Trial Outcomes & Findings for Bladder Scan of Residual Urine With New Catheter (NCT NCT01048541)
NCT ID: NCT01048541
Last Updated: 2012-09-03
Results Overview
Residual urine was mesured by ultrasound measurement of bladder content after intermittent catherisation
Recruitment status
COMPLETED
Study phase
PHASE2/PHASE3
Target enrollment
37 participants
Primary outcome timeframe
3 catheterisations on 1 day
Results posted on
2012-09-03
Participant Flow
Participant milestones
| Measure |
AB: First Test Catheter Then Standard Catheter
Test catheter (SpeediCath Compact Male)used in the first period and Standard catheter (SpediCath straight) used in the second period
|
BA: First Standard Catheter Then Test Catheter
Standard catheter (SpediCath straight)used in the first period and test catheter (SpeediCath Compact Male) used in the second period
|
|---|---|---|
|
Overall Study
STARTED
|
18
|
19
|
|
Overall Study
COMPLETED
|
17
|
19
|
|
Overall Study
NOT COMPLETED
|
1
|
0
|
Reasons for withdrawal
| Measure |
AB: First Test Catheter Then Standard Catheter
Test catheter (SpeediCath Compact Male)used in the first period and Standard catheter (SpediCath straight) used in the second period
|
BA: First Standard Catheter Then Test Catheter
Standard catheter (SpediCath straight)used in the first period and test catheter (SpeediCath Compact Male) used in the second period
|
|---|---|---|
|
Overall Study
Withdrawal by Subject
|
1
|
0
|
Baseline Characteristics
Bladder Scan of Residual Urine With New Catheter
Baseline characteristics by cohort
| Measure |
Entire Study
n=37 Participants
Includes groups randomized to standard catheter first and test catheter first
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
36 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
1 Participants
n=5 Participants
|
|
Age Continuous
|
39.5 years
STANDARD_DEVIATION 12.7 • n=5 Participants
|
|
Sex: Female, Male
Female
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
37 Participants
n=5 Participants
|
|
Region of Enrollment
Germany
|
37 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 3 catheterisations on 1 dayResidual urine was mesured by ultrasound measurement of bladder content after intermittent catherisation
Outcome measures
| Measure |
Test Catheter
n=36 Participants
Test catheter (SpeediCath Compact Male)
|
Standard Catheter
n=36 Participants
Standard catheter (SpediCath straight)
|
|---|---|---|
|
Mean Residual Urine Volume
|
12.4 mL
Standard Deviation 15.7
|
9.4 mL
Standard Deviation 11.4
|
SECONDARY outcome
Timeframe: Study periodOutcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: 3 catheterisations on 1 dayOutcome measures
Outcome data not reported
Adverse Events
Test Catheter
Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths
Standard Catheter
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Test Catheter
n=36 participants at risk
Test catheter (SpeediCath Compact Male)
|
Standard Catheter
n=36 participants at risk
Standard catheter (SpediCath straight)
|
|---|---|---|
|
Renal and urinary disorders
Burning in the urethra
|
2.8%
1/36
|
0.00%
0/36
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: LTE60