Trial Outcomes & Findings for Therapy Cool Path Ablate VT (NCT NCT00925522)

NCT ID: NCT00925522

Last Updated: 2019-04-17

Results Overview

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

20 participants

Primary outcome timeframe

7 days

Results posted on

2019-04-17

Participant Flow

Patients with atrial flutter are evaluated by ablation doctor in the EP clinic and assessed for possibility of meeting other inclusion criteria.

This is a non-randomized study.

Participant milestones

Participant milestones
Measure
Therapy Cool Path Duo Cardiac Ablation System
Therapy Cool Path Duo Cardiac Ablation System: Consists of following three system components: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids to patient's vascular system through the cat
Overall Study
STARTED
20
Overall Study
COMPLETED
0
Overall Study
NOT COMPLETED
20

Reasons for withdrawal

Reasons for withdrawal
Measure
Therapy Cool Path Duo Cardiac Ablation System
Therapy Cool Path Duo Cardiac Ablation System: Consists of following three system components: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids to patient's vascular system through the cat
Overall Study
study closed -- business decision
20

Baseline Characteristics

Therapy Cool Path Ablate VT

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Therapy Cool Path Duo Cardiac Ablation System
n=20 Participants
Therapy Cool Path Duo Cardiac Ablation System: Consists three system components: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
15 Participants
n=5 Participants
Age, Categorical
>=65 years
5 Participants
n=5 Participants
Age, Continuous
66.8 years
STANDARD_DEVIATION 10.5 • n=5 Participants
Sex: Female, Male
Female
0 Participants
n=5 Participants
Sex: Female, Male
Male
20 Participants
n=5 Participants
Region of Enrollment
United States
20 participants
n=5 Participants

PRIMARY outcome

Timeframe: 7 days

Outcome measures

Outcome measures
Measure
Therapy Cool Path Duo Catheter
n=20 Participants
All patients received RF (radio frequency) therapy from ablation catheter.
Primary Safety is Defined as the Incidence of Intra-procedural, Acute or Sub-chronic, Serious Cardiac Adverse Events, up to 7 Days Post-procedure.
7 participants

Adverse Events

Therapy Cool Path Duo Cardiac Ablation System

Serious events: 7 serious events
Other events: 16 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Therapy Cool Path Duo Cardiac Ablation System
n=20 participants at risk
Therapy Cool Path Duo Cardiac Ablation System: Consists of: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids to patient's vascular system through the catheter in the hospital environment.
Cardiac disorders
Arrhythmia(s) including AFL, AF or VT
25.0%
5/20 • Number of events 5 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Congestive Heart Failure
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Hypotension
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Cardiac Effusion without Tamponade
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Cardiac Effusion with Tamponade
10.0%
2/20 • Number of events 2 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used

Other adverse events

Other adverse events
Measure
Therapy Cool Path Duo Cardiac Ablation System
n=20 participants at risk
Therapy Cool Path Duo Cardiac Ablation System: Consists of: A flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. The 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode. The Cool Point ™ Irrigation Pump and Cool Point™ Tubing Set is an external volumetric pump that is intended for use in administration of general I.V. fluids to patient's vascular system through the catheter in the hospital environment.
Cardiac disorders
Hypotension
15.0%
3/20 • Number of events 3 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Hypertension
15.0%
3/20 • Number of events 3 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Arrythmia
10.0%
2/20 • Number of events 2 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Blood and lymphatic system disorders
Infection
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
Chest Pains Palpitations
10.0%
2/20 • Number of events 3 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Psychiatric disorders
Anxiety/Depression
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Cardiac disorders
CHF symptoms
10.0%
2/20 • Number of events 3 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Nervous system disorders
Sleep Apnea
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used
Blood and lymphatic system disorders
Hyperlipidemia
5.0%
1/20 • Number of events 1 • 6 months
Definition of AEs and SAEs per clinicaltrials.gov were used

Additional Information