Trial Outcomes & Findings for FAST as a Treatment for Obstructive Sleep Apnea (NCT NCT01611779)

NCT ID: NCT01611779

Last Updated: 2017-02-24

Results Overview

Ability to place the implant and stabilize the tongue

Recruitment status

TERMINATED

Study phase

PHASE4

Target enrollment

5 participants

Primary outcome timeframe

Up to 7 weeks after the procedure

Results posted on

2017-02-24

Participant Flow

Participant milestones

Participant milestones
Measure
Tongue Suspension
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Overall Study
STARTED
5
Overall Study
COMPLETED
5
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

FAST as a Treatment for Obstructive Sleep Apnea

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Tongue Suspension
n=5 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
5 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Gender
Female
1 Participants
n=5 Participants
Gender
Male
4 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
4 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Region of Enrollment
United States
5 participants
n=5 Participants
Body Mass Index (BMI)
26.98 kg/m^2
STANDARD_DEVIATION 3.78 • n=5 Participants
Prior OSA Surgery
5 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Up to 7 weeks after the procedure

Ability to place the implant and stabilize the tongue

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=5 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Place the Implant and Stabilize the Tongue
5 Participants

PRIMARY outcome

Timeframe: 3 months

Patient will be examined by the investigator at each of the follow-up visits for the presence of any untoward or unintended response to the device.

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=5 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Number of Participants Experiencing Complications
Headache
1 Participants
Number of Participants Experiencing Complications
No Complications
4 Participants

SECONDARY outcome

Timeframe: Baseline, 3, and 12 months

Population: Three (3) of 5 subjects returned for 3 month follow-up and 2 of 5 subjects returned for the 12 month follow-up.

0 to \>30/hour (high value represents worse outcome)

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=5 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Apnea Hypopnea Index
Baseline
24.2 events/hour
Standard Deviation 10.3
Apnea Hypopnea Index
3 months
15.9 events/hour
Standard Deviation 7.8
Apnea Hypopnea Index
12 months
34.6 events/hour
Standard Deviation 6.6

SECONDARY outcome

Timeframe: Baseline, 1, 3, 12 months

Population: Four (4) of 5 total subjects had data at baseline and follow-up data.

Questionnaire: 0 to 120 (high value represents better outcome)

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=4 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Functional Outcomes and Sleep Questionnaire (FOSQ)
Baseline
17.2 units on a scale
Standard Deviation 1.6
Functional Outcomes and Sleep Questionnaire (FOSQ)
1 month
18.2 units on a scale
Standard Deviation 1.9
Functional Outcomes and Sleep Questionnaire (FOSQ)
3 months
18.1 units on a scale
Standard Deviation 2.1
Functional Outcomes and Sleep Questionnaire (FOSQ)
12 months
17.2 units on a scale
Standard Deviation 2.4

SECONDARY outcome

Timeframe: Baseline, 1 week; 1 month, 3 months, 12 months

Population: Four (4) of 5 total subjects had data at baseline. Four (4) of 5 subjects had follow-up data at 1 week, 4 had follow-up data at 1 month, 3 had follow-up data at 3 months, and 2 had follow-up data at 12 months.

0 to 10 (high value represents worse outcome)

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=4 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Snoring Scale (VAS)
Baseline
5.5 units on a scale
Standard Deviation 3.4
Snoring Scale (VAS)
1 week
2.6 units on a scale
Standard Deviation 2.3
Snoring Scale (VAS)
1 month
1.8 units on a scale
Standard Deviation 1.0
Snoring Scale (VAS)
3 months
1.8 units on a scale
Standard Deviation 1.6
Snoring Scale (VAS)
12 months
1.7 units on a scale
Standard Deviation 2.4

SECONDARY outcome

Timeframe: Baseline, 1, 3, 12 months

Population: Five (5) of 5 total subjects had data at baseline. Four (4) of 5 subjects had follow-up data at 1 week, 4 had follow-up data at 1 month, 4 had follow-up data at 3 months, and 2 had follow-up data at 12 months.

0 to 24 (high value represents worse outcome)

Outcome measures

Outcome measures
Measure
Tongue Suspension
n=5 Participants
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
Epworth Sleeping Scale (ESS)
Baseline
12.4 units on a scale
Standard Deviation 4.0
Epworth Sleeping Scale (ESS)
1 month
7.7 units on a scale
Standard Deviation 2.2
Epworth Sleeping Scale (ESS)
3 months
7 units on a scale
Standard Deviation 2.4
Epworth Sleeping Scale (ESS)
12 months
6.5 units on a scale
Standard Deviation 2.1

Adverse Events

Tongue Suspension

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Tongue Suspension
n=5 participants at risk
Tongue-based suspension Encore Tongue Suspension System: The primary components of the Encore Tongue Suspension System consist of a suture passer, suspension line and bone screw.
General disorders
Headache
20.0%
1/5 • Number of events 1 • 1 year

Additional Information

Michael Kolber

Siesta Medical, Inc.

Phone: 408-320-9424

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place