Trial Outcomes & Findings for Evaluation of the Precision of the Microperimetry Function of the Spectral OCT/SLO (NCT NCT01692938)
NCT ID: NCT01692938
Last Updated: 2017-05-10
Results Overview
A precision study was conducted that used 3 devices (each with a different operator). Each device was used to measure 4 normal subjects and 4 subjects with relevant eye pathology, with a total of 24 subject eyes (12 normal, 12 with pathology) measured across all three devices. For each subject, one eye was evaluated using 3 tests with repositioning at the start of each test. Test results were given in decibels which is the unit used in microperimetry testing. Overall mean and standard deviation in decibels were calculated for the two groups: normal and with pathology.
COMPLETED
32 participants
1 Month
2017-05-10
Participant Flow
Recruitment Dates - 10/5/2012 to 11/5/2012. 3 Ophthalmic Clinics.
Participant milestones
| Measure |
No Retinal Disease
|
Retinal Disease
|
|---|---|---|
|
Overall Study
STARTED
|
16
|
16
|
|
Overall Study
COMPLETED
|
12
|
12
|
|
Overall Study
NOT COMPLETED
|
4
|
4
|
Reasons for withdrawal
| Measure |
No Retinal Disease
|
Retinal Disease
|
|---|---|---|
|
Overall Study
Protocol Deviation
|
1
|
1
|
|
Overall Study
Study completed prior to site completion
|
3
|
3
|
Baseline Characteristics
Evaluation of the Precision of the Microperimetry Function of the Spectral OCT/SLO
Baseline characteristics by cohort
| Measure |
No Retinal Disease
n=12 Participants
|
Retinal Disease
n=12 Participants
|
Total
n=24 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Customized
21-35 years
|
4 participants
n=5 Participants
|
3 participants
n=7 Participants
|
7 participants
n=5 Participants
|
|
Age, Customized
36-50 years
|
4 participants
n=5 Participants
|
3 participants
n=7 Participants
|
7 participants
n=5 Participants
|
|
Age, Customized
51-64 years
|
3 participants
n=5 Participants
|
3 participants
n=7 Participants
|
6 participants
n=5 Participants
|
|
Age, Customized
65 years or great
|
1 participants
n=5 Participants
|
3 participants
n=7 Participants
|
4 participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
8 Participants
n=5 Participants
|
6 Participants
n=7 Participants
|
14 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
4 Participants
n=5 Participants
|
6 Participants
n=7 Participants
|
10 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1 MonthA precision study was conducted that used 3 devices (each with a different operator). Each device was used to measure 4 normal subjects and 4 subjects with relevant eye pathology, with a total of 24 subject eyes (12 normal, 12 with pathology) measured across all three devices. For each subject, one eye was evaluated using 3 tests with repositioning at the start of each test. Test results were given in decibels which is the unit used in microperimetry testing. Overall mean and standard deviation in decibels were calculated for the two groups: normal and with pathology.
Outcome measures
| Measure |
No Retinal Disease
n=12 Participants
|
Retinal Disease
n=12 Participants
|
|---|---|---|
|
Standard Deviation and Mean Test Results of Normal and With Pathology Participants Taken by 3 Different Operator-device Configurations
|
16.16 decibels
Standard Deviation 1.735
|
12.23 decibels
Standard Deviation 3.508
|
PRIMARY outcome
Timeframe: 1 MonthA precision study was conducted that used 3 devices (each with a different operator). Each device was used to measure 4 normal subjects and 4 subjects with relevant eye pathology, with a total of 24 subject eyes (12 normal, 12 with pathology) measured across all three devices. Test results were given in decibels as that is the standard measurement in microperimetry testing. For each subject, one eye was evaluated using 3 microperimetry tests with repositioning at the start of each test. Overall Repeatability SD and Repeatability SD Limit were calculated for the two groups: normal and with pathology.
Outcome measures
| Measure |
No Retinal Disease
n=12 Participants
|
Retinal Disease
n=12 Participants
|
|---|---|---|
|
Repeatability of Microperimetry Tests in Normal and With Pathology Participants Based on 3 Microperimetry Tests Taken for Each Participant for a Given Operator-device Combination.
Repeatability SD
|
16.16 decibels
Standard Deviation 0.531
|
12.23 decibels
Standard Deviation 0.682
|
|
Repeatability of Microperimetry Tests in Normal and With Pathology Participants Based on 3 Microperimetry Tests Taken for Each Participant for a Given Operator-device Combination.
Repeatability SD Limit
|
16.16 decibels
Standard Deviation 1.49
|
12.23 decibels
Standard Deviation 1.91
|
Adverse Events
No Retinal Disease
Retinal Disease
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information