Trial Outcomes & Findings for Carestream Digital Radiography Long Length Imaging Software Data Collection Protocol (NCT NCT01592435)
NCT ID: NCT01592435
Last Updated: 2017-04-06
Results Overview
1-Non-diagnostic - Unacceptable for diagnostic purposes. 2-Limited - Acceptable, with some technical defect. 3-Diagnostic - Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary - Good, most adequate for diagnostic purposes.
COMPLETED
76 participants
5 weeks after completion of data collection
2017-04-06
Participant Flow
Participant milestones
| Measure |
Pred. & Invest.-All Study Participants
Cedera AccuStitch Software is standard of care software currently used at sites.
Carestream DR LLI software is investigational software used for reconstruction.
|
|---|---|
|
Overall Study
STARTED
|
76
|
|
Overall Study
COMPLETED
|
76
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Carestream Digital Radiography Long Length Imaging Software Data Collection Protocol
Baseline characteristics by cohort
| Measure |
Pred. & Invest.-All Study Participants
n=76 Participants
Cedera AccuStitch Software is standard of care software currently used at sites.
Carestream DR LLI software is investigational software used for reconstruction
|
|---|---|
|
Age, Categorical
<=18 years
|
19 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
57 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
44 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
32 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 5 weeks after completion of data collection1-Non-diagnostic - Unacceptable for diagnostic purposes. 2-Limited - Acceptable, with some technical defect. 3-Diagnostic - Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary - Good, most adequate for diagnostic purposes.
Outcome measures
| Measure |
Predicate - Cedera AccuStitch Software
n=76 Participants
Cedera AccuStitch Software is standard of care software currently used at sites.
|
|---|---|
|
Radlex Scale for Diagnostic Capability Ratings - Cedara Accustitch Software (Predicate)
|
3.01 units on a scale
Standard Error 0.04
|
PRIMARY outcome
Timeframe: 5 weeks after completion of data collection1-Non-diagnostic - Unacceptable for diagnostic purposes. 2-Limited - Acceptable, with some technical defect. 3-Diagnostic - Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary - Good, most adequate for diagnostic purposes.
Outcome measures
| Measure |
Predicate - Cedera AccuStitch Software
n=76 Participants
Cedera AccuStitch Software is standard of care software currently used at sites.
|
|---|---|
|
Radlex Scale for Diagnostic Capability Ratings - Carestream DR LLI Software (Investigational)
|
3.16 units on a scale
Standard Error 0.07
|
Adverse Events
Predicate - Cedera AccuStitch Software
Invest. - Carestream DR LLI Software
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Susan Pate, MS, RN, RAC, Senior Manager, Clinical Affairs
Carestream Health, Inc.
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place