Trial Outcomes & Findings for A Study Evaluating OsseoSpeed TX Using Different Drilling Protocols With Immediate Loading (NCT NCT01340170)

NCT ID: NCT01340170

Last Updated: 2024-11-15

Results Overview

Marginal Bone Level alterations will be determined from radiographs and expressed as the difference from a reference point on the implant to the most coronal bone-to-implant contact on the mesial and distal aspect of the implant.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

168 participants

Primary outcome timeframe

Implant insertion and 12 months post implant insertion

Results posted on

2024-11-15

Participant Flow

Unit of analysis: Implants

Participant milestones

Participant milestones
Measure
Standard Drilling Protocol (Group A)
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Group A + B
Subjects with implants from both Group A and Group B
Overall Study
STARTED
79 83
86 99
3 6
Overall Study
Permanent Restoration (Start + 3 Months)
79 83
81 94
3 6
Overall Study
COMPLETED
76 79
79 91
3 6
Overall Study
NOT COMPLETED
3 4
7 8
0 0

Reasons for withdrawal

Reasons for withdrawal
Measure
Standard Drilling Protocol (Group A)
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Group A + B
Subjects with implants from both Group A and Group B
Overall Study
Lost to Follow-up
2
4
0
Overall Study
Implant failure
1
3
0

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Standard Drilling Protocol (Group A)
n=79 Participants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=86 Participants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Group A + B
n=3 Participants
Subjects with implants from both Group A and Group B
Total
n=168 Participants
Total of all reporting groups
Age, Continuous
46.6 Years
STANDARD_DEVIATION 12.6 • n=79 Participants
47.5 Years
STANDARD_DEVIATION 12.3 • n=86 Participants
54.3 Years
STANDARD_DEVIATION 17.7 • n=3 Participants
47.2 Years
STANDARD_DEVIATION 12.5 • n=168 Participants
Sex: Female, Male
Female
43 Participants
n=79 Participants
44 Participants
n=86 Participants
3 Participants
n=3 Participants
90 Participants
n=168 Participants
Sex: Female, Male
Male
36 Participants
n=79 Participants
42 Participants
n=86 Participants
0 Participants
n=3 Participants
78 Participants
n=168 Participants
Race and Ethnicity Not Collected
0 Participants
Race and Ethnicity were not collected from any participant.

PRIMARY outcome

Timeframe: Implant insertion and 12 months post implant insertion

Population: ITT subjects with MBL data available (MBL data missing for 14 implants in Group A, and from 10 implants in group B). Outcome measure 1 is analyzed on implant level only and does not contain subject level group "Group A + B".

Marginal Bone Level alterations will be determined from radiographs and expressed as the difference from a reference point on the implant to the most coronal bone-to-implant contact on the mesial and distal aspect of the implant.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=68 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=84 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Marginal Bone Level Change
-0.25 mm
Standard Deviation 0.9
-0.2 mm
Standard Deviation 0.67

SECONDARY outcome

Timeframe: Implant insertion and 3 months post implant insertion

Population: ITT subjects with MBL data available (MBL data missing from 10 implants in Group A, from 9 implants in group B). Outcome measure 2 is analyzed on implant level only and does not contain subject level group "Group A + B".

Marginal Bone Level alterations will be determined from radiographs and expressed as the difference from a reference point on the implant to the most coronal bone-to-implant contact on the mesial and distal aspect of the implant.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=76 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=88 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Marginal Bone Level Change
-0.2 mm
Standard Deviation 0.62
-0.29 mm
Standard Deviation 0.67

SECONDARY outcome

Timeframe: 12 months after baseline, i.e., implant insertion

Population: ITT subjects at 12 months as well as known implant failures already terminated are included in the analysis. Outcome measure 3 is analyzed on implant level only and does not contain subject level group "Group A + B".

Implant survival rate will be evaluated by both clinically and radiographically counting the number of remaining implants from implant placement to one year after loading.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=84 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=98 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Implant Survival Rate
Implants confirmed success
82 Implants
94 Implants
Implant Survival Rate
Implants confirmed failure
2 Implants
4 Implants

SECONDARY outcome

Timeframe: At implant insertion

Population: ITT population. Outcome measure 4 is analyzed on implant level only and does not contain subject level group "Group A + B".

Evaluated clinically/manually and recorded as stable yes/no.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=86 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=102 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Implant Stability Assessed Clinically/Manually
Implant stability - No
0 Implants
5 Implants
Implant Stability Assessed Clinically/Manually
Implant stability - Yes
86 Implants
97 Implants

SECONDARY outcome

Timeframe: At implant insertion.

Population: ITT population. Outcome measure 5 is analyzed on implant level only and does not contain subject level group "Group A + B".

The Osstell® instrument "Osstell® ISQ" will be used to measure RFA. RFA measures the frequency with which a device vibrates. By comparing resonance frequencies the stability of a dental implant can be determined as the resonance frequency changes with different stability. The Osstell® ISQ instrument translates the RFA value into an Implant Stability Quotient (ISQ) scale value. ISQ is a scale from 1 to 100. The ISQ scale has a non-linear correlation to micro mobility, with a higher value corresponding to increased stability.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=86 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=102 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Implant Stability Using Resonance Frequency Analysis (RFA)
Mesio-Distal
78.6 ISQ on implant level
Standard Deviation 6.2
77.5 ISQ on implant level
Standard Deviation 9.6
Implant Stability Using Resonance Frequency Analysis (RFA)
Bucco-Palatinal
78.8 ISQ on implant level
Standard Deviation 5.9
77.7 ISQ on implant level
Standard Deviation 9.4

SECONDARY outcome

Timeframe: After implant insertion

Population: ITT subjects. Outcome measure 6 is analyzed on implant level only and does not contain subject level group "Group A + B".

After implant insertion, a questionnaire will be completed for each study implant by the surgeon. The questionnaire contains following statements "The implant was guided into the prepared osteotomy", "The implant followed the prepared osteotomy" and "The implant has a good primary stability". The surgeon records his/her response to each statement with a number from 1 to 10 , where 1= totally disagree with the statement, and 10= totally agree with the statement.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=86 Implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=102 Implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Perception During Implant Insertion - "The Implant Was Guided Into the Prepared Osteotomy"
7.5 score on a scale
Standard Deviation 2.6
7.1 score on a scale
Standard Deviation 2.4

SECONDARY outcome

Timeframe: After implant insertion

Population: ITT subjects. Outcome measure 7 is analyzed on implant level only and does not contain subject level group "Group A + B".

After implant insertion, a questionnaire will be completed for each study implant by the surgeon. The questionnaire contains following statements "The implant was guided into the prepared osteotomy", "The implant followed the prepared osteotomy" and "The implant has a good primary stability". The surgeon records his/her response to each statement with a number from 1 to 10 , where 1= totally disagree with the statement, and 10= totally agree with the statement.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=86 implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=102 implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Perception During Implant Insertion - "The Implant Followed the Prepared Osteotomy"
8.2 score on a scale
Standard Deviation 1.6
7.6 score on a scale
Standard Deviation 2.0

SECONDARY outcome

Timeframe: After implant insertion

Population: ITT subjects. Outcome measure 8 is analyzed on implant level only and does not contain subject level group "Group A + B".

After implant insertion, a questionnaire will be completed for each study implant by the surgeon. The questionnaire contains following statements "The implant was guided into the prepared osteotomy", "The implant followed the prepared osteotomy" and "The implant has a good primary stability". The surgeon records his/her response to each statement with a number from 1 to 10 , where 1= totally disagree with the statement, and 10= totally agree with the statement.

Outcome measures

Outcome measures
Measure
Standard Drilling Protocol (Group A)
n=86 implants
A standard drilling protocol will be used in bone quality 1 and 2 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Soft Bone Drilling Protocol (Group B)
n=102 implants
A soft bone drilling protocol will be used in bone quality 3 and 4 OsseoSpeed TX: OsseoSpeed TX dental implants, 6-17 mm
Perception During Implant Insertion - "The Implant Has a Good Primary Stability"
8.7 score on a scale
Standard Deviation 1.3
8.1 score on a scale
Standard Deviation 1.7

Adverse Events

Standard Drilling Protocol (Group A)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Soft Bone Drilling Protocol (Group B)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Group A + B

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Manager of Global Clinical Reseach

Dentsply Sirona Implants

Phone: +46313763500

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place