Trial Outcomes & Findings for Titan® One Touch Release Inflatable Penile Prosthesis (NCT NCT00731939)

NCT ID: NCT00731939

Last Updated: 2014-04-14

Results Overview

The study's primary endpoint was to demonstrate that at least 64% of subjects were mostly or completely satisfied with the ability to deflate the device at the 6-month follow-up. The study's primary objective was to assess the ease of deflation of the Titan® OTR pump via subject questionnaire at 6-month follow-up. Subjects were asked via questionnaire how satisfied they were with ease of deflation of their implant. Success criteria was a response of "satisfactory" or "somewhat satisfactory". Other possible responses were "neither satisfactory nor unsatisfactory", "somewhat unsatisfactory", and "very unsatisfactory".

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

124 participants

Primary outcome timeframe

6 months

Results posted on

2014-04-14

Participant Flow

Participant milestones

Participant milestones
Measure
Titan® OTR IPP
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Overall Study
STARTED
124
Overall Study
COMPLETED
90
Overall Study
NOT COMPLETED
34

Reasons for withdrawal

Reasons for withdrawal
Measure
Titan® OTR IPP
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Overall Study
Adverse Event
6
Overall Study
Lost to Follow-up
15
Overall Study
Not implanted with study device
9
Overall Study
Withdrawal by Subject
2
Overall Study
Did not meet enrollment criteria
2

Baseline Characteristics

Titan® One Touch Release Inflatable Penile Prosthesis

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Titan® OTR IPP
n=124 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
76 Participants
n=5 Participants
Age, Categorical
>=65 years
48 Participants
n=5 Participants
Age, Continuous
61.0 years
STANDARD_DEVIATION 9.1 • n=5 Participants
Sex: Female, Male
Female
0 Participants
n=5 Participants
Sex: Female, Male
Male
124 Participants
n=5 Participants
Region of Enrollment
United States
71 participants
n=5 Participants
Region of Enrollment
Canada
27 participants
n=5 Participants
Region of Enrollment
United Kingdom
26 participants
n=5 Participants

PRIMARY outcome

Timeframe: 6 months

The study's primary endpoint was to demonstrate that at least 64% of subjects were mostly or completely satisfied with the ability to deflate the device at the 6-month follow-up. The study's primary objective was to assess the ease of deflation of the Titan® OTR pump via subject questionnaire at 6-month follow-up. Subjects were asked via questionnaire how satisfied they were with ease of deflation of their implant. Success criteria was a response of "satisfactory" or "somewhat satisfactory". Other possible responses were "neither satisfactory nor unsatisfactory", "somewhat unsatisfactory", and "very unsatisfactory".

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Deflation of the Titan® OTR Pump
70.8 % of participants meeting success criter

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - overall function

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 1
72.4 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - overall function

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 1
90.6 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - overall function

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 1
90.0 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - soft enough to conceal when deflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 2
62.8 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - soft enough to conceal when deflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 2
76.0 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - soft enough to conceal when deflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 2
82.2 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - ease of locating the deflation touch pads

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 3
60.6 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - ease of locating the deflation touch pads

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 3
78.2 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - ease of locating the deflation touch pads

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=89 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 3
77.6 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - ease of inflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 4
68.1 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - ease of inflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=95 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 4
85.3 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - ease of inflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=88 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 4
86.4 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - ease of deflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 5
60.6 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - ease of deflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 5
70.8 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - ease of deflation

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 5
73.3 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - hardness of erection when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 6
82.0 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - hardness of erection when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 6
91.5 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - hardness of erection when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=89 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 6
93.2 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - width when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 7
77.6 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - width when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 7
86.1 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - width when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=89 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 7
79.8 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - length when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 8
54.3 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - length when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=95 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 8
61.1 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - length when inflated

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 8
68.9 % satisfactory or somewhat satisfactory

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 9
79.8 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 9
86.5 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - would you recommend this penile implant device to men with the same erectile difficulty that you had?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 9
87.8 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: 3 months post-surgery

Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=94 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 10
72.4 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: 6 months post-surgery

Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=96 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 10
83.3 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: 12 months post-surgery

Subject Satisfaction - if you had the decision to make again, would you undergo this penile implant procedure again?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=90 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Evaluate User Acceptance of Titan® OTR - Question 10
86.7 % responding "yes" and "probably"

SECONDARY outcome

Timeframe: At implant

Titan OTR pre-implant product preparation was straightforward/simple?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=113 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Implant of the Titan® OTR Question 1
Completely disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR Question 1
Somewhat disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR Question 1
Neutral
2.7 percentage of responses
Assess the Ease of Implant of the Titan® OTR Question 1
Somewhat agree
22.1 percentage of responses
Assess the Ease of Implant of the Titan® OTR Question 1
Completely agree
75.2 percentage of responses

SECONDARY outcome

Timeframe: At implant

Titan OTR pre-implant product preparation was easier than your usual pump of choice?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=113 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Implant of the Titan® OTR - Question 2
Completely disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 2
Somewhat disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 2
Neutral
10.6 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 2
Somewhat agree
40.7 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 2
Completely agree
48.7 percentage of responses

SECONDARY outcome

Timeframe: At implant

The subject was easily able to accommodate the OTR pump?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=113 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Implant of the Titan® OTR - Question 3
Completely disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 3
Somewhat disagree
0.0 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 3
Neutral
2.7 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 3
Somewhat agree
20.4 percentage of responses
Assess the Ease of Implant of the Titan® OTR - Question 3
Completely agree
77.0 percentage of responses

SECONDARY outcome

Timeframe: 6 weeks

It was easy for the subject to find the inflation bulb?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 1
Not at all
0.0 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 1
A little
0.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 1
Somewhat
4.6 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 1
Very
13.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 1
Extremely
80.6 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

It was easy for the subject to find the deflation touch pads?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 2
Not at all
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 2
A little
3.7 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 2
Somewhat
9.3 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 2
Very
26.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 2
Extremely
58.3 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

It was easy for the subject to inflate the device?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 3
Not at all
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 3
A little
0.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 3
Somewhat
4.6 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 3
Very
17.6 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 3
Extremely
75.0 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

It was easy for the subject to compress the deflation touch pads?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 4
Not at all
2.8 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 4
A little
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 4
Somewhat
10.2 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 4
Very
25.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 4
Extremely
59.3 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

Subject training with OTR pump was easier than with previous pump?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 5
Not at all
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 5
A little
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 5
Somewhat
16.7 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 5
Very
16.7 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 5
Extremely
63.0 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

The OTR pump was easy to use at 1st cycling?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 6
Not at all
1.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 6
A little
5.6 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 6
Somewhat
8.3 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 6
Very
21.3 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 6
Extremely
63.0 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

The subject likes the OTR pump?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=108 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 7
Not at all
0.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 7
A little
3.7 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 7
Somewhat
13.0 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 7
Very
20.4 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 7
Extremely
62.0 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

How easy was it for the subject to learn?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=107 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 8
Somewhat
7.5 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 8
Not at all
2.8 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 8
A little
8.4 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 8
Very
29.9 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 8
Extremely
51.4 percentage of participants

SECONDARY outcome

Timeframe: 6 weeks

How likely the subject will need continued training or retraining?

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=107 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess the Ease of Training for Titan® OTR - Question 9
Not at all
5.6 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 9
A little
10.3 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 9
Somewhat
11.2 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 9
Very
25.2 percentage of participants
Assess the Ease of Training for Titan® OTR - Question 9
Extremely
47.7 percentage of participants

SECONDARY outcome

Timeframe: Baseline

The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline.

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=59 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess Partner Satisfaction (Where Applicable)
Ease of erection
22.9 units on a scale
Standard Deviation 28.3
Assess Partner Satisfaction (Where Applicable)
Erectile function
18.1 units on a scale
Standard Deviation 23.9
Assess Partner Satisfaction (Where Applicable)
Pleasure from sexual activity
35.6 units on a scale
Standard Deviation 29.8
Assess Partner Satisfaction (Where Applicable)
Satisfaction with orgasm
35.2 units on a scale
Standard Deviation 33.2
Assess Partner Satisfaction (Where Applicable)
Confidence to complete sexual activity
23.2 units on a scale
Standard Deviation 27.0

SECONDARY outcome

Timeframe: 6 months

The Partner Treatment Satisfaction Scale (TSS) was used. The TSS provides scores ranging from 0 to 100 in 5 domains of Ease of erection, Erectile function, Pleasure from sexual activity, Satisfaction with orgasm and Confidence to complete sexual activity with higher scores indicative of worse symptoms. Available data is summarized for each domain at each visit along with change from baseline.

Outcome measures

Outcome measures
Measure
Titan® OTR IPP
n=39 Participants
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Assess Partner Satisfaction (Where Applicable)
Ease of erection
67.8 units on a scale
Standard Deviation 31.3
Assess Partner Satisfaction (Where Applicable)
Erectile function
74.8 units on a scale
Standard Deviation 20.4
Assess Partner Satisfaction (Where Applicable)
Pleasure from sexual activity
64.7 units on a scale
Standard Deviation 24.8
Assess Partner Satisfaction (Where Applicable)
Satisfaction with orgasm
62.2 units on a scale
Standard Deviation 29.2
Assess Partner Satisfaction (Where Applicable)
Confidence to complete sexual activity
71.2 units on a scale
Standard Deviation 28.4

Adverse Events

Titan® OTR IPP

Serious events: 10 serious events
Other events: 14 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Titan® OTR IPP
n=113 participants at risk
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Surgical and medical procedures
Infection
3.5%
4/113 • Number of events 4 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.
Surgical and medical procedures
Hematoma
1.8%
2/113 • Number of events 2 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.
Surgical and medical procedures
Device malfunction
1.8%
2/113 • Number of events 2 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.
Surgical and medical procedures
Chronic pain
0.88%
1/113 • Number of events 1 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.
Surgical and medical procedures
Cylinder extrusion
0.88%
1/113 • Number of events 1 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.
Surgical and medical procedures
Scrotal abscess
0.88%
1/113 • Number of events 1 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.

Other adverse events

Other adverse events
Measure
Titan® OTR IPP
n=113 participants at risk
Subjects implanted with Titan® One Touch Release (OTR) Inflatable Penile Prosthesis (IPP) Inflatable Penile Prosthesis : Hydraulic system designed to be surgically implanted into the penis for the management of erectile dysfunction.
Surgical and medical procedures
Auto inflation
12.4%
14/113 • Number of events 14 • 12 months
The analysis is based on the per protocol population (n=113). Of 124 consented subjects, nine (9) subjects did not meet inclusion/exclusion criteria and were not implanted with the study device. Two (2) subjects did not meet enrollment criteria and these two subjects were not followed for safety or efficacy.

Additional Information

Clinical Trial Manager

Coloplast Corp

Phone: 612-302-4990

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place