body surface area. Normal cardiac output index in healthy volunteers over 60 years of age was reported between 1.1 and 3.2 L/min/m\^2. In general, a higher value is considered better, and a lower value is considered worse, but the value is dependent on the size of the patient.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Cardiac Index Measurement Over Time
Discharge
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.7 Liters/minutes/meters squared
Standard Deviation 0.5
2.6 Liters/minutes/meters squared
Standard Deviation 0.6
2.7 Liters/minutes/meters squared
Standard Deviation 0.6
Subject's Cardiac Index Measurement Over Time
3 months
2.6 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.4
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
Subject's Cardiac Index Measurement Over Time
1 Year
2.5 Liters/minutes/meters squared
Standard Deviation 0.6
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.4
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.5 Liters/minutes/meters squared
Standard Deviation 0.6
Subject's Cardiac Index Measurement Over Time
2 Years
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.4
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.4
Subject's Cardiac Index Measurement Over Time
3 Years
2.3 Liters/minutes/meters squared
Standard Deviation 0.4
2.3 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.4
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
Subject's Cardiac Index Measurement Over Time
4 Years
2.3 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.4
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
Subject's Cardiac Index Measurement Over Time
5 Years
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.4
Subject's Cardiac Index Measurement Over Time
6 Years
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.4
2.4 Liters/minutes/meters squared
Standard Deviation 0.6
2.4 Liters/minutes/meters squared
Standard Deviation 0.6
2.4 Liters/minutes/meters squared
Standard Deviation 0.6
Subject's Cardiac Index Measurement Over Time
7 Years
2.3 Liters/minutes/meters squared
Standard Deviation 0.4
2.2 Liters/minutes/meters squared
Standard Deviation 0.4
2.5 Liters/minutes/meters squared
Standard Deviation 0.5
2.4 Liters/minutes/meters squared
Standard Deviation 0.5
2.3 Liters/minutes/meters squared
Standard Deviation 0.3
Subject's Cardiac Index Measurement Over Time
8 Years
1.5 Liters/minutes/meters squared
Standard Deviation 0.0
2.6 Liters/minutes/meters squared
Standard Deviation 0.9
2.6 Liters/minutes/meters squared
Standard Deviation 0.6
2.4 Liters/minutes/meters squared
Standard Deviation 0.6
2.8 Liters/minutes/meters squared
Standard Deviation 0.2

SECONDARY outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Valvular regurgitation occurs when the valve in the heart does not close tightly allowing some of the blood that was pumped out of the heart to leak back into it. Valvular regurgitation is evaluated by echocardiography over time. It is assessed on a scale from 0 to 4, where 0 represents no regurgitation and 4 represents severe regurgitation. Higher numbers on the scale show a worsening outcome.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=49 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
n=179 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
n=291 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
n=261 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
n=105 Participants
Subjects who received an INTUITY study surgical aortic heart valve.
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 months · Mild (+2)
2 Participants
12 Participants
16 Participants
18 Participants
4 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 months · Moderate (+3)
0 Participants
0 Participants
4 Participants
6 Participants
4 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 months · Severe (+4)
0 Participants
0 Participants
1 Participants
1 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
1 Year · None (0)/Trival (+1)
40 Participants
156 Participants
245 Participants
200 Participants
89 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
1 Year · Mild (+2)
5 Participants
9 Participants
18 Participants
21 Participants
5 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
1 Year · Moderate (+3)
0 Participants
1 Participants
5 Participants
2 Participants
1 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
1 Year · Severe (+4)
0 Participants
0 Participants
0 Participants
1 Participants
1 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
Discharge · Moderate (+3)
0 Participants
2 Participants
4 Participants
0 Participants
2 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
2 Years · None (0)/Trival (+1)
38 Participants
138 Participants
229 Participants
193 Participants
86 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
6 Years · Moderate (+3)
0 Participants
0 Participants
1 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
2 Years · Mild (+2)
5 Participants
8 Participants
19 Participants
23 Participants
4 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
2 Years · Moderate (+3)
0 Participants
0 Participants
4 Participants
0 Participants
2 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
2 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
1 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 Years · None (0)/Trival (+1)
33 Participants
120 Participants
211 Participants
179 Participants
79 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 Years · Mild (+2)
5 Participants
7 Participants
18 Participants
18 Participants
5 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 Years · Moderate (+3)
0 Participants
0 Participants
3 Participants
0 Participants
2 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
4 Years · None (0)/Trival (+1)
31 Participants
104 Participants
191 Participants
160 Participants
77 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
4 Years · Mild (+2)
4 Participants
9 Participants
17 Participants
19 Participants
5 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
4 Years · Moderate (+3)
0 Participants
0 Participants
1 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
4 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
5 Years · None (0)/Trival (+1)
26 Participants
87 Participants
146 Participants
129 Participants
61 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
5 Years · Moderate (+3)
0 Participants
0 Participants
0 Participants
1 Participants
1 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
5 Years · Mild (+2)
2 Participants
7 Participants
15 Participants
18 Participants
4 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
5 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
1 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
Discharge · Severe (+4)
0 Participants
0 Participants
1 Participants
2 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
3 months · None (0)/Trival (+1)
43 Participants
156 Participants
245 Participants
219 Participants
97 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
6 Years · Mild (+2)
2 Participants
2 Participants
9 Participants
10 Participants
2 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
6 Years · None (0)/Trival (+1)
20 Participants
53 Participants
87 Participants
78 Participants
31 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
Discharge · None (0)/Trival (+1)
46 Participants
171 Participants
266 Participants
235 Participants
97 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
Discharge · Mild (+2)
2 Participants
3 Participants
12 Participants
13 Participants
3 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
6 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
7 Years · None (0)/Trival (+1)
7 Participants
25 Participants
30 Participants
22 Participants
6 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
7 Years · Mild (+2)
2 Participants
1 Participants
5 Participants
6 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
7 Years · Moderate (+3)
0 Participants
0 Participants
0 Participants
1 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
7 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
8 Years · None (0)/Trival (+1)
1 Participants
6 Participants
10 Participants
4 Participants
2 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
8 Years · Mild (+2)
0 Participants
1 Participants
0 Participants
2 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
8 Years · Moderate (+3)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants
Amount of Aortic Valvular Regurgitation in Subjects Over Time by Valve Size
8 Years · Severe (+4)
0 Participants
0 Participants
0 Participants
0 Participants
0 Participants

OTHER_PRE_SPECIFIED outcome

Timeframe: Baseline and one year follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve where data is available at both the baseline and 1 year follow-up visits. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

The Medical Outcomes Study Short-Form 12 (SF-12) - physical and metal states. The SF-12 questionnaire scale ranges from 100, which reflects the best health status to 0, which reflects the worst health status.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=524 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey
Physical Health - Baseline
41.8 Units on a scale
Standard Deviation 10.2
Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey
Mental Health - Baseline
51.3 Units on a scale
Standard Deviation 9.9
Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey
Mental Health - 1 year follow-up
54.3 Units on a scale
Standard Deviation 8.5
Subject's Average Score at Baseline and 1 Year on the Quality of Life Survey
Physical Health - 1 year follow-up
47.6 Units on a scale
Standard Deviation 9.7

OTHER_PRE_SPECIFIED outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Laboratory analysis of White Blood Cell (WBC) count on blood drawn from subject; WBC fight infection. The normal reference range for WBC count is 4.5 to 10 10\^3 cells/microliters. In general, results either lower or higher than the normal range can indicate something is wrong.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average White Blood Cell Count Over Time
Discharge
9.54 10^3 cells/microliters
Standard Deviation 3.0
Subject's Average White Blood Cell Count Over Time
3 months
7.27 10^3 cells/microliters
Standard Deviation 2.5
Subject's Average White Blood Cell Count Over Time
1 Year
7.31 10^3 cells/microliters
Standard Deviation 2.8
Subject's Average White Blood Cell Count Over Time
2 Years
7.47 10^3 cells/microliters
Standard Deviation 4.5
Subject's Average White Blood Cell Count Over Time
3 Years
7.47 10^3 cells/microliters
Standard Deviation 5.7
Subject's Average White Blood Cell Count Over Time
4 Years
7.41 10^3 cells/microliters
Standard Deviation 3.0
Subject's Average White Blood Cell Count Over Time
5 Years
7.37 10^3 cells/microliters
Standard Deviation 3.6
Subject's Average White Blood Cell Count Over Time
6 Years
7.47 10^3 cells/microliters
Standard Deviation 6.1
Subject's Average White Blood Cell Count Over Time
7 Years
8.05 10^3 cells/microliters
Standard Deviation 9.8
Subject's Average White Blood Cell Count Over Time
8 Years
14.0 10^3 cells/microliters
Standard Deviation 36.8

OTHER_PRE_SPECIFIED outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

The red blood cell (RBC) count is a test that measures the number of oxygen-carrying blood cells in your blood. The red blood cell count normal range for women is 4.2 to 5.4 10\^6 cells/microliters and for men is 4.7 to 6.1 10\^6 cells/microliters. A high red blood cell count may indicate that you have a condition that's preventing you from getting enough oxygen. A low RBC may be caused by an infection, or a medical condition related to anemia. In general, results either lower or higher than the normal range can indicate something is wrong.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Red Blood Cell Count Over Time
Discharge
3.31 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
3 months
4.51 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
1 Year
4.46 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
2 Years
4.45 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
3 Years
4.46 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
4 Years
4.44 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
5 Years
4.43 10^6 cells/microliters
Standard Deviation 0.6
Subject's Average Red Blood Cell Count Over Time
6 Years
4.38 10^6 cells/microliters
Standard Deviation 0.5
Subject's Average Red Blood Cell Count Over Time
7 Years
4.33 10^6 cells/microliters
Standard Deviation 0.6
Subject's Average Red Blood Cell Count Over Time
8 Years
4.28 10^6 cells/microliters
Standard Deviation 0.7

OTHER_PRE_SPECIFIED outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Laboratory Analysis of Hematocrit Percentage on blood drawn from subjects. A hematocrit test measures the proportion of red blood cells in your blood. Red blood cells carry oxygen throughout your body. Having too few or too many red blood cells can be a sign of certain diseases. Normal hematocrit ranges for adult females are 36% to 44% and for adult males are 41% to 50%. In general, results either lower or higher than the normal range can indicate something is wrong.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Hematocrit Percentage Over Time
Discharge
29.94 Percentage of RBC
Standard Deviation 4.0
Subject's Average Hematocrit Percentage Over Time
3 months
39.28 Percentage of RBC
Standard Deviation 4.2
Subject's Average Hematocrit Percentage Over Time
1 Year
40.25 Percentage of RBC
Standard Deviation 4.4
Subject's Average Hematocrit Percentage Over Time
2 Years
40.30 Percentage of RBC
Standard Deviation 4.4
Subject's Average Hematocrit Percentage Over Time
3 Years
40.42 Percentage of RBC
Standard Deviation 4.1
Subject's Average Hematocrit Percentage Over Time
4 Years
40.40 Percentage of RBC
Standard Deviation 4.2
Subject's Average Hematocrit Percentage Over Time
5 Years
40.66 Percentage of RBC
Standard Deviation 4.5
Subject's Average Hematocrit Percentage Over Time
6 Years
40.41 Percentage of RBC
Standard Deviation 4.4
Subject's Average Hematocrit Percentage Over Time
7 Years
40.00 Percentage of RBC
Standard Deviation 4.8
Subject's Average Hematocrit Percentage Over Time
8 Years
39.31 Percentage of RBC
Standard Deviation 5.4

OTHER_PRE_SPECIFIED outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Laboratory Analysis of Hemoglobin Count on blood drawn from subjects. Hemoglobin is an oxygen-carrying protein in red blood cells. In males, a healthy hemoglobin level is between 13.2 and 16.6 g/dL. In females, a hemoglobin count of between 11.6 and 15 g/dL is considered normal and healthy. In general, results either lower or higher than the normal range can indicate something is wrong.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Hemoglobin Percentage Over Time
Discharge
9.95 g/dL
Standard Deviation 1.4
Subject's Average Hemoglobin Percentage Over Time
3 months
12.82 g/dL
Standard Deviation 1.6
Subject's Average Hemoglobin Percentage Over Time
1 Year
13.33 g/dL
Standard Deviation 1.6
Subject's Average Hemoglobin Percentage Over Time
2 Years
13.37 g/dL
Standard Deviation 1.6
Subject's Average Hemoglobin Percentage Over Time
3 Years
13.42 g/dL
Standard Deviation 1.6
Subject's Average Hemoglobin Percentage Over Time
4 Years
13.39 g/dL
Standard Deviation 1.6
Subject's Average Hemoglobin Percentage Over Time
5 Years
13.22 g/dL
Standard Deviation 2.2
Subject's Average Hemoglobin Percentage Over Time
6 Years
13.20 g/dL
Standard Deviation 1.8
Subject's Average Hemoglobin Percentage Over Time
7 Years
13.10 g/dL
Standard Deviation 1.8
Subject's Average Hemoglobin Percentage Over Time
8 Years
12.87 g/dL
Standard Deviation 1.9

OTHER_PRE_SPECIFIED outcome

Timeframe: Discharge, 3 months, 1-, 2-, 3-, 4-, 5-, 6-, 7-, and 8- Years follow-up

Population: This outcome is reported for subjects who received an Edwards INTUITY surgical aortic heart valve. Note that the number of participants analyzed is different from that in the participant flow as there was some missing data.

Laboratory Analysis of Plasma Free Hemoglobin (Hgb) of blood drawn from subject. This blood test measures the level of free hemoglobin in the plasma (liquid portion of the blood). The reference range for plasma free hemoglobin is 0.0 to 15.2 mg/dL. In general, values higher than this can indicate something is wrong.

Outcome measures

Outcome measures
Measure
Edwards INTUITY Surgical Aortic Heart Valve
n=885 Participants
Subjects who received an Edwards INTUITY surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class II
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class III
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - Class IV
Subjects who received an INTUITY study surgical aortic heart valve.
Edwards INTUITY Surgical Aortic Heart Valve - 27mm
Subjects who received an INTUITY study surgical aortic heart valve.
Subject's Average Plasma Free Hemoglobin Over Time
Discharge
11.89 mg/dL
Standard Deviation 14.6
Subject's Average Plasma Free Hemoglobin Over Time
3 months
13.12 mg/dL
Standard Deviation 13.5
Subject's Average Plasma Free Hemoglobin Over Time
1 Year
12.08 mg/dL
Standard Deviation 12.9
Subject's Average Plasma Free Hemoglobin Over Time
2 Years
12.06 mg/dL
Standard Deviation 13.5
Subject's Average Plasma Free Hemoglobin Over Time
3 Years
11.90 mg/dL
Standard Deviation 15.7
Subject's Average Plasma Free Hemoglobin Over Time
4 Years
11.88 mg/dL
Standard Deviation 14.5
Subject's Average Plasma Free Hemoglobin Over Time
5 Years
12.73 mg/dL
Standard Deviation 16.9
Subject's Average Plasma Free Hemoglobin Over Time
6 Years
16.32 mg/dL
Standard Deviation 14.4
Subject's Average Plasma Free Hemoglobin Over Time
7 Years
18.43 mg/dL
Standard Deviation 14.9
Subject's Average Plasma Free Hemoglobin Over Time
8 Years
22.61 mg/dL
Standard Deviation 11.4

Adverse Events

Edwards TRANSFORM INTUITY Aortic Valve Study Cohort

Serious events: 756 serious events
Other events: 868 other events
Deaths: 178 deaths

Commercially Available Aortic Valve Study Cohort

Serious events: 37 serious events
Other events: 47 other events
Deaths: 8 deaths

Serious adverse events

Serious adverse events
Measure
Edwards TRANSFORM INTUITY Aortic Valve Study Cohort
n=885 participants at risk
The data from the 885 subjects who received the investigational Edwards INTUITY surgical aortic heart valve.
Commercially Available Aortic Valve Study Cohort
n=49 participants at risk
The data from the 49 subjects who received an alternate FDA approved surgical aortic heart valve.
Blood and lymphatic system disorders
ANEMIA - BLEEDING RELATED - MAJOR
3.5%
31/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
ANEMIA - BLEEDING RELATED - MINOR
5.1%
45/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
10.2%
5/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
ANEMIA - NON-BLEEDING RELATED
3.2%
28/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - CARDIOVASCULAR - MAJOR
2.3%
20/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - CARDIOVASCULAR - MINOR
1.4%
12/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GASTROINTESTINAL LOWER - MAJOR
1.8%
16/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GASTROINTESTINAL LOWER - MINOR
1.9%
17/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GASTROINTESTINAL UPPER - MAJOR
2.0%
18/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GASTROINTESTINAL UPPER - MINOR
1.1%
10/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GENITOURINARY - MAJOR
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - GENITOURINARY - MINOR
0.79%
7/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - MUSCULOSKELETAL/DERMATOLOGICAL - MAJOR (E.G. ECCHYMOSIS)
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - MUSCULOSKELETAL/DERMATOLOGICAL - MINOR (E.G. ECCHYMOSIS)
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - NEUROLOGICAL - MAJOR (E.G. CVA)
1.7%
15/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - NEUROLOGICAL - MINOR (E.G. CVA)
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - PERIPHERAL VASCULAR - MAJOR (E.G. NOSEBLEEDS; HEMATOMAS)
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - PULMONARY/RESPIRATORY - MAJOR (E.G. HEMOTHORAX)
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLEEDING - PULMONARY/RESPIRATORY - MINOR (E.G. HEMOTHORAX)
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLOOD SEPSIS
5.5%
49/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLOOD/LYMPHATIC - OTHER
2.6%
23/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
THROMBOCYTOPENIA - HEPARIN INDUCED (HIT)
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
THROMBOCYTOPENIA - NON-HEPARIN INDUCED
2.6%
23/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ANGINA, STABLE
2.4%
21/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ANGINA, UNSTABLE
2.5%
22/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
AORTIC DISSECTION
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - ATRIAL FLUTTER
4.2%
37/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - AV BLOCK - 1ST DEGREE
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - AV BLOCK - 2ND DEGREE
1.9%
17/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - AV BLOCK - 3RD DEGREE
10.6%
94/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BRADYCARDIA
3.2%
28/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BUNDLE BRANCH BLOCK - LEFT
1.0%
9/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BUNDLE BRANCH BLOCK - RIGHT
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - OTHER
3.8%
34/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PACEMAKER/ICD MALFUNCTION
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PAROXYSMAL ATRIAL FIBRILLATION (PAF)
10.3%
91/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
24.5%
12/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PAROXYSMAL ATRIAL TACHYCARDIA (PAT)
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PERMANENT ATRIAL FIBRILLATION
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PERSISTENT ATRIAL FIBRILLATION
7.7%
68/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
10.2%
5/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - SUPRAVENTRICULAR TACHYCARDIA (SVT)
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - TACHY-BRADYCARDIA
1.4%
12/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - TACHYCARDIA - NON-VENTRICULAR
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - TACHYCARDIA - VENTRICULAR
0.90%
8/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - VENTRICULAR FIBRILLATION
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CARDIAC ARREST
3.1%
27/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CARDIAC DECOMPENSATION
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CARDIOGENIC SHOCK
1.6%
14/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CARDIOVASCULAR - OTHER
11.9%
105/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CORONARY ARTERY OSTIAL OBSTRUCTION
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ENDOCARDITIS
1.2%
11/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HEART FAILURE - ACUTE
7.2%
64/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HEART FAILURE - CHRONIC (CHF)
4.7%
42/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HYPERTENSION - SYSTEMIC
1.8%
16/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HYPOTENSION
2.8%
25/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
MYOCARDIAL INFARCTION
3.8%
34/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERFORATION - VENTRICULAR
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERICARDIAL EFFUSION - MAJOR
1.6%
14/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERICARDIAL EFFUSION - MINOR
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERICARDIAL TAMPONADE
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERICARDITIS
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - AORTIC-CENTRAL/TRANSVALVULAR +4
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - MITRAL-CENTRAL/TRANSVALVULAR +3
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - MITRAL-CENTRAL/TRANSVALVULAR +4
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - MITRAL-INDETERMINATE +3
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - MITRAL-INDETERMINATE +4
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - TRICUSPID-CENTRAL/TRANSVALVULAR +3
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - TRICUSPID-CENTRAL/TRANSVALVULAR +4
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - TRICUSPID-INDETERMINATE +3
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
REGURGITATION - TRICUSPID-INDETERMINATE +4
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
STENOSIS - MITRAL - SEVERE
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
THROMBOEMBOLIC EVENT - OTHER - CENTRAL - NO PARESIS
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
THROMBOEMBOLIC EVENT - STROKE
6.3%
56/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
12.2%
6/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
THROMBOEMBOLIC EVENT - TRANSIENT ISCHEMIC ATTACK (TIA)
3.1%
27/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
VALVE THROMBOSIS - AORTIC
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Ear and labyrinth disorders
HEARING DISORDER
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Endocrine disorders
ENDOCRINE COMPLICATIONS
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Endocrine disorders
PANCREATIC COMPLICATION
1.7%
15/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Eye disorders
VISION DISORDER
4.0%
35/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Gastrointestinal disorders
GASTROINTESTINAL - INFECTION
1.9%
17/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Gastrointestinal disorders
GASTROINTESTINAL - OTHER
12.2%
108/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
ALLERGIC REACTION - MEDICATION RELATED
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
ALLERGIC REACTION - OTHER
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
ANAPHYLACTIC REACTION
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
CANCER - NEWLY DIAGNOSED
8.7%
77/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
CANCER - PROGRESSION OF UNDERLYING DISEASE
2.0%
18/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
FEVER - UNKNOWN ORIGIN
1.1%
10/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
MULTI-SYSTEM ORGAN FAILURE
0.90%
8/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NONSPECIFIC, UNKNOWN, OR OTHER BODY SYSTEM - OTHER COMPLICATION
16.9%
150/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
14.3%
7/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NSD - PARAVALVULAR LEAK +2
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NSD - PARAVALVULAR LEAK +3
0.90%
8/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NSD - PARAVALVULAR LEAK +4
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NSD - PARAVALVULAR LEAK WITH HEMOLYSIS
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NSD - STUDY VALVE STENOSIS - SEVERE
1.0%
9/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
SPEECH DISORDER
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
SVD - STUDY VALVE CALCIFICATION
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Hepatobiliary disorders
BILIARY (GALLBLADDER)
3.3%
29/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Hepatobiliary disorders
HEPATIC COMPLICATION - OTHER
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Hepatobiliary disorders
LIVER FAILURE - ACUTE
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Infections and infestations
INFECTION/INFLAMMATION - OTHER
5.5%
49/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Infections and infestations
STERNAL WOUND/THORACIC INFECTION
1.2%
11/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Infections and infestations
WOUND INFECTION - OTHER
3.5%
31/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Metabolism and nutrition disorders
METABOLIC COMPLICATIONS
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Musculoskeletal and connective tissue disorders
BONE FRACTURE/BREAK
6.4%
57/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Musculoskeletal and connective tissue disorders
MUSCULAR SKELETAL/DERMATOLOGIC - OTHER
20.0%
177/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Psychiatric disorders
PSYCHIATRIC - OTHER
2.3%
20/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Psychiatric disorders
PSYCHIATRIC DISORDER
0.68%
6/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Psychiatric disorders
SUICIDE
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
GENITOURINARY - OTHER
3.5%
31/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL - OTHER
3.7%
33/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL DYSFUNCTION
1.4%
12/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL FAILURE - ACUTE
4.6%
41/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL FAILURE - CHRONIC
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
URINARY TRACT INFECTION (UTI)
4.9%
43/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
ATELECTASIS
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
HYPOXEMIA
1.5%
13/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PLEURAL EFFUSION - BILATERAL
3.2%
28/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PLEURAL EFFUSION - LEFT
2.4%
21/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PLEURAL EFFUSION - RIGHT
3.7%
33/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PNEUMOTHORAX
1.2%
11/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY EDEMA
0.68%
6/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY EMBOLISM - BILATERAL
0.79%
7/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY EMBOLISM - LEFT
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY EMBOLISM - RIGHT
0.68%
6/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY HYPERTENSION
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY/RESPIRATORY - OTHER
3.5%
31/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY DYSFUNCTION/INSUFFICIENCY
1.7%
15/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS)
0.90%
8/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - ASTHMA
0.34%
3/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - COPD
1.8%
16/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - EMPHYSEMA
0.11%
1/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - HEMOTHORAX
0.23%
2/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - OTHER
5.6%
50/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - PNEUMONIA
3.8%
34/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY FAILURE - PNEUMOTHORAX
0.56%
5/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY INFECTION - PNEUMONIA
7.8%
69/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY INFECTION - UPPER (URI)
1.9%
17/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Vascular disorders
VASCULAR - ACCESS SITE COMPLICATION
0.45%
4/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Vascular disorders
VASCULAR - DEEP VEIN THROMBOSIS (DVT)
1.7%
15/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Vascular disorders
VASCULAR - OTHER
5.0%
44/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERFORATION - OTHER
0.00%
0/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.

Other adverse events

Other adverse events
Measure
Edwards TRANSFORM INTUITY Aortic Valve Study Cohort
n=885 participants at risk
The data from the 885 subjects who received the investigational Edwards INTUITY surgical aortic heart valve.
Commercially Available Aortic Valve Study Cohort
n=49 participants at risk
The data from the 49 subjects who received an alternate FDA approved surgical aortic heart valve.
Blood and lymphatic system disorders
THROMBOCYTOPENIA - NON-HEPARIN INDUCED
49.9%
442/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
57.1%
28/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
BLOOD/LYMPHATIC - OTHER
29.3%
259/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
ANEMIA - BLEEDING RELATED - MINOR
22.1%
196/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
20.4%
10/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
ANEMIA - NON-BLEEDING RELATED
9.4%
83/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
18.4%
9/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PAROXYSMAL ATRIAL FIBRILLATION (PAF)
29.6%
262/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
30.6%
15/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BUNDLE BRANCH BLOCK - LEFT
31.5%
279/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
18.4%
9/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - AV BLOCK - 1ST DEGREE
26.2%
232/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
22.4%
11/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - OTHER
28.0%
248/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
24.5%
12/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BRADYCARDIA
15.5%
137/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
10.2%
5/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - BUNDLE BRANCH BLOCK - RIGHT
10.4%
92/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
12.2%
6/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HYPOTENSION
18.2%
161/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
12.2%
6/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
CARDIOVASCULAR - OTHER
17.4%
154/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HYPERTENSION - SYSTEMIC
9.0%
80/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Gastrointestinal disorders
GASTROINTESTINAL - OTHER
15.5%
137/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
10.2%
5/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
URINARY TRACT INFECTION (UTI)
11.9%
105/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
14.3%
7/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Musculoskeletal and connective tissue disorders
MUSCULAR SKELETAL/DERMATOLOGIC - OTHER
26.4%
234/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
16.3%
8/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
General disorders
NONSPECIFIC, UNKNOWN, OR OTHER BODY SYSTEM - OTHER COMPLICATION
28.6%
253/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
16.3%
8/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Psychiatric disorders
PSYCHIATRIC - OTHER
8.7%
77/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
ATELECTASIS
39.3%
348/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
28.6%
14/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PLEURAL EFFUSION - BILATERAL
7.1%
63/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Endocrine disorders
ENDOCRINE COMPLICATIONS
6.9%
61/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL DYSFUNCTION
6.9%
61/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL - OTHER
6.7%
59/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
RENAL FAILURE - ACUTE
6.4%
57/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Infections and infestations
INFECTION/INFLAMMATION - OTHER
9.9%
88/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Musculoskeletal and connective tissue disorders
BONE FRACTURE/BREAK
6.1%
54/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
2.0%
1/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Vascular disorders
VASCULAR - OTHER
6.4%
57/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
4.1%
2/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
PULMONARY/RESPIRATORY - OTHER
12.2%
108/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Respiratory, thoracic and mediastinal disorders
RESPIRATORY INFECTION - UPPER (URI)
6.8%
60/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Blood and lymphatic system disorders
THROMBOCYTOPENIA - HEPARIN INDUCED (HIT)
1.4%
12/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - ATRIAL FLUTTER
5.0%
44/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - PERSISTENT ATRIAL FIBRILLATION
4.3%
38/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
ARRHYTHMIA - TACHYCARDIA - NON-VENTRICULAR
4.1%
36/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
HEART FAILURE - ACUTE
3.3%
29/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Cardiac disorders
PERICARDITIS
1.5%
13/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
6.1%
3/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Renal and urinary disorders
GENITOURINARY - OTHER
4.0%
35/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
8.2%
4/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
Eye disorders
VISION DISORDER
5.3%
47/885 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.
0.00%
0/49 • Events occurring from the day of aortic heart valve surgery up through 8 years post implant.

Additional Information

Jennifer Currin, Senior VP of Medical and Clinical Affairs, Surgical Structural Heart

Edwards Lifesciences, LLC

Phone: 949-250-2500

Results disclosure agreements

  • Principal investigator is a sponsor employee PI can't publish/present on overall study results not yet published but may publish his own data subject to EW review prior to submission/presentation, but data analyses of site-specific results can occur only in intervals as specified in the CTA. Publication/presentation of the PI's site-specific results of devices which haven't been market released and which still may be undergoing development, shall not include claims of device safety/effectiveness and will require the review/approval of EW.
  • Publication restrictions are in place

Restriction type: OTHER