Trial Outcomes & Findings for X-ray Dose Reduction in Electrophysiology (NCT NCT01593852)

NCT ID: NCT01593852

Last Updated: 2022-03-08

Results Overview

Percentage reduction of reduced X-ray dose settings (Allura Clarity) vs. regular X-ray dose settings (AlluraXper) in DAP.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

136 participants

Primary outcome timeframe

Day 0

Results posted on

2022-03-08

Participant Flow

Participant milestones

Participant milestones
Measure
Reduced X-ray Dose Settings
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Overall Study
STARTED
68
68
Overall Study
COMPLETED
68
68
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

X-ray Dose Reduction in Electrophysiology

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Reduced X-ray Dose Settings
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Total
n=136 Participants
Total of all reporting groups
Age, Continuous
58 years
STANDARD_DEVIATION 14 • n=5 Participants
65 years
STANDARD_DEVIATION 9 • n=7 Participants
61 years
STANDARD_DEVIATION 12 • n=5 Participants
Sex: Female, Male
Female
18 Participants
n=5 Participants
17 Participants
n=7 Participants
35 Participants
n=5 Participants
Sex: Female, Male
Male
50 Participants
n=5 Participants
51 Participants
n=7 Participants
101 Participants
n=5 Participants
Region of Enrollment
Netherlands
68 participants
n=5 Participants
68 participants
n=7 Participants
136 participants
n=5 Participants

PRIMARY outcome

Timeframe: Day 0

Percentage reduction of reduced X-ray dose settings (Allura Clarity) vs. regular X-ray dose settings (AlluraXper) in DAP.

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Cumulative Dose Area Product (DAP) Value
8.8 Gy*cm^2
Interval 4.62 to 12.8
15.3 Gy*cm^2
Interval 9.14 to 25.53

PRIMARY outcome

Timeframe: Day 0

Percentage reduction of reducted X-ray dose settings (Allura Clarity) vs. regular X-ray dose settings (AlluraXper) in AK.

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Cumulative Air Kerma (AK) Value
75 mGy
Interval 42.85 to 122.73
126 mGy
Interval 85.79 to 226.4

SECONDARY outcome

Timeframe: Day 0

Staff dose measured by Electronic Personal Dosimeter (EPD) worn by the operator over the lead apron (EPD operator) and the other mounted at a fixed location in the EP laboratory (EPD fixed)

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Staff Dose Measured by DoseAware and Electronic Personal Dosimeter (EPD)
EPD operator
3 µSv
Interval 1.0 to 6.3
6 µSv
Interval 3.3 to 10.0
Staff Dose Measured by DoseAware and Electronic Personal Dosimeter (EPD)
EPD fixed
10 µSv
Interval 5.8 to 18.0
19 µSv
Interval 9.0 to 38.5

SECONDARY outcome

Timeframe: Day 0

Measurement of professional judgment (yes/no) of the treating physician.

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Physician Professional Judgment on Procedural Success
91 percentage of procedural success
91 percentage of procedural success

SECONDARY outcome

Timeframe: Day 0

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Procedure Duration
160 minutes
Standard Deviation 58
162 minutes
Standard Deviation 58

SECONDARY outcome

Timeframe: Day 0

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Fluoroscopy Time
24 minutes
Standard Deviation 13
26 minutes
Standard Deviation 13

SECONDARY outcome

Timeframe: Day 0

If fluoroscopy time (see results in other secondary outcomes), number of exposure frames (see below) and procedure duration (see results in other secondary outcomes) are equivalent between the two groups, this will indicate that image quality (IQ) is equally adequate in both groups.

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Physician Professional Judgment on Adequacy of Images for Performing the EP Procedure
83 Number of exposure frames
Standard Deviation 89
61 Number of exposure frames
Standard Deviation 66

SECONDARY outcome

Timeframe: Day 0

In both groups, default fluoroscopy dose settings were 'low', but could be changed by the operator to a medium or high setting for better imaging. The necessity to increase fluoroscopy dose for better imaging to medium or high was registered as a percentage of total number of fluoroscopy frames.

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Usage of Physician Controlled Dose Settings
Low
86 percentage of total # of fluoro frames
Standard Deviation 28
88 percentage of total # of fluoro frames
Standard Deviation 20
Usage of Physician Controlled Dose Settings
Medium
2 percentage of total # of fluoro frames
Standard Deviation 8
2 percentage of total # of fluoro frames
Standard Deviation 9
Usage of Physician Controlled Dose Settings
High
12 percentage of total # of fluoro frames
Standard Deviation 24
9 percentage of total # of fluoro frames
Standard Deviation 19

SECONDARY outcome

Timeframe: Day 0 if any

Outcome measures

Outcome measures
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 Participants
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 Participants
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Serious Adverse Events
1 participants
0 participants

Adverse Events

Reduced X-ray Dose Settings (Allura Clarity)

Serious events: 1 serious events
Other events: 0 other events
Deaths: 0 deaths

Regular X-ray Dose Settings (AlluraXper)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Reduced X-ray Dose Settings (Allura Clarity)
n=68 participants at risk
For patients in this group x-ray images are acquired with reduced dose settings of the x-ray system and advanced image processing Advanced image processing : Acquisition of x-ray images with reduced X-ray dose and advanced image processing
Regular X-ray Dose Settings (AlluraXper)
n=68 participants at risk
For patients in this group x-ray images are acquired with regular dose settings of the x-ray system and regular image processing
Cardiac disorders
Pericardial tamponade
1.5%
1/68 • Number of events 1 • During procedure
0.00%
0/68 • During procedure

Other adverse events

Adverse event data not reported

Additional Information

Lukas Dekker, MD, PhD

Catharina Hospital Eindhoven

Phone: +31 40 239 7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place