Trial Outcomes & Findings for User Study of an Investigational Blood Glucose Monitoring System (NCT NCT01447121)
NCT ID: NCT01447121
Last Updated: 2016-02-29
Results Overview
Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).
COMPLETED
NA
118 participants
1 hour
2016-02-29
Participant Flow
Participant milestones
| Measure |
Intended Users of the System
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick.
Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
|
|---|---|
|
Overall Study
STARTED
|
118
|
|
Overall Study
COMPLETED
|
118
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
User Study of an Investigational Blood Glucose Monitoring System
Baseline characteristics by cohort
| Measure |
Intended Users of the System
n=116 Participants
Study results from 2 subjects were excluded from all data analysis, because study staff inadvertently performed study tasks that could be considered 'training' them. (According to protocol, training was not allowed.)
|
|---|---|
|
Age, Customized
|
52.3 years
n=5 Participants
|
|
Sex: Female, Male
Female
|
61 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
55 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
116 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1 hourPopulation: 115(115x1)test results are possible. 118 subjects tested one of 3 test strip lots on the BGM system. Study results from 2 subjects were excluded from all data analyses as already described. One subject test exceeded time interval (defined in protocol) between meter test and reference method.
Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).
Outcome measures
| Measure |
Intended Users of the System
n=115 BG test results
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick.
Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
|
|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
|
115 BG test results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 115 (115x1) test results are possible. Staff tested blood from 118 subjects (one of 3 test strip lots). All study results from 2 subjects were excluded from data analyses as already described. Also, one test exceeded time interval (defined in protocol) between meter test and reference method.
Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results.
Outcome measures
| Measure |
Intended Users of the System
n=115 Participants
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick.
Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
|
|---|---|
|
Number of Study Staff Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method When Testing Subject Blood Glucose (BG)
|
115 BG test results
|
Adverse Events
Intended Users of the System
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Intended Users of the System
n=118 participants at risk
Untrained subjects with diabetes use an investigational blood glucose monitoring system (Tatsu/Tradewind Investigational BG Monitoring System) to self-test capillary blood obtained from fingerstick.
Tatsu/Tradewind Investigational BG Monitoring System : Tradewind is a Bayer investigational meter that used an investigational sensor.
|
|---|---|
|
Endocrine disorders
hypoglycemia
|
4.2%
5/118 • Number of events 5
|
Additional Information
Carmine Greene, Senior Clinical Research Scientist
Ascensia Diabetes Care
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60