An Inpatient Performance Evaluation of a New Subcutaneous Glucose Sensor

NCT ID: NCT01112696

Last Updated: 2018-01-16

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

100 participants

Study Classification

INTERVENTIONAL

Study Start Date

2010-04-30

Study Completion Date

2010-11-30

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The purpose of the study is to assess the performance of a new subcutaneous glucose sensor over a seven-day sensor life when used with currently marketed Medtronic Diabetes devices. In addition performance will be calculated for use of the new sensor with proposed new devices using new calibration algorithms.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

The first generation Medtronic MiniMed Subcutaneous Glucose Sensor (Sof-Sensor) was originally approved by the FDA for commercialization as part of the Continuous Glucose Monitoring System (CGMS) on June 15, 1999 (PMA 980022). The Sensor is composed of a microelectrode with a thin coating of glucose oxidase beneath several layers of biocompatible membrane. This same sensor was used as part of subsequent continuous glucose monitoring (CGM) systems, such as the Guardian REAL-Time and Paradigm REAL-Time sensor augmented insulin pump. The current configuration of sensor has undergone in vitro and in vivo testing. A new second-generation glucose sensor (herein referred to as the Comfort Sensor) has been developed. The first-generation glucose sensor was approved with reported sensor accuracy (MAD) of 18%; it was labeled for a maximum use duration of 72 hours, using only the abdomen as an insertion site. The new sensor is shorter and has a smaller diameter, with a smaller gauge introducer needle. The new sensor inserter is designed to be used with the new sensor. The objectives of this study are to 1) Assess performance of the Comfort Sensor when used over a period of seven days with currently available devices, and 2) Assess performance of the Comfort Sensor when used over a period of seven days using new calibration algorithms (post-processed with algorithm for future devices). Accuracy data will be calculated based on comparing calibrated glucose sensor values to a "gold standard" (YSI plasma glucose values) in adult subjects during in-clinic testing, and on comparing calibrated glucose sensor values to glucose meter values during in-home testing. The devices that will be used for gathering sensor data during this study will be: 1) the Comfort Sensor, 2) the MiniLink (Transmitter), 3) the Guardian REAL-Time Display Device and 4) the CGMS iPro.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Type 1 Diabetes Mellitus Type 2 Diabetes Mellitus

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NA

Intervention Model

SINGLE_GROUP

Primary Study Purpose

OTHER

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Sensor

All subjects that wear sensors (all subjects)

Group Type OTHER

Sensor wear

Intervention Type DEVICE

All subjects to wear sensors

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Sensor wear

All subjects to wear sensors

Intervention Type DEVICE

Other Intervention Names

Discover alternative or legacy names that may be used to describe the listed interventions across different sources.

Enlite Sensor

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

1. Males and females aged 18 through 75 inclusive
2. Diagnosed with Type 1 or type 2 Diabetes Mellitus, using insulin to treat their diabetes (Each site's and overall study enrollment of subjects with Type 2 diabetes will target 20% of the total number)
3. Willingness to perform required study and data collection procedures and to adhere to operating requirements of the Guardian REAL-TIME and CGMS iPro Systems
4. Willingness to perform at least 4 capillary blood glucose tests per day while wearing the Guardian REAL-TIME and CGMS iPro Systems
5. Willingness to participate in a 10 hour frequent blood sampling session during the study
6. Subject agrees to comply with the study protocol requirements
7. Informed Consent, HIPAA Authorization, and California Experimental Subject Bill of Rights (if applicable) signed by the subject
8. The Subject is willing to wear both the Guardian REAL-Time and CGMS iPro Systems for 14 days (\~340 hours)

Exclusion Criteria

1. The Subject has a history of tape allergies that have not been resolved
2. The Subject has any skin abnormality (e.g. psoriasis, rash, staphylococcus infection) that has not been resolved and would inhibit them from wearing the sensors
3. Any additional condition(s) that in the Investigator's opinion would warrant exclusion from the study or prevent the Subject from completing the study
4. Subject is currently participating in an investigational study (drug or device)
Minimum Eligible Age

18 Years

Maximum Eligible Age

75 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Medtronic Diabetes

INDUSTRY

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Responsibility Role SPONSOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Scott Lee, MD

Role: STUDY_DIRECTOR

Medtronic Diabetes

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

AMCR Institute, Inc.

Escondido, California, United States

Site Status

Sansum Diabetes Research Institute

Santa Barbara, California, United States

Site Status

Rainier Clinical Research Center, Inc.

Renton, Washington, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

CEP 221

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.

The Monitoring Study
NCT02969863 COMPLETED NA