Trial Outcomes & Findings for Evaluation of Capsule Endoscopy With PillCam® COLON 2 in Visualization of the Colon (NCT NCT01372878)

NCT ID: NCT01372878

Last Updated: 2019-08-01

Results Overview

Sensitivity and specificity of PillCam Platform with the PillCam COLON 2 capsule in detecting subjects with polyps equal to or larger than 6 mm. For a given polyp, a match between the PillCam Colon 2 Capsule and optical colonoscopy was considered if the polyp size was assessed within plus or minus 50% of the size of the estimate of the OC measurement and the polyp as appearing within the same colon segment or in adjacent segments. The polyp size measurement by optical colonoscopy was used as the reference standard. Sensitivity measures the proportion of actual positives which are correctly identified as such. Specificity measures the proportion of negatives which are correctly identified as such. positive event defined as patients with polyps ≥6 mm detected by OC procedure.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

884 participants

Primary outcome timeframe

1 year, same as study duration

Results posted on

2019-08-01

Participant Flow

Participant milestones

Participant milestones
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Overall Study
STARTED
884
Overall Study
COMPLETED
700
Overall Study
NOT COMPLETED
184

Reasons for withdrawal

Reasons for withdrawal
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Overall Study
Withdrawal by Subject
63
Overall Study
Protocol Violation
17
Overall Study
incomplete study procedure
104

Baseline Characteristics

Evaluation of Capsule Endoscopy With PillCam® COLON 2 in Visualization of the Colon

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
n=884 Participants
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Age, Continuous
57.29 years
STANDARD_DEVIATION 5.84 • n=113 Participants
Sex: Female, Male
Female
487 Participants
n=113 Participants
Sex: Female, Male
Male
397 Participants
n=113 Participants
Region of Enrollment
United States
558 participants
n=113 Participants
Region of Enrollment
Israel
326 participants
n=113 Participants

PRIMARY outcome

Timeframe: 1 year, same as study duration

Sensitivity and specificity of PillCam Platform with the PillCam COLON 2 capsule in detecting subjects with polyps equal to or larger than 6 mm. For a given polyp, a match between the PillCam Colon 2 Capsule and optical colonoscopy was considered if the polyp size was assessed within plus or minus 50% of the size of the estimate of the OC measurement and the polyp as appearing within the same colon segment or in adjacent segments. The polyp size measurement by optical colonoscopy was used as the reference standard. Sensitivity measures the proportion of actual positives which are correctly identified as such. Specificity measures the proportion of negatives which are correctly identified as such. positive event defined as patients with polyps ≥6 mm detected by OC procedure.

Outcome measures

Outcome measures
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
n=700 Participants
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Sensitivity and Specificity of PillCam Platform With the PillCam COLON 2 Capsule in Detecting Patients With Polyps ≥6 mm Where OC Considered as the Gold Standard Reference
sensitivity of patients with Polyp>= 6 mm
69 percentage of participants
Interval 63.0 to 74.0
Sensitivity and Specificity of PillCam Platform With the PillCam COLON 2 Capsule in Detecting Patients With Polyps ≥6 mm Where OC Considered as the Gold Standard Reference
specificty of patients with Polyp>= 6 mm
87 percentage of participants
Interval 85.0 to 89.0

SECONDARY outcome

Timeframe: 1 year, same as study duration

Sensitivity and specificity of PillCam Platform with the PillCam COLON 2 capsule in detecting subjects with polyps equal to or larger than 6 mm. For a given polyp, a match between the PillCam Colon 2 Capsule and optical colonoscopy was considered if the polyp size was assessed within plus or minus 50% of the size of the estimate of the OC measurement and the polyp as appearing within the same colon segment or in adjacent segments. The polyp size measurement by optical colonoscopy was used as the reference standard.

Outcome measures

Outcome measures
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
n=700 Participants
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Sensitivity and Specificity of PillCam Platform With the PillCam COLON 2 Capsule in Detecting Patients With Polyps ≥10 mm Where OC Considered as the Gold Standard Reference
sensitivity of patients with Polyp>= 10 mm
64 percentage of participants
Interval 55.0 to 73.0
Sensitivity and Specificity of PillCam Platform With the PillCam COLON 2 Capsule in Detecting Patients With Polyps ≥10 mm Where OC Considered as the Gold Standard Reference
specificity of patients with Polyp>= 10 mm
95 percentage of participants
Interval 94.0 to 96.0

Adverse Events

Colon Capsule Endoscopy, Then Standard Colonoscopy

Serious events: 1 serious events
Other events: 110 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
n=884 participants at risk
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Gastrointestinal disorders
abdominal pain related to OC procedure
0.11%
1/884 • Number of events 1

Other adverse events

Other adverse events
Measure
Colon Capsule Endoscopy, Then Standard Colonoscopy
n=884 participants at risk
Capsule endoscopy was ingested following colon preparation without colon insufflation or sedation.The purpose was to detect patients with polyps equal or larger than 6mm. Patients subsequently had standard colonoscopy as "gold standard" comparison
Gastrointestinal disorders
Adverse event related to preperation
11.4%
101/884 • Number of events 128
Gastrointestinal disorders
Adverse event related to Capsule endoscopy
0.34%
3/884 • Number of events 3
Gastrointestinal disorders
Adverse event related to Colonoscopy
0.68%
6/884 • Number of events 11

Additional Information

Ravit Peled | Biostatistics and Data Management Manager

Given Imaging

Phone: +972 (4) 909-7894

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place