Trial Outcomes & Findings for PMP-300E (Smart Watch): Portable Monitoring Device Study (NCT NCT01364740)
NCT ID: NCT01364740
Last Updated: 2017-01-27
Results Overview
Apnea-Hypopnea Index (number of events/hour of sleep). Hypopneas scored without EEG arousals.
COMPLETED
PHASE3
10 participants
One night
2017-01-27
Participant Flow
Patients were recruited through the patient population at the Stanford Sleep Medicine Center
Participant milestones
| Measure |
SmartWatch PMP-300E
Investigational device Smart Watch PMP-300E to measure sleep-related breathing was tested against conventional gold-standard In-Lab Polysomnography (sleep studies)
Smart Watch PMP-300E: Data collected from a Level 3 portable monitoring device
In-lab Polysomnography: Data collected from Type I In-Lab Polysomnography
|
|---|---|
|
Overall Study
STARTED
|
10
|
|
Overall Study
COMPLETED
|
10
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
PMP-300E (Smart Watch): Portable Monitoring Device Study
Baseline characteristics by cohort
| Measure |
PMP-300E, In-Lab PSG
n=10 Participants
PMP-300E, In-Lab PSG: PMP-300E, A 7-channel (nasal pressure, effort, snoring, SpO2, pulse rate, body position and movement) Level 3 portable monitor (11.2 x 3.3 x 5.5cm, 80g, Pacific Medico Co., LTD) to measure sleep-related breathing will be tested against conventional gold-standard In-Lab Polysomnography (sleep study)
PMP-300E: Data collected from Level 3 device
In-lab PSG: Data collected from Type I In-Lab Polysomnography
|
|---|---|
|
Age, Customized
Men
|
54.3 years
STANDARD_DEVIATION 15.46 • n=5 Participants
|
|
Age, Customized
Women
|
56.8 years
STANDARD_DEVIATION 14.24 • n=5 Participants
|
|
Gender
Female
|
4 Participants
n=5 Participants
|
|
Gender
Male
|
6 Participants
n=5 Participants
|
|
Body Mass Index
Men
|
29.5 kg/m^2
STANDARD_DEVIATION 9.54 • n=5 Participants
|
|
Body Mass Index
Women
|
30.8 kg/m^2
STANDARD_DEVIATION 11.76 • n=5 Participants
|
|
Epworth Sleepiness Scale score
|
6.9 units on a scale
STANDARD_DEVIATION 5.61 • n=5 Participants
|
PRIMARY outcome
Timeframe: One nightApnea-Hypopnea Index (number of events/hour of sleep). Hypopneas scored without EEG arousals.
Outcome measures
| Measure |
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
|
|---|---|
|
AHI
PMP-300E
|
19.2 events/hour
Standard Deviation 14.50
|
|
AHI
In-Lab PSG
|
19.3 events/hour
Standard Deviation 14.5
|
SECONDARY outcome
Timeframe: One nightNumber of events/hour of sleep
Outcome measures
| Measure |
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
|
|---|---|
|
Apnea Index
PMP-300E
|
7.7 events/hour
Standard Deviation 11.06
|
|
Apnea Index
In-Lab PSG
|
9.8 events/hour
Standard Deviation 13.32
|
SECONDARY outcome
Timeframe: One nightNumber of events/hour of sleep. Hypopneas scored without EEG arousals.
Outcome measures
| Measure |
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
|
|---|---|
|
Hypopnea Index
PMP-300E
|
8.8 events/hour
Standard Deviation 4.96
|
|
Hypopnea Index
In-Lab PSG
|
14.1 events/hour
Standard Deviation 19.0
|
SECONDARY outcome
Timeframe: One nightMean oxygen saturation
Outcome measures
| Measure |
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
|
|---|---|
|
Oxygen Saturation
PMP-300E
|
94.8 percent
Standard Deviation 2.10
|
|
Oxygen Saturation
In-Lab PSG
|
95.0 percent
Standard Deviation 2.11
|
OTHER_PRE_SPECIFIED outcome
Timeframe: one nightQuestionnaire consisting of multiple questions, scored as 1 = disagree completely to 5 = agree completely, revealed the following mean scores
Outcome measures
| Measure |
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
|
|---|---|
|
Attitude Toward Device Use Questionnaire Scores
Confident in using device at home
|
4.3 units on a scale
Standard Deviation 0.82
|
|
Attitude Toward Device Use Questionnaire Scores
Confident in applying nasal cannula on own
|
4.4 units on a scale
Standard Deviation 0.84
|
|
Attitude Toward Device Use Questionnaire Scores
confident in applying finger probe on own
|
4.9 units on a scale
Standard Deviation 0.32
|
|
Attitude Toward Device Use Questionnaire Scores
Can read window display without difficulty
|
3.9 units on a scale
Standard Deviation 0.99
|
|
Attitude Toward Device Use Questionnaire Scores
Can hear the on/off beep without difficulty
|
3.8 units on a scale
Standard Deviation 1.03
|
|
Attitude Toward Device Use Questionnaire Scores
Can strap on chest belt without difficulty
|
4.5 units on a scale
Standard Deviation 0.71
|
|
Attitude Toward Device Use Questionnaire Scores
Importance of finding out more about own sleep
|
4.8 units on a scale
Standard Deviation 0.42
|
|
Attitude Toward Device Use Questionnaire Scores
Bed partner is a restless sleeper or loud snorer
|
1.2 units on a scale
Standard Deviation 1.40
|
Adverse Events
PMP-300E, In-Lab Polysomnography
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place