Trial Outcomes & Findings for Comparison Evaluation of Blood Glucose Meter Systems (NCT NCT01548885)
NCT ID: NCT01548885
Last Updated: 2016-02-29
Results Overview
Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
COMPLETED
NA
146 participants
10 hours
2016-02-29
Participant Flow
Participant milestones
| Measure |
Study Staff Test BGMSs
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; ACCU-CHEK® Aviva BGMS; CONTOUR® NEXT EZ BGMS.
|
|---|---|
|
Overall Study
STARTED
|
146
|
|
Overall Study
COMPLETED
|
146
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Comparison Evaluation of Blood Glucose Meter Systems
Baseline characteristics by cohort
| Measure |
Study Staff Test BGMSs
n=146 Participants
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; ACCU-CHEK® Aviva BGMS; CONTOUR® NEXT EZ BGMS.
|
|---|---|
|
Age, Continuous
|
62 years
n=5 Participants
|
|
Sex: Female, Male
Female
|
85 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
61 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
146 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 10 hoursPopulation: Same number 388(393-3-2)BG results possible for each BGMS.Subjects provided 1,2,or 3 capillary samples-total 393 samples. 3 glycolyzed samples were not analyzed - glycolysis exceeded the protocol-defined time. All meter data for 2 glycolyzed samples were not evaluable (not analyzed)- their YSI results were less than the meter operating ranges.
Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
Outcome measures
| Measure |
Study Staff Test BGMSs
n=388 BG test results
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; ACCU-CHEK® Aviva BGMS; CONTOUR® NEXT EZ BGMS.
|
|---|---|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
CONTOUR® NEXT EZ BGMS
|
4.7 percentage
Interval 0.0 to 11.4
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
ACCU-CHEK® Aviva BGMS
|
6.3 percentage
Interval 0.0 to 13.0
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
FreeStyle Freedom Lite® BGMS
|
18.3 percentage
Interval 11.6 to 25.0
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
OneTouch® Ultra®2 BGMS
|
23.4 percentage
Interval 16.7 to 30.1
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
TRUEtrack® BGMS
|
26.2 percentage
Interval 19.5 to 32.9
|
SECONDARY outcome
Timeframe: 10 hoursPopulation: Same number (190) of BG results was possible for each BGMS. Subjects provided 1,2,or 3 capillary samples, of which 190 samples were less than 70 mg/dL.
Using fresh and glycolyzed samples with Blood Glucose(BG) below 70 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
Outcome measures
| Measure |
Study Staff Test BGMSs
n=190 BG test results
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; ACCU-CHEK® Aviva BGMS; CONTOUR® NEXT EZ BGMS.
|
|---|---|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
CONTOUR® NEXT EZ BGMS
|
0.7 percentage
Interval 0.0 to 6.8
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
ACCU-CHEK® Aviva BGMS
|
2.5 percentage
Interval 0.0 to 8.7
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
FreeStyle Freedom Lite® BGMS
|
18.3 percentage
Interval 12.1 to 28.6
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
OneTouch® Ultra®2 BGMS
|
22.4 percentage
Interval 16.2 to 28.6
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
TRUEtrack® BGMS
|
33.2 percentage
Interval 27.1 to 39.4
|
Adverse Events
Study Staff Test BGMSs
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Study Staff Test BGMSs
n=146 participants at risk
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; ACCU-CHEK® Aviva BGMS; CONTOUR® NEXT EZ BGMS.
|
|---|---|
|
Endocrine disorders
hypoglycemia
|
2.7%
4/146 • Number of events 4
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place