Trial Outcomes & Findings for PMP-300E (Smart Watch): Portable Monitoring Device Study (NCT NCT01364740)

NCT ID: NCT01364740

Last Updated: 2017-01-27

Results Overview

Apnea-Hypopnea Index (number of events/hour of sleep). Hypopneas scored without EEG arousals.

Recruitment status

COMPLETED

Study phase

PHASE3

Target enrollment

10 participants

Primary outcome timeframe

One night

Results posted on

2017-01-27

Participant Flow

Patients were recruited through the patient population at the Stanford Sleep Medicine Center

Participant milestones

Participant milestones
Measure
SmartWatch PMP-300E
Investigational device Smart Watch PMP-300E to measure sleep-related breathing was tested against conventional gold-standard In-Lab Polysomnography (sleep studies) Smart Watch PMP-300E: Data collected from a Level 3 portable monitoring device In-lab Polysomnography: Data collected from Type I In-Lab Polysomnography
Overall Study
STARTED
10
Overall Study
COMPLETED
10
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

PMP-300E (Smart Watch): Portable Monitoring Device Study

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
PMP-300E, In-Lab PSG
n=10 Participants
PMP-300E, In-Lab PSG: PMP-300E, A 7-channel (nasal pressure, effort, snoring, SpO2, pulse rate, body position and movement) Level 3 portable monitor (11.2 x 3.3 x 5.5cm, 80g, Pacific Medico Co., LTD) to measure sleep-related breathing will be tested against conventional gold-standard In-Lab Polysomnography (sleep study) PMP-300E: Data collected from Level 3 device In-lab PSG: Data collected from Type I In-Lab Polysomnography
Age, Customized
Men
54.3 years
STANDARD_DEVIATION 15.46 • n=5 Participants
Age, Customized
Women
56.8 years
STANDARD_DEVIATION 14.24 • n=5 Participants
Gender
Female
4 Participants
n=5 Participants
Gender
Male
6 Participants
n=5 Participants
Body Mass Index
Men
29.5 kg/m^2
STANDARD_DEVIATION 9.54 • n=5 Participants
Body Mass Index
Women
30.8 kg/m^2
STANDARD_DEVIATION 11.76 • n=5 Participants
Epworth Sleepiness Scale score
6.9 units on a scale
STANDARD_DEVIATION 5.61 • n=5 Participants

PRIMARY outcome

Timeframe: One night

Apnea-Hypopnea Index (number of events/hour of sleep). Hypopneas scored without EEG arousals.

Outcome measures

Outcome measures
Measure
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
AHI
PMP-300E
19.2 events/hour
Standard Deviation 14.50
AHI
In-Lab PSG
19.3 events/hour
Standard Deviation 14.5

SECONDARY outcome

Timeframe: One night

Number of events/hour of sleep

Outcome measures

Outcome measures
Measure
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
Apnea Index
PMP-300E
7.7 events/hour
Standard Deviation 11.06
Apnea Index
In-Lab PSG
9.8 events/hour
Standard Deviation 13.32

SECONDARY outcome

Timeframe: One night

Number of events/hour of sleep. Hypopneas scored without EEG arousals.

Outcome measures

Outcome measures
Measure
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
Hypopnea Index
PMP-300E
8.8 events/hour
Standard Deviation 4.96
Hypopnea Index
In-Lab PSG
14.1 events/hour
Standard Deviation 19.0

SECONDARY outcome

Timeframe: One night

Mean oxygen saturation

Outcome measures

Outcome measures
Measure
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
Oxygen Saturation
PMP-300E
94.8 percent
Standard Deviation 2.10
Oxygen Saturation
In-Lab PSG
95.0 percent
Standard Deviation 2.11

OTHER_PRE_SPECIFIED outcome

Timeframe: one night

Questionnaire consisting of multiple questions, scored as 1 = disagree completely to 5 = agree completely, revealed the following mean scores

Outcome measures

Outcome measures
Measure
PMP-300E, In-Lab Polysomnography
n=10 Participants
Patient data from one night with the PMP-300E, compared to In-Lab PSG data
Attitude Toward Device Use Questionnaire Scores
Confident in using device at home
4.3 units on a scale
Standard Deviation 0.82
Attitude Toward Device Use Questionnaire Scores
Confident in applying nasal cannula on own
4.4 units on a scale
Standard Deviation 0.84
Attitude Toward Device Use Questionnaire Scores
confident in applying finger probe on own
4.9 units on a scale
Standard Deviation 0.32
Attitude Toward Device Use Questionnaire Scores
Can read window display without difficulty
3.9 units on a scale
Standard Deviation 0.99
Attitude Toward Device Use Questionnaire Scores
Can hear the on/off beep without difficulty
3.8 units on a scale
Standard Deviation 1.03
Attitude Toward Device Use Questionnaire Scores
Can strap on chest belt without difficulty
4.5 units on a scale
Standard Deviation 0.71
Attitude Toward Device Use Questionnaire Scores
Importance of finding out more about own sleep
4.8 units on a scale
Standard Deviation 0.42
Attitude Toward Device Use Questionnaire Scores
Bed partner is a restless sleeper or loud snorer
1.2 units on a scale
Standard Deviation 1.40

Adverse Events

PMP-300E, In-Lab Polysomnography

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Clete A. Kushida, MD, PhD

Stanford University

Phone: 650-721-7560

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place