Trial Outcomes & Findings for Evaluation of the Efficacy of the Bipolar Sealer Aquamantys 6.0 in Patients Undergoing Total Hip Arthroplasty (NCT NCT01027195)

NCT ID: NCT01027195

Last Updated: 2012-11-29

Results Overview

Recruitment status

COMPLETED

Study phase

PHASE4

Target enrollment

140 participants

Primary outcome timeframe

daily during hospital stay (an expected average of 4 days)

Results posted on

2012-11-29

Participant Flow

Participant milestones

Participant milestones
Measure
Bipolar Radiofrequency
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Overall Study
STARTED
71
69
Overall Study
COMPLETED
71
69
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Evaluation of the Efficacy of the Bipolar Sealer Aquamantys 6.0 in Patients Undergoing Total Hip Arthroplasty

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Total
n=140 Participants
Total of all reporting groups
Age Continuous
55.4 years
STANDARD_DEVIATION 10.9 • n=5 Participants
55.7 years
STANDARD_DEVIATION 11.0 • n=7 Participants
55.5 years
STANDARD_DEVIATION 11.0 • n=5 Participants
Sex: Female, Male
Female
35 Participants
n=5 Participants
31 Participants
n=7 Participants
66 Participants
n=5 Participants
Sex: Female, Male
Male
36 Participants
n=5 Participants
38 Participants
n=7 Participants
74 Participants
n=5 Participants
Body-Mass Index
Females
28.6 kg/m2
STANDARD_DEVIATION 7.1 • n=5 Participants
29.2 kg/m2
STANDARD_DEVIATION 6.1 • n=7 Participants
28.9 kg/m2
STANDARD_DEVIATION 6.6 • n=5 Participants
Body-Mass Index
Males
29.4 kg/m2
STANDARD_DEVIATION 4.7 • n=5 Participants
30.4 kg/m2
STANDARD_DEVIATION 5.1 • n=7 Participants
29.9 kg/m2
STANDARD_DEVIATION 4.9 • n=5 Participants
Baseline Hemoglobin
14.6 g/dL
STANDARD_DEVIATION 1.3 • n=5 Participants
14.2 g/dL
STANDARD_DEVIATION 1.2 • n=7 Participants
14.4 g/dL
STANDARD_DEVIATION 1.3 • n=5 Participants

PRIMARY outcome

Timeframe: daily during hospital stay (an expected average of 4 days)

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Number of Patients Managed With Blood Transfusion
15 participants
14 participants

SECONDARY outcome

Timeframe: daily during hospital stay (an expected average of 4 days)

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Number of Units Transfused
0.38 units of blood
Standard Deviation 0.83
0.44 units of blood
Standard Deviation 0.98

SECONDARY outcome

Timeframe: Intraoperative (day of surgery)

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Estimated Blood Loss
315.2 mL
Standard Deviation 203.4
368.5 mL
Standard Deviation 277.8

SECONDARY outcome

Timeframe: within 30 days before surgery (preop), day of hospital discharge

Change in hemoglobin level from baseline to the day of hospital discharge.

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Change in Hemoglobin Level
-5.4 g/dL
Standard Deviation 1.6
-5.3 g/dL
Standard Deviation 1.8

SECONDARY outcome

Timeframe: daily during hospital stay (an expected average of 4 days)

Sum of daily morphine-equivalent mg narcotic usage during hospital stay.

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Total Narcotic Usage (Morphine-equivalent mg) During Hospital Stay
156.1 total morphine-equivalent mg
Standard Deviation 130.2
164.9 total morphine-equivalent mg
Standard Deviation 123.4

SECONDARY outcome

Timeframe: day of hospital discharge

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Length of Stay
3.3 days
Standard Deviation 1.1
3.4 days
Standard Deviation 1.6

SECONDARY outcome

Timeframe: within 30 days before surgery (preop), 4 weeks after surgery, 12 weeks after surgery

A scoring system used to evaluate the outcome after total hip replacement. Domains include pain (44 points), function (47 points), deformity (4 points), and range of motion (5 points). A total score is computed by summing the individual domain scores, with a maximum of 100 points. Higher values represent better outcomes. A total Harris Hip Score below 70 points is generally considered a poor result, 70 to 80 fair, 80 to 90 good, and 90 to 100 excellent.

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Harris Hip Score (Outcome Score)
12 Weeks Postoperative
92.8 points on a scale
Standard Deviation 8.2
95.1 points on a scale
Standard Deviation 5.4
Harris Hip Score (Outcome Score)
Preoperative
59.0 points on a scale
Standard Deviation 12.2
60.4 points on a scale
Standard Deviation 11.6
Harris Hip Score (Outcome Score)
4 Weeks Postoperative
78.4 points on a scale
Standard Deviation 9.8
78.4 points on a scale
Standard Deviation 11.2

SECONDARY outcome

Timeframe: within 30 days before surgery (preop), 4 weeks after surgery, 12 weeks after surgery

A single scoring system used to evaluate overall pain on a scale of integers 0 to 10, with 0 representing "no pain" and 10 representing "unbearable pain." Thus, in this context, lower values represent better outcomes.

Outcome measures

Outcome measures
Measure
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Pain Score Scale
Preoperative
6.2 points on a scale
Standard Deviation 2.2
6.4 points on a scale
Standard Deviation 2.0
Pain Score Scale
4 Weeks Postoperative
1.7 points on a scale
Standard Deviation 1.7
1.7 points on a scale
Standard Deviation 1.8
Pain Score Scale
12 Weeks Postoperative
1.2 points on a scale
Standard Deviation 1.7
0.9 points on a scale
Standard Deviation 1.2

Adverse Events

Bipolar Radiofrequency

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

Standard Bovie Electrocautery

Serious events: 0 serious events
Other events: 4 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Bipolar Radiofrequency
n=71 participants at risk
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
Standard Bovie Electrocautery
n=69 participants at risk
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
Infections and infestations
Deep Infection
0.00%
0/71
1.4%
1/69 • Number of events 1
Cardiac disorders
Coronary Ischemic Event
1.4%
1/71 • Number of events 1
1.4%
1/69 • Number of events 1
Vascular disorders
Deep-Vein Thrombosis
0.00%
0/71
1.4%
1/69 • Number of events 1
Vascular disorders
Pulmonary Embolism
1.4%
1/71 • Number of events 2
1.4%
1/69 • Number of events 1

Additional Information

Wael K Barsoum, MD

The Cleveland Clinic

Phone: 216-444-7515

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place