Trial Outcomes & Findings for Effect of Cyclic Loading (Vibration) on Orthodontic Tooth Movement (NCT NCT00830947)

NCT ID: NCT00830947

Last Updated: 2013-12-23

Results Overview

Recruitment status

COMPLETED

Study phase

PHASE3

Target enrollment

45 participants

Primary outcome timeframe

Time to Space Closure, an average of 22 weeks

Results posted on

2013-12-23

Participant Flow

Participant milestones

Participant milestones
Measure
OrthoAccel Device
OrthoAccel Device : The OA device provides a light vibration at 0.2 Newtons and 30 Hz. frequency for 20 minutes daily.
Sham (Inactive Device)
Sham Device : Inactive sham device that is held in the mouth for 20 minutes.
Overall Study
STARTED
23
22
Overall Study
COMPLETED
19
18
Overall Study
NOT COMPLETED
4
4

Reasons for withdrawal

Reasons for withdrawal
Measure
OrthoAccel Device
OrthoAccel Device : The OA device provides a light vibration at 0.2 Newtons and 30 Hz. frequency for 20 minutes daily.
Sham (Inactive Device)
Sham Device : Inactive sham device that is held in the mouth for 20 minutes.
Overall Study
Lost to Follow-up
4
4

Baseline Characteristics

Effect of Cyclic Loading (Vibration) on Orthodontic Tooth Movement

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
OrthoAccel Device
n=23 Participants
OrthoAccel Device : The OA device provides a light vibration at 0.2 Newtons and 30 Hz. frequency for 20 minutes daily.
Sham (Inactive Device)
n=22 Participants
Sham Device : Inactive sham device that is held in the mouth for 20 minutes.
Total
n=45 Participants
Total of all reporting groups
Age Continuous
21.8 years
STANDARD_DEVIATION 8.7 • n=5 Participants
21.4 years
STANDARD_DEVIATION 7.8 • n=7 Participants
21.6 years
STANDARD_DEVIATION 8.2 • n=5 Participants
Sex: Female, Male
Female
17 Participants
n=5 Participants
11 Participants
n=7 Participants
28 Participants
n=5 Participants
Sex: Female, Male
Male
6 Participants
n=5 Participants
11 Participants
n=7 Participants
17 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Time to Space Closure, an average of 22 weeks

Outcome measures

Outcome measures
Measure
OrthoAccel Device
n=23 Participants
OrthoAccel Device : The OA device provides a light vibration at 0.2 Newtons and 30 Hz. frequency for 20 minutes daily.
Sham (Inactive Device)
n=22 Participants
Sham Device : Inactive sham device that is held in the mouth for 20 minutes.
The Rate of Orthodontic Movement of a Maxillary Canine Tooth Being Distalized to Close an Extraction Space.
0.27 mm/week
Standard Deviation 0.04
0.18 mm/week
Standard Deviation 0.04

Adverse Events

OrthoAccel Device

Serious events: 0 serious events
Other events: 5 other events
Deaths: 0 deaths

Sham (Inactive Device)

Serious events: 0 serious events
Other events: 5 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
OrthoAccel Device
n=23 participants at risk
OrthoAccel Device : The OA device provides a light vibration at 0.2 Newtons and 30 Hz. frequency for 20 minutes daily.
Sham (Inactive Device)
n=22 participants at risk
Sham Device : Inactive sham device that is held in the mouth for 20 minutes.
Gastrointestinal disorders
Hypoaestesia oral
0.00%
0/23
4.5%
1/22
Gastrointestinal disorders
Nausea
4.3%
1/23
0.00%
0/22
Gastrointestinal disorders
Retching
0.00%
0/23
4.5%
1/22
Gastrointestinal disorders
Toothache
8.7%
2/23
0.00%
0/22
General disorders
Malaise
4.3%
1/23
0.00%
0/22
General disorders
Sensation of pressure
4.3%
1/23
0.00%
0/22
Injury, poisoning and procedural complications
Medical device discomfort
0.00%
0/23
4.5%
1/22
Injury, poisoning and procedural complications
Medical device pain
8.7%
2/23
4.5%
1/22
Injury, poisoning and procedural complications
TAD Failure
4.3%
1/23
4.5%
1/22
Musculoskeletal and connective tissue disorders
Pain in jaw
0.00%
0/23
4.5%
1/22
Nervous system disorders
Headache
4.3%
1/23
9.1%
2/22
Nervous system disorders
Hypoaesthesia
4.3%
1/23
0.00%
0/22
Respiratory, thoracic and mediastinal disorders
Sneezing
4.3%
1/23
0.00%
0/22

Additional Information

Zaffer Syed, Director of Clinical & Regulatory Affairs

OrthoAccel Technologies, Inc.

Phone: (832) 803-0339

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: LTE60