Trial Outcomes & Findings for Trilogy Comparison Study - Adults (NCT NCT00810498)
NCT ID: NCT00810498
Last Updated: 2021-02-03
Results Overview
Blood gas was obtained for this study one hour after use of each device.
COMPLETED
NA
11 participants
After one hour of ventilator use
2021-02-03
Participant Flow
Participant milestones
| Measure |
Trilogy First, Then Standard of Care
These participants used Trilogy first, then Standard of Care.
|
Standard of Care First, Then Trilogy
these participants used Standard of Care first, then Trilogy.
|
|---|---|---|
|
Overall Study
STARTED
|
6
|
5
|
|
Overall Study
COMPLETED
|
6
|
5
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Trilogy Comparison Study - Adults
Baseline characteristics by cohort
| Measure |
All Participants Consented
n=11 Participants
All participants that were consented.
|
|---|---|
|
Age, Continuous
|
57.8 years
STANDARD_DEVIATION 8.82 • n=5 Participants
|
|
Sex: Female, Male
Female
|
5 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
6 Participants
n=5 Participants
|
|
Region of Enrollment
Canada
|
11 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: After one hour of ventilator usePopulation: Only 10 out of 11 total participants' data were used for blood gases due to one participant's venous line occluding.
Blood gas was obtained for this study one hour after use of each device.
Outcome measures
| Measure |
Trilogy
n=10 Participants
Trilogy Device
Trilogy: Exposure for one hour on the Trilogy ventilator
|
Standard of Care
n=10 Participants
Participants prescribed ventilator
Standard of Care: Exposure to participants current ventilator
|
|---|---|---|
|
Comparable Level of Gas Exchange (Measured as Partial Pressure of CO2 in Arterial or Venous Blood).
|
42 mm Hg
Standard Deviation 14
|
43 mm Hg
Standard Deviation 14
|
PRIMARY outcome
Timeframe: After one hour of ventilator usePopulation: Only 10 out of the 11 total participants' data were used for SpO2 due to missing data. Only 8 out of the 11 total participants data were used for SpO2 Standard of care analysis due to missing data.
Pulse Oximetry was measured after one hour of use of each ventilator.
Outcome measures
| Measure |
Trilogy
n=10 Participants
Trilogy Device
Trilogy: Exposure for one hour on the Trilogy ventilator
|
Standard of Care
n=8 Participants
Participants prescribed ventilator
Standard of Care: Exposure to participants current ventilator
|
|---|---|---|
|
Comparable Level of Gas Exchange SpO2 Levels During Various Modes of Ventilation.
|
98 mm Hg
Standard Deviation 2
|
97 mm Hg
Standard Deviation 2
|
SECONDARY outcome
Timeframe: One hour during ventilator use.Population: Missing one data set for Standard of care.
Tidal volume is the amount of air that moves in or out of the lungs with each respiratory cycle. It measures around 500 mL in an average healthy adult male and approximately 400 mL in a healthy female. Average tidal volume was compare after a hour of use of each device.
Outcome measures
| Measure |
Trilogy
n=11 Participants
Trilogy Device
Trilogy: Exposure for one hour on the Trilogy ventilator
|
Standard of Care
n=10 Participants
Participants prescribed ventilator
Standard of Care: Exposure to participants current ventilator
|
|---|---|---|
|
Tidal Volume
|
479 mL/kg
Standard Deviation 182
|
498 mL/kg
Standard Deviation 128
|
SECONDARY outcome
Timeframe: One hour during ventilator use.Population: Missing one data set for Standard of Care.
Minute ventilation is the volume of gas inhaled or exhaled from a person's lungs per minute. This was measured after one hour of use of each device.
Outcome measures
| Measure |
Trilogy
n=11 Participants
Trilogy Device
Trilogy: Exposure for one hour on the Trilogy ventilator
|
Standard of Care
n=10 Participants
Participants prescribed ventilator
Standard of Care: Exposure to participants current ventilator
|
|---|---|---|
|
Minute Ventilation
|
10.11 L/m
Standard Deviation 1.94
|
10.11 L/m
Standard Deviation 2.12
|
SECONDARY outcome
Timeframe: One hour during ventilator use.Number of beats per minute. Measured after one hour of use of each device.
Outcome measures
| Measure |
Trilogy
n=11 Participants
Trilogy Device
Trilogy: Exposure for one hour on the Trilogy ventilator
|
Standard of Care
n=11 Participants
Participants prescribed ventilator
Standard of Care: Exposure to participants current ventilator
|
|---|---|---|
|
Hemodynamics (Heart Rate)
|
89.9 beats per minute
Standard Deviation 19.0
|
90.4 beats per minute
Standard Deviation 19.7
|
Adverse Events
Trilogy Device
Standard of Care
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place