Trial Outcomes & Findings for Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism. (NCT NCT00769808)

NCT ID: NCT00769808

Last Updated: 2020-09-11

Results Overview

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

72 participants

Primary outcome timeframe

3 months

Results posted on

2020-09-11

Participant Flow

Seventy two myopic participants (144 eyes) were enrolled.

Participant milestones

Participant milestones
Measure
Technolas 217z Excimer Laser
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Overall Study
STARTED
72
Overall Study
COMPLETED
64
Overall Study
NOT COMPLETED
8

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Technolas 217z Laser With Zyoptix Aspheric Algorithm Used for the LASIK Treatment of Myopia and Myopic Astigmatism.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Technolas 217z Excimer Laser
n=65 Participants
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
65 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Age, Continuous
32.0 years
STANDARD_DEVIATION 7.1 • n=5 Participants
Sex: Female, Male
Female
29 Participants
n=5 Participants
Sex: Female, Male
Male
36 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
21 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
2 Participants
n=5 Participants
Race (NIH/OMB)
White
40 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
2 Participants
n=5 Participants

PRIMARY outcome

Timeframe: 3 months

Population: There were 59 participants with visual acuity data available for analysis.

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=59 Participants
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage of Eyes With Uncorrected High Contrast Distance Visual Acuity of 20/40 or Better
99.2 percentage of eyes
Interval 97.5 to 100.0

PRIMARY outcome

Timeframe: 3 months

Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Within 0.50 D
109 eyes
Percentage of Eyes With Manifest Subjective Refraction (MRSE) Within ± 1.00 Diopter (D) and 0.50 D of Emmetropia
Within 1.00 D
117 eyes

PRIMARY outcome

Timeframe: Baseline, 3 months

Mean change in best-corrected high contrast distance visual acuity. Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Preservation of Manifest Best Corrected High Contrast Distance Visual Acuity
-0.055 units on a scale
Standard Deviation 0.103

PRIMARY outcome

Timeframe: 3 months

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage of Eyes With Manifest Best-corrected Distance Photopic Visual Acuity Decrease of 2 or More Lines
1 eyes

PRIMARY outcome

Timeframe: 3 months

Manifest refractions (eyeglass prescriptions) have three components: sphere, cylinder, and axis. Sphere and cylinder are both expressed in diopters, which is a continuous measure of the focusing power needed to make the image sharp on the retina. Manifest refraction spherical equivalent (MRSE) is equal to sphere + ½ x cylinder. It is expressed in diopters (D).

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage of Eyes With > 2.00 Diopter of Induced Postoperative Refractive Cylinder
0 eyes

PRIMARY outcome

Timeframe: 3 months

Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity. Acuity was averaged across visits.

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage of Eyes That Had High Contrast Best Spectacle Corrected Visual Acuity (HCBSCVA) of 20/20 or Better at Baseline Had a Visual Outcome Worse Than 20/40 at 3 Months
0 eyes

SECONDARY outcome

Timeframe: Baseline, 3 months

The aberrations in the eye's optical system were measured over the entire pupil by the Bausch \& Lomb Zywave II WaveFront Aberrometer, a Hartmann-Shack type device. The pupil should have been at least 6 mm in diameter using 2.5% phenylephrine as a dilating agent to maximize the pupil diameter and hence the amount of wavefront information collected by the Zywave II. The wavefront data was then fitted with Zernike polynomials representing the 2nd, 3rd, 4th and 5th order Zernike terms. These terms represent the wavefront aberration of the eye.

Outcome measures

Outcome measures
Measure
Technolas 217z Excimer Laser
n=118 eyes
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Percentage Change From Baseline in Higher Order Aberrations.
24.9 Percentage change
Standard Deviation 0.1413

Adverse Events

Technolas 217z Excimer Laser

Serious events: 0 serious events
Other events: 51 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Technolas 217z Excimer Laser
n=66 participants at risk
Bausch \& Lomb Zyoptix Aspheric Algorithm for LASIK correction of myopia and myopic astigmatism. Bausch \& Lomb 217z Excimer Laser: LASIK correction of myopia and myopic astigmatism
Eye disorders
Bowman's membrane disorder
12.1%
8/66 • 6 months
There were 66 treated participants evaluated for adverse events.
Eye disorders
Corneal edema
6.1%
4/66 • 6 months
There were 66 treated participants evaluated for adverse events.
Eye disorders
Keratitis
24.2%
16/66 • 6 months
There were 66 treated participants evaluated for adverse events.
Eye disorders
Punctate keratitis
54.5%
36/66 • 6 months
There were 66 treated participants evaluated for adverse events.
Eye disorders
Eye operation complication
31.8%
21/66 • 6 months
There were 66 treated participants evaluated for adverse events.

Additional Information

Study Director

Bausch & Lomb Incorporated

Phone: 9083009926

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: OTHER