Trial Outcomes & Findings for Three Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction (NCT NCT01225952)

NCT ID: NCT01225952

Last Updated: 2013-01-18

Results Overview

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Recruitment status

COMPLETED

Study phase

PHASE4

Target enrollment

93 participants

Primary outcome timeframe

Postoperative visit 3 (day 30-60)

Results posted on

2013-01-18

Participant Flow

This study was an evaluation of three United States Food and Drug Administration (FDA) approved IOLs. A total of 93 participants(186 eyes) were enrolled at one investigative site in the Philippines. First participant was enrolled 7/15/2010 and last participant exited the study 4/28/2011.

186 participant eyes were enrolled in this study 154 participant eyes completed the study. Participants underwent bilateral phacoemulsification and IOL implantation. Participants were randomized to treatment groups prior to surgery.

Participant milestones

Participant milestones
Measure
Crystalens AO
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Overall Study
STARTED
31
31
31
Overall Study
COMPLETED
25
27
25
Overall Study
NOT COMPLETED
6
4
6

Reasons for withdrawal

Reasons for withdrawal
Measure
Crystalens AO
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Overall Study
Lost to Follow-up
1
0
0
Overall Study
Study Eye Explanted
0
0
1
Overall Study
Withdrawal by Subject
4
4
5
Overall Study
Surgical complications
1
0
0

Baseline Characteristics

Three Intraocular Lenses Designed to Improve Distance, Intermediate and Near Vision Following Lens Extraction

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Crystalens AO
n=31 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=31 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=31 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Total
n=93 Participants
Total of all reporting groups
Age, Customized
< 40 years
0 participants
n=5 Participants
0 participants
n=7 Participants
0 participants
n=5 Participants
0 participants
n=4 Participants
Age, Customized
40 - 49 years
2 participants
n=5 Participants
1 participants
n=7 Participants
0 participants
n=5 Participants
3 participants
n=4 Participants
Age, Customized
50 - 59 years
6 participants
n=5 Participants
8 participants
n=7 Participants
7 participants
n=5 Participants
21 participants
n=4 Participants
Age, Customized
60 - 69 years
13 participants
n=5 Participants
14 participants
n=7 Participants
16 participants
n=5 Participants
43 participants
n=4 Participants
Age, Customized
70 - 79 years
10 participants
n=5 Participants
8 participants
n=7 Participants
7 participants
n=5 Participants
25 participants
n=4 Participants
Age, Customized
>/= 80 years
0 participants
n=5 Participants
0 participants
n=7 Participants
1 participants
n=5 Participants
1 participants
n=4 Participants
Sex: Female, Male
Female
23 Participants
n=5 Participants
25 Participants
n=7 Participants
22 Participants
n=5 Participants
70 Participants
n=4 Participants
Sex: Female, Male
Male
8 Participants
n=5 Participants
6 Participants
n=7 Participants
9 Participants
n=5 Participants
23 Participants
n=4 Participants
Region of Enrollment
Philippines
31 participants
n=5 Participants
31 participants
n=7 Participants
31 participants
n=5 Participants
93 participants
n=4 Participants

PRIMARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=52 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=54 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=50 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree)
1.5 cycles/degree
1.612 log contrast sensitivity units
Standard Deviation 0.229
1.582 log contrast sensitivity units
Standard Deviation 0.231
1.527 log contrast sensitivity units
Standard Deviation 0.248
Monocular Mesopic Contrast Sensitivity Without Glare (1.5, 3 Cycles/Degree)
3 cycles/degree
1.658 log contrast sensitivity units
Standard Deviation 0.195
1.627 log contrast sensitivity units
Standard Deviation 0.224
1.629 log contrast sensitivity units
Standard Deviation 0.240

PRIMARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set, non-missing

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=51 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=50 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=48 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
1.641 log contrast sensitivity units
Standard Deviation 0.210
1.611 log contrast sensitivity units
Standard Deviation 0.237
1.641 log contrast sensitivity units
Standard Deviation 0.286

PRIMARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set, non-missing

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=48 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=43 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=42 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
1.325 log contrast sensitivity units
Standard Deviation 0.226
1.227 log contrast sensitivity units
Standard Deviation 0.198
1.324 log contrast sensitivity units
Standard Deviation 0.281

PRIMARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set, non-missing

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=34 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=26 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=30 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
0.977 log contrast sensitivity units
Standard Deviation 0.266
0.985 log contrast sensitivity units
Standard Deviation 0.185
0.973 log contrast sensitivity units
Standard Deviation 0.342

PRIMARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=50 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=54 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=50 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
1.5 cycles/degree
1.696 log contrast sensitivity units
Standard Deviation 0.213
1.621 log contrast sensitivity units
Standard Deviation 0.222
1.527 log contrast sensitivity units
Standard Deviation 0.190
Monocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
3 cycles/degree
1.678 log contrast sensitivity units
Standard Deviation 0.159
1.670 log contrast sensitivity units
Standard Deviation 0.236
1.630 log contrast sensitivity units
Standard Deviation 0.203

PRIMARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=48 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=51 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=48 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
1.579 log contrast sensitivity units
Standard Deviation 0.239
1.499 log contrast sensitivity units
Standard Deviation 0.230
1.610 log contrast sensitivity units
Standard Deviation 0.228

PRIMARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=39 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=35 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=40 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
1.280 log contrast sensitivity units
Standard Deviation 0.245
1.225 log contrast sensitivity units
Standard Deviation 0.253
1.230 log contrast sensitivity units
Standard Deviation 0.248

PRIMARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

A single eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=25 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=20 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=18 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
0.974 log contrast sensitivity units
Standard Deviation 0.303
0.980 log contrast sensitivity units
Standard Deviation 0.246
0.976 log contrast sensitivity units
Standard Deviation 0.207

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set

In two eye (Binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=26 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=27 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=25 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
1.5 cycles/degree
1.770 log contrast sensitivity units
Standard Deviation 0.187
1.656 log contrast sensitivity units
Standard Deviation 0.238
1.605 log contrast sensitivity units
Standard Deviation 0.203
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
3 cycles/degree
1.807 log contrast sensitivity units
Standard Deviation 0.199
1.679 log contrast sensitivity units
Standard Deviation 0.275
1.689 log contrast sensitivity units
Standard Deviation 0.189

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=25 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=26 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=25 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
1.825 log contrast sensitivity units
Standard Deviation 0.216
1.591 log contrast sensitivity units
Standard Deviation 0.274
1.700 log contrast sensitivity units
Standard Deviation 0.284

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=25 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=21 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=22 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
1.425 log contrast sensitivity units
Standard Deviation 0.221
1.335 log contrast sensitivity units
Standard Deviation 0.218
1.361 log contrast sensitivity units
Standard Deviation 0.278

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=19 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=15 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=16 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
1.029 log contrast sensitivity units
Standard Deviation 0.250
1.001 log contrast sensitivity units
Standard Deviation 0.213
0.981 log contrast sensitivity units
Standard Deviation 0.362

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=25 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=27 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=25 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
1.5 cycles/degree
1.797 log contrast sensitivity units
Standard Deviation 0.183
1.751 log contrast sensitivity units
Standard Deviation 0.238
1.577 log contrast sensitivity units
Standard Deviation 0.206
Binocular Mesopic Contrast Sensitivity Without Glare(1.5, 3 Cycles/Degree)
3 cycles/degree
1.851 log contrast sensitivity units
Standard Deviation 0.178
1.723 log contrast sensitivity units
Standard Deviation 0.239
1.702 log contrast sensitivity units
Standard Deviation 0.219

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=24 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=25 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=24 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(6 Cycles/Degree)
1.758 log contrast sensitivity units
Standard Deviation 0.280
1.640 log contrast sensitivity units
Standard Deviation 0.236
1.690 log contrast sensitivity units
Standard Deviation 0.263

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=21 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=18 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=22 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(12 Cycles/Degree)
1.371 log contrast sensitivity units
Standard Deviation 0.247
1.269 log contrast sensitivity units
Standard Deviation 0.257
1.310 log contrast sensitivity units
Standard Deviation 0.286

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full Analysis Set

In two eyes (binocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=16 Participants
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=12 Participants
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=12 Participants
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Binocular Mesopic Contrast Sensitivity Without Glare(18 Cycles/Degree)
1.115 log contrast sensitivity units
Standard Deviation 0.299
0.995 log contrast sensitivity units
Standard Deviation 0.384
1.041 log contrast sensitivity units
Standard Deviation 0.330

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=52 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=50 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=49 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
1.591 log contrast sensitivity units
Standard Deviation 0.215
1.519 log contrast sensitivity units
Standard Deviation 0.218
1.482 log contrast sensitivity units
Standard Deviation 0.268

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=52 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=50 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=47 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare (3 Cycles/Degree)
1.687 log contrast sensitivity units
Standard Deviation 0.229
1.581 log contrast sensitivity units
Standard Deviation 0.204
1.588 log contrast sensitivity units
Standard Deviation 0.264

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=50 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=49 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=45 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
1.674 log contrast sensitivity units
Standard Deviation 0.237
1.531 log contrast sensitivity units
Standard Deviation 0.238
1.626 log contrast sensitivity units
Standard Deviation 0.293

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=45 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=35 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=34 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
1.328 log contrast sensitivity units
Standard Deviation 0.224
1.230 log contrast sensitivity units
Standard Deviation 0.223
1.340 log contrast sensitivity units
Standard Deviation 0.268

SECONDARY outcome

Timeframe: Postoperative visit 3 (day 30-60)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=27 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=24 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=25 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
1.024 log contrast sensitivity units
Standard Deviation 0.196
0.945 log contrast sensitivity units
Standard Deviation 0.158
0.834 log contrast sensitivity units
Standard Deviation 0.283

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=50 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=50 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=48 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(1.5 Cycles/Degree)
1.550 log contrast sensitivity units
Standard Deviation 0.263
1.488 log contrast sensitivity units
Standard Deviation 0.226
1.449 log contrast sensitivity units
Standard Deviation 0.206

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=50 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=50 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=49 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(3 Cycles/Degree)
1.664 log contrast sensitivity units
Standard Deviation 0.258
1.572 log contrast sensitivity units
Standard Deviation 0.225
1.535 log contrast sensitivity units
Standard Deviation 0.220

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=48 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=46 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=45 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(6 Cycles/Degree)
1.578 log contrast sensitivity units
Standard Deviation 0.252
1.493 log contrast sensitivity units
Standard Deviation 0.267
1.519 log contrast sensitivity units
Standard Deviation 0.260

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=39 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=30 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=31 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(12 Cycles/Degree)
1.335 log contrast sensitivity units
Standard Deviation 0.298
1.246 log contrast sensitivity units
Standard Deviation 0.278
1.194 log contrast sensitivity units
Standard Deviation 0.237

SECONDARY outcome

Timeframe: Postoperative visit 4 (day 120-180)

Population: Full analysis set

In one eye (monocular) in dim light (mesopic) measuring contrast sensitivity; The participant is presented with a sine-wave grating target of a given spatial frequency (cycles/degree of visual angle) where the smaller the number of cycles/degree the wider apart the gradations (vertical lines of grayness). The participant's ability to detect changes in contrast is determined. Higher mean indicates improved contrast sensitivity.

Outcome measures

Outcome measures
Measure
Crystalens AO
n=25 eyes
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=16 eyes
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=16 eyes
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Monocular Mesopic Contrast Sensitivity With Glare(18 Cycles/Degree)
0.919 log contrast sensitivity units
Standard Deviation 0.283
1.019 log contrast sensitivity units
Standard Deviation 0.367
0.935 log contrast sensitivity units
Standard Deviation 0.207

Adverse Events

Crystalens AO

Serious events: 0 serious events
Other events: 12 other events
Deaths: 0 deaths

ReSTOR 3.0

Serious events: 0 serious events
Other events: 5 other events
Deaths: 0 deaths

AMO Tecnis Multifocal

Serious events: 0 serious events
Other events: 6 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Crystalens AO
n=53 participants at risk
Bausch \& Lomb silicone multi-piece accommodating IOL is a modified plate haptic lens
ReSTOR 3.0
n=54 participants at risk
An aspheric multifocal IOL (Alcon Laboratories) combines the functions of an apodized diffractive region and a refractive region.
AMO Tecnis Multifocal
n=51 participants at risk
A foldable hydrophobic acrylic IOL,(Abbott Medical Optics), is an ultraviolet light-absorbing posterior chamber IOL
Eye disorders
Increased Intraocular Pressure
20.8%
11/53 • Number of events 11 • 120-180 days
Adverse events measured by eye
9.3%
5/54 • Number of events 5 • 120-180 days
Adverse events measured by eye
5.9%
3/51 • Number of events 3 • 120-180 days
Adverse events measured by eye
Eye disorders
Cystoid Macular Edema
1.9%
1/53 • Number of events 1 • 120-180 days
Adverse events measured by eye
0.00%
0/54 • 120-180 days
Adverse events measured by eye
5.9%
3/51 • Number of events 3 • 120-180 days
Adverse events measured by eye

Additional Information

Omid Khodai, OD,RAC

Bausch & Lomb Incorporated

Phone: (949) 521-7894

Results disclosure agreements

  • Principal investigator is a sponsor employee All study data generated as a result of this study will be regarded as confidential, until appropriate analysis and review by the Sponsor or its designee and the Investigator(s) are completed. The results of the study may be published or presented by the Investigator(s)after the review by, and in consultation and agreement with the Sponsor.
  • Publication restrictions are in place

Restriction type: OTHER