Trial Outcomes & Findings for Evaluation of Blood Glucose Monitoring Systems (NCT NCT01699763)

NCT ID: NCT01699763

Last Updated: 2016-02-29

Results Overview

Using the overall Blood Glucose (BG) range (34 to 561 mg/dL according to YSI subject plasma results), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI plasma results (BG reference) were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

112 participants

Primary outcome timeframe

8 hours

Results posted on

2016-02-29

Participant Flow

Participant milestones

Participant milestones
Measure
Subjects With and Without Diabetes
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
Overall Study
STARTED
112
Overall Study
COMPLETED
112
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Evaluation of Blood Glucose Monitoring Systems

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Subjects With and Without Diabetes
n=112 Participants
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
Age, Continuous
59 years
n=5 Participants
Sex: Female, Male
Female
67 Participants
n=5 Participants
Sex: Female, Male
Male
45 Participants
n=5 Participants
Region of Enrollment
United States
112 participants
n=5 Participants

PRIMARY outcome

Timeframe: 8 hours

Population: Same number 333 (336-3) BG results possible for each BGMS. Staff collected 3 capillary samples from each subject - total 336 samples. Three samples from one subject were not analyzed. Subject hematocrit (58.5) was above the study evaluable limit of 55.

Using the overall Blood Glucose (BG) range (34 to 561 mg/dL according to YSI subject plasma results), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI plasma results (BG reference) were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.

Outcome measures

Outcome measures
Measure
Subjects With and Without Diabetes
n=333 BG test results
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
OneTouch® UltraLink® BGMS
10.8 Percent Difference
Standard Error 0.33
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
Nova Max Link® BGMS
6.4 Percent Difference
Standard Error 0.33
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
Contour® NEXT LINK BGMS
2.7 Percent Difference
Standard Error 0.33

SECONDARY outcome

Timeframe: 8 hours

Population: Same number (107) of BG results was possible for each BGMS. Staff collected 3 capillary samples from each subject (total 333), of which 107 samples were less than or equal to 80 mg/dL.

Using fresh and glycolyzed samples with YSI plasma Blood Glucose (BG) ≤80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.

Outcome measures

Outcome measures
Measure
Subjects With and Without Diabetes
n=107 BG test results
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range (<=80 mg/dL)
Contour® NEXT LINK BGMS
2.63 Percent Difference
Standard Error 0.69
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range (<=80 mg/dL)
OneTouch® UltraLink® BGMS
14.92 Percent Difference
Standard Error 0.69
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range (<=80 mg/dL)
Nova Max Link® BGMS
8.81 Percent Difference
Standard Error 0.69

SECONDARY outcome

Timeframe: 8 hours

Population: Same number (106) of BG results was possible for each BGMS. Staff collected 3 capillary samples from each subject (total 333), of which 106 samples were greater than 180 mg/dL.

Using samples with YSI plasma Blood Glucose (BG) \>180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.

Outcome measures

Outcome measures
Measure
Subjects With and Without Diabetes
n=106 BG test results
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)
Contour® NEXT LINK BGMS
2.66 Percent Difference
Standard Error 0.76
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)
OneTouch® UltraLink® BGMS
8.76 Percent Difference
Standard Error 0.76
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)
Nova Max Link® BGMS
4.52 Percent Difference
Standard Error 0.76

Adverse Events

Subjects With and Without Diabetes

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Subjects With and Without Diabetes
n=112 participants at risk
All testing and lancing were performed by the study staff; subjects with and without diabetes did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
Endocrine disorders
hypoglycemia
1.8%
2/112 • Number of events 2