Trial Outcomes & Findings for Evaluation of the Efficacy of the Bipolar Sealer Aquamantys 6.0 in Patients Undergoing Total Hip Arthroplasty (NCT NCT01027195)
NCT ID: NCT01027195
Last Updated: 2012-11-29
Results Overview
COMPLETED
PHASE4
140 participants
daily during hospital stay (an expected average of 4 days)
2012-11-29
Participant Flow
Participant milestones
| Measure |
Bipolar Radiofrequency
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Overall Study
STARTED
|
71
|
69
|
|
Overall Study
COMPLETED
|
71
|
69
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Evaluation of the Efficacy of the Bipolar Sealer Aquamantys 6.0 in Patients Undergoing Total Hip Arthroplasty
Baseline characteristics by cohort
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
Total
n=140 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age Continuous
|
55.4 years
STANDARD_DEVIATION 10.9 • n=5 Participants
|
55.7 years
STANDARD_DEVIATION 11.0 • n=7 Participants
|
55.5 years
STANDARD_DEVIATION 11.0 • n=5 Participants
|
|
Sex: Female, Male
Female
|
35 Participants
n=5 Participants
|
31 Participants
n=7 Participants
|
66 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
36 Participants
n=5 Participants
|
38 Participants
n=7 Participants
|
74 Participants
n=5 Participants
|
|
Body-Mass Index
Females
|
28.6 kg/m2
STANDARD_DEVIATION 7.1 • n=5 Participants
|
29.2 kg/m2
STANDARD_DEVIATION 6.1 • n=7 Participants
|
28.9 kg/m2
STANDARD_DEVIATION 6.6 • n=5 Participants
|
|
Body-Mass Index
Males
|
29.4 kg/m2
STANDARD_DEVIATION 4.7 • n=5 Participants
|
30.4 kg/m2
STANDARD_DEVIATION 5.1 • n=7 Participants
|
29.9 kg/m2
STANDARD_DEVIATION 4.9 • n=5 Participants
|
|
Baseline Hemoglobin
|
14.6 g/dL
STANDARD_DEVIATION 1.3 • n=5 Participants
|
14.2 g/dL
STANDARD_DEVIATION 1.2 • n=7 Participants
|
14.4 g/dL
STANDARD_DEVIATION 1.3 • n=5 Participants
|
PRIMARY outcome
Timeframe: daily during hospital stay (an expected average of 4 days)Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Number of Patients Managed With Blood Transfusion
|
15 participants
|
14 participants
|
SECONDARY outcome
Timeframe: daily during hospital stay (an expected average of 4 days)Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Number of Units Transfused
|
0.38 units of blood
Standard Deviation 0.83
|
0.44 units of blood
Standard Deviation 0.98
|
SECONDARY outcome
Timeframe: Intraoperative (day of surgery)Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Estimated Blood Loss
|
315.2 mL
Standard Deviation 203.4
|
368.5 mL
Standard Deviation 277.8
|
SECONDARY outcome
Timeframe: within 30 days before surgery (preop), day of hospital dischargeChange in hemoglobin level from baseline to the day of hospital discharge.
Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Change in Hemoglobin Level
|
-5.4 g/dL
Standard Deviation 1.6
|
-5.3 g/dL
Standard Deviation 1.8
|
SECONDARY outcome
Timeframe: daily during hospital stay (an expected average of 4 days)Sum of daily morphine-equivalent mg narcotic usage during hospital stay.
Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Total Narcotic Usage (Morphine-equivalent mg) During Hospital Stay
|
156.1 total morphine-equivalent mg
Standard Deviation 130.2
|
164.9 total morphine-equivalent mg
Standard Deviation 123.4
|
SECONDARY outcome
Timeframe: day of hospital dischargeOutcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Length of Stay
|
3.3 days
Standard Deviation 1.1
|
3.4 days
Standard Deviation 1.6
|
SECONDARY outcome
Timeframe: within 30 days before surgery (preop), 4 weeks after surgery, 12 weeks after surgeryA scoring system used to evaluate the outcome after total hip replacement. Domains include pain (44 points), function (47 points), deformity (4 points), and range of motion (5 points). A total score is computed by summing the individual domain scores, with a maximum of 100 points. Higher values represent better outcomes. A total Harris Hip Score below 70 points is generally considered a poor result, 70 to 80 fair, 80 to 90 good, and 90 to 100 excellent.
Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Harris Hip Score (Outcome Score)
12 Weeks Postoperative
|
92.8 points on a scale
Standard Deviation 8.2
|
95.1 points on a scale
Standard Deviation 5.4
|
|
Harris Hip Score (Outcome Score)
Preoperative
|
59.0 points on a scale
Standard Deviation 12.2
|
60.4 points on a scale
Standard Deviation 11.6
|
|
Harris Hip Score (Outcome Score)
4 Weeks Postoperative
|
78.4 points on a scale
Standard Deviation 9.8
|
78.4 points on a scale
Standard Deviation 11.2
|
SECONDARY outcome
Timeframe: within 30 days before surgery (preop), 4 weeks after surgery, 12 weeks after surgeryA single scoring system used to evaluate overall pain on a scale of integers 0 to 10, with 0 representing "no pain" and 10 representing "unbearable pain." Thus, in this context, lower values represent better outcomes.
Outcome measures
| Measure |
Bipolar Radiofrequency
n=71 Participants
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 Participants
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Pain Score Scale
Preoperative
|
6.2 points on a scale
Standard Deviation 2.2
|
6.4 points on a scale
Standard Deviation 2.0
|
|
Pain Score Scale
4 Weeks Postoperative
|
1.7 points on a scale
Standard Deviation 1.7
|
1.7 points on a scale
Standard Deviation 1.8
|
|
Pain Score Scale
12 Weeks Postoperative
|
1.2 points on a scale
Standard Deviation 1.7
|
0.9 points on a scale
Standard Deviation 1.2
|
Adverse Events
Bipolar Radiofrequency
Standard Bovie Electrocautery
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Bipolar Radiofrequency
n=71 participants at risk
Aquamantys 6.0 bipolar sealer \[Salient Surgical Technologies, Portsmouth, New Hampshire\] used on surgical site during primary total hip arthroplasty to deliver radiofrequency energy coupled with saline solution irrigation for hemostatic sealing (i.e. shrinking of collagen in the walls of tissue vessels) at lower temperatures (\<100 degrees Celsius) than standard Bovie electrocautery.
|
Standard Bovie Electrocautery
n=69 participants at risk
Standard Bovie electrocautery \[Valleylab, Boulder, Colorado\] used on surgical site during primary total hip arthroplasty to deliver high frequency electrical current to seal tissues and blood vessels.
|
|---|---|---|
|
Infections and infestations
Deep Infection
|
0.00%
0/71
|
1.4%
1/69 • Number of events 1
|
|
Cardiac disorders
Coronary Ischemic Event
|
1.4%
1/71 • Number of events 1
|
1.4%
1/69 • Number of events 1
|
|
Vascular disorders
Deep-Vein Thrombosis
|
0.00%
0/71
|
1.4%
1/69 • Number of events 1
|
|
Vascular disorders
Pulmonary Embolism
|
1.4%
1/71 • Number of events 2
|
1.4%
1/69 • Number of events 1
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place