Trial Outcomes & Findings for A Study on M2a Magnum Total Hip Arthroplasty (NCT NCT01010763)

NCT ID: NCT01010763

Last Updated: 2019-06-21

Results Overview

Hip Flexion Angle at 1 year postoperatively

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

184 participants

Primary outcome timeframe

1 year

Results posted on

2019-06-21

Participant Flow

Period of Recruitment was from December 2009 until December 2011. Recruitements were done at medical clinics by providing detailed explanation of procedure as well as study protocol, including randomization \& right of withdrawal.

Four patients withdrer before randomization.

Participant milestones

Participant milestones
Measure
M2a Magnum
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Overall Study
STARTED
96
84
Overall Study
COMPLETED
61
54
Overall Study
NOT COMPLETED
35
30

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Preoperative Harris Hip Score of two patients in Magnum arm were failed to obtain.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
M2a Magnum
n=96 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=84 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Total
n=180 Participants
Total of all reporting groups
Age, Continuous
65 years
n=96 Participants
67 years
n=84 Participants
66 years
n=180 Participants
Sex: Female, Male
Female
72 Participants
n=96 Participants
61 Participants
n=84 Participants
133 Participants
n=180 Participants
Sex: Female, Male
Male
24 Participants
n=96 Participants
23 Participants
n=84 Participants
47 Participants
n=180 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Race (NIH/OMB)
Asian
96 Participants
n=96 Participants
84 Participants
n=84 Participants
180 Participants
n=180 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Race (NIH/OMB)
White
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=96 Participants
0 Participants
n=84 Participants
0 Participants
n=180 Participants
Region of Enrollment
Japan
96 participants
n=96 Participants
84 participants
n=84 Participants
180 participants
n=180 Participants
Harris Hip Score
52.3 units on a scale
n=94 Participants • Preoperative Harris Hip Score of two patients in Magnum arm were failed to obtain.
51.2 units on a scale
n=84 Participants • Preoperative Harris Hip Score of two patients in Magnum arm were failed to obtain.
51.7 units on a scale
n=178 Participants • Preoperative Harris Hip Score of two patients in Magnum arm were failed to obtain.
EQ-5D
0.620 units on a scale
n=95 Participants • Preoperative EQ-5D questionnaire was not completed by two subjects (one subject each arm).
0.647 units on a scale
n=83 Participants • Preoperative EQ-5D questionnaire was not completed by two subjects (one subject each arm).
0.633 units on a scale
n=178 Participants • Preoperative EQ-5D questionnaire was not completed by two subjects (one subject each arm).
UCLA Activity Score
3.1 units on a scale
n=96 Participants
3.0 units on a scale
n=84 Participants
3.0 units on a scale
n=180 Participants

PRIMARY outcome

Timeframe: 1 year

Population: 2 patients in M2a Magnum group and 4 patients in M2a Taper group were failed to obtain Range of Motion data at 1 year follow-up visit.

Hip Flexion Angle at 1 year postoperatively

Outcome measures

Outcome measures
Measure
M2a Magnum
n=94 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=80 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Range of Motion
109.2 degrees
Interval 70.0 to 130.0
110.2 degrees
Interval 70.0 to 130.0

SECONDARY outcome

Timeframe: 3 month

Population: 3 patients in M2a Taper group were failed to obtain Metal Ion data at 3 month follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=96 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=81 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
0.7 microgram/Litter
Interval 0.0 to 2.3
0.9 microgram/Litter
Interval 0.0 to 2.7

SECONDARY outcome

Timeframe: 5 year

Population: 59 patients in M2a Magnum group and 59 patients in M2a Taper group were failed to obtain UCLA Activity Score data at 5 year follow-up visit.

UCLA is physician assessing scoring system to assess patient's activity level. Minimum score is 0 (completely inactive) and maximum score is 9 (Regularly participates in impact sports) with 1 point increments depending on patient's activity level. Higher score means higher activity level achieved.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=37 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=25 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
UCLA Activity Score
4.97 units on a scale
Interval 3.0 to 9.0
4.8 units on a scale
Interval 1.0 to 8.0

SECONDARY outcome

Timeframe: 5 year

Population: 45 patients in M2a Magnum group and 36 patients in M2a Taper group were failed to be assessed radiographically at 5 year follow-up visit.

number of participants, whose x-ray shows any signs of radiographic changes (osteolysis, heterotopic ossification and/or radiolucency)

Outcome measures

Outcome measures
Measure
M2a Magnum
n=51 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=48 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Radiographic Assessment
Acetabular Osteolysis
0 Participants
0 Participants
Radiographic Assessment
Acetabular Radiolucency
0 Participants
0 Participants
Radiographic Assessment
Stem Osteolysis
0 Participants
0 Participants
Radiographic Assessment
Stem Radiolucency
2 Participants
7 Participants
Radiographic Assessment
Heterotopic Ossification
2 Participants
1 Participants
Radiographic Assessment
no abnormalities
47 Participants
40 Participants

SECONDARY outcome

Timeframe: 5 year

Population: 68 patients in M2a Magnum group and 62 patients in M2a Taper group were failed to obtain Harris Hip Score data at 5 year follow-up visit.

Harris Hip Score is physician assessing hip functional scoring system, which includes 4 components (pain, function, deformity and range of motion) with minimum score of 0 \& maximum score of 100. Maximum \& Minimum score of each subscale is pain: 0-44, function: 0-47, Deformity: 0-4 and Range of motion: 0-5. Current report uses summed score of each patient to compare total Harris Hip Score. Higher score means better outcomes.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=28 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=22 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Harris Hip Score
96.3 units on a scale
Interval 72.0 to 100.0
94.5 units on a scale
Interval 64.0 to 100.0

SECONDARY outcome

Timeframe: 6 month

Population: 4 patients in M2a Taper group were failed to obtain Metal Ion data at 6 month follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=96 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=80 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
0.9 microgram/Litter
Interval 0.0 to 5.1
1.2 microgram/Litter
Interval 0.0 to 8.9

SECONDARY outcome

Timeframe: 1 year

Population: 7 patients in M2a Magnum group and 9 patients in M2a Taper group were failed to obtain Metal Ion data at 1 year follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=89 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=75 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
1.2 microgram/Litter
Interval 0.0 to 12.9
1.8 microgram/Litter
Interval 0.0 to 15.4

SECONDARY outcome

Timeframe: 2 year

Population: 12 patients in M2a Magnum group and 10 patients in M2a Taper group were failed to obtain Metal Ion data at 2 year follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=84 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=74 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
1.2 microgram/Litter
Interval 0.13 to 6.3
2.0 microgram/Litter
Interval 0.2 to 11.0

SECONDARY outcome

Timeframe: 3 year

Population: 21 patients in M2a Magnum group and 19 patients in M2a Taper group were failed to obtain Metal Ion data at 3 year follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=75 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=65 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
0.98 microgram/Litter
Interval 0.0 to 4.9
2.9 microgram/Litter
Interval 0.0 to 44.0

SECONDARY outcome

Timeframe: 4 year

Population: 50 patients in M2a Magnum group and 46 patients in M2a Taper group were failed to obtain Metal Ion data at 4 year follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=46 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=38 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
1.09 microgram/Litter
Interval 0.11 to 4.0
3.1 microgram/Litter
Interval 0.0 to 22.0

SECONDARY outcome

Timeframe: 5 year

Population: 33 patients in M2a Magnum group and 30 patients in M2a Taper group were failed to obtain Metal Ion data at 5 year follow-up visit.

Blood Cobalt Ion Concentration 0 was imputed for the measurement result being below detection limit.

Outcome measures

Outcome measures
Measure
M2a Magnum
n=63 Participants
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=54 Participants
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Metal Ion
1.07 microgram/Litter
Interval 0.11 to 7.6
3.8 microgram/Litter
Interval 0.16 to 71.0

Adverse Events

M2a Magnum

Serious events: 1 serious events
Other events: 2 other events
Deaths: 0 deaths

M2a Taper

Serious events: 1 serious events
Other events: 3 other events
Deaths: 3 deaths

Serious adverse events

Serious adverse events
Measure
M2a Magnum
n=96 participants at risk
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=84 participants at risk
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Musculoskeletal and connective tissue disorders
Revision due to Adverse Reaction to Metal Debris
1.0%
1/96 • Number of events 1 • 5 years
0.00%
0/84 • 5 years
Musculoskeletal and connective tissue disorders
Femoral Epicondyle Fracture
0.00%
0/96 • 5 years
1.2%
1/84 • Number of events 1 • 5 years

Other adverse events

Other adverse events
Measure
M2a Magnum
n=96 participants at risk
Total HIp Arthroplasty using with the M2a Magnum Large Metal Articulation is an ultra-high performance metal-on-metal articulation with a big ball (greater than or equal to 38mm) in acetabulums as small as 44mm. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
M2a Taper
n=84 participants at risk
Total Hip Arthroplasty using with the M2a Taper Acetabular System consists of a titanium outer shell with cobalt chromium (Co-Cr-Mo) metallic liner, which articulates with with a cobalt chromium (Co-Cr-Mo) modular femoral head. Total Hip Arthroplasty: Degenerated hip is replaced with implantable devices, which include femoral stem, acetabular cup, femoral head and acetabular liner(control group only).
Musculoskeletal and connective tissue disorders
Dislocation
0.00%
0/96 • 5 years
3.6%
3/84 • Number of events 3 • 5 years
Vascular disorders
Cerebral Infarction
1.0%
1/96 • Number of events 1 • 5 years
0.00%
0/84 • 5 years
Surgical and medical procedures
Intra-operative fracture
1.0%
1/96 • Number of events 1 • 5 years
0.00%
0/84 • 5 years

Additional Information

Manager of Clinical Affairs

Zimmer Biomet G.K.

Phone: +81-3-6402-6610

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place