Trial Outcomes & Findings for User Study of an Investigational Blood Glucose Monitoring System (NCT NCT01447121)

NCT ID: NCT01447121

Last Updated: 2016-02-29

Results Overview

Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

118 participants

Primary outcome timeframe

1 hour

Results posted on

2016-02-29

Participant Flow

Participant milestones

Participant milestones
Measure
Intended Users of the System
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick. Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
Overall Study
STARTED
118
Overall Study
COMPLETED
118
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

User Study of an Investigational Blood Glucose Monitoring System

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Intended Users of the System
n=116 Participants
Study results from 2 subjects were excluded from all data analysis, because study staff inadvertently performed study tasks that could be considered 'training' them. (According to protocol, training was not allowed.)
Age, Customized
52.3 years
n=5 Participants
Sex: Female, Male
Female
61 Participants
n=5 Participants
Sex: Female, Male
Male
55 Participants
n=5 Participants
Region of Enrollment
United States
116 participants
n=5 Participants

PRIMARY outcome

Timeframe: 1 hour

Population: 115(115x1)test results are possible. 118 subjects tested one of 3 test strip lots on the BGM system. Study results from 2 subjects were excluded from all data analyses as already described. One subject test exceeded time interval (defined in protocol) between meter test and reference method.

Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).

Outcome measures

Outcome measures
Measure
Intended Users of the System
n=115 BG test results
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick. Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
115 BG test results

SECONDARY outcome

Timeframe: 1 hour

Population: 115 (115x1) test results are possible. Staff tested blood from 118 subjects (one of 3 test strip lots). All study results from 2 subjects were excluded from data analyses as already described. Also, one test exceeded time interval (defined in protocol) between meter test and reference method.

Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results.

Outcome measures

Outcome measures
Measure
Intended Users of the System
n=115 Participants
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick. Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
Number of Study Staff Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method When Testing Subject Blood Glucose (BG)
115 BG test results

Adverse Events

Intended Users of the System

Serious events: 0 serious events
Other events: 5 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Intended Users of the System
n=118 participants at risk
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick. Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
Endocrine disorders
hypoglycemia
4.2%
5/118 • Number of events 5