Trial Outcomes & Findings for Ease of Use and Blood Clotting for the Gambro Polyflux HD-C4 Big Dialyzer (NCT NCT00659724)

NCT ID: NCT00659724

Last Updated: 2025-07-24

Results Overview

For every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.

Recruitment status

COMPLETED

Target enrollment

33 participants

Primary outcome timeframe

Priming at each treatment

Results posted on

2025-07-24

Participant Flow

Participant milestones

Participant milestones
Measure
Hemodialysis Patients
Hemodialysis patients 18 years of age or older receiving routine dialysis treatment for chronic renal failure.
Weeks 1-2: Assessment of 180NR or 200NR
STARTED
33
Weeks 1-2: Assessment of 180NR or 200NR
COMPLETED
33
Weeks 1-2: Assessment of 180NR or 200NR
NOT COMPLETED
0
Weeks 3-4: Assessment of HD-C4 Big
STARTED
33
Weeks 3-4: Assessment of HD-C4 Big
COMPLETED
33
Weeks 3-4: Assessment of HD-C4 Big
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Ease of Use and Blood Clotting for the Gambro Polyflux HD-C4 Big Dialyzer

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Hemodialysis Patients
n=33 Participants
Hemodialysis patients 18 years of age or older receiving routine dialysis treatment for chronic renal failure.
Age, Continuous
66.0 years
STANDARD_DEVIATION 11.1 • n=5 Participants
Sex: Female, Male
Female
18 Participants
n=5 Participants
Sex: Female, Male
Male
15 Participants
n=5 Participants
Height
65.6 inches
STANDARD_DEVIATION 4.3 • n=5 Participants

PRIMARY outcome

Timeframe: Priming at each treatment

For every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.

Outcome measures

Outcome measures
Measure
Revaclear MAX
n=219 Dialyzers
Optiflux F180NR
n=95 Dialyzers
Optiflux F200NR
n=97 Dialyzers
Ease of Use: Priming Blood Side
Acceptable
80 Number of Dialyzers
82 Number of Dialyzers
84 Number of Dialyzers
Ease of Use: Priming Blood Side
Very Easy
139 Number of Dialyzers
13 Number of Dialyzers
13 Number of Dialyzers

PRIMARY outcome

Timeframe: Priming at each treatment

For every study treatment/dialyzer, the nursing staff rated the ease of priming dialysate side as follows: Dialysate Side at 500 ml Priming Volume (check one): Perfect: No air visible. Acceptable: Some air visible, but considered insignificant. Not Acceptable: Additional actions needed to sufficiently remove air.

Outcome measures

Outcome measures
Measure
Revaclear MAX
n=220 Dialyzers
Optiflux F180NR
n=95 Dialyzers
Optiflux F200NR
n=97 Dialyzers
Ease of Use: Priming Dialysate Side
Perfect
131 Number of Dialyzers
4 Number of Dialyzers
2 Number of Dialyzers
Ease of Use: Priming Dialysate Side
Acceptable
89 Number of Dialyzers
91 Number of Dialyzers
95 Number of Dialyzers

PRIMARY outcome

Timeframe: Each treatment: assessment of the condition of the dialyzer fibers after rinse-back

For every study treatment/dialyzer, the nursing staff rated the condition of the dialyzer fibers after rinse-back as follows: Very Good:All fibers appear white. Good:Few fibers (less than 10) appear PINK / RED. (check one) Poor:Several fibers (more than 10) appear PINK / RED. (check one) Very Poor:Most fibers (more than 75%) appear PINK / RED. (check one)

Outcome measures

Outcome measures
Measure
Revaclear MAX
n=218 Dialyzers
Optiflux F180NR
n=95 Dialyzers
Optiflux F200NR
n=97 Dialyzers
Dialyzer Assessment: Fiber Condition
Very Good
126 Number of Dialyzers
23 Number of Dialyzers
15 Number of Dialyzers
Dialyzer Assessment: Fiber Condition
Good
75 Number of Dialyzers
44 Number of Dialyzers
45 Number of Dialyzers
Dialyzer Assessment: Fiber Condition
Poor
12 Number of Dialyzers
17 Number of Dialyzers
30 Number of Dialyzers
Dialyzer Assessment: Fiber Condition
Very Poor
5 Number of Dialyzers
11 Number of Dialyzers
7 Number of Dialyzers

PRIMARY outcome

Timeframe: Each treatment: assessment of the condition of the arterial header after rinse-back

For every study treatment/dialyzer, the nursing staff rated the condition of the arterial header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.

Outcome measures

Outcome measures
Measure
Revaclear MAX
n=218 Dialyzers
Optiflux F180NR
n=95 Dialyzers
Optiflux F200NR
n=96 Dialyzers
Dialyzer Assessment: Arterial Header Condition
Very Good
155 Number of Dialyzers
0 Number of Dialyzers
0 Number of Dialyzers
Dialyzer Assessment: Arterial Header Condition
Good
49 Number of Dialyzers
66 Number of Dialyzers
64 Number of Dialyzers
Dialyzer Assessment: Arterial Header Condition
Poor
13 Number of Dialyzers
26 Number of Dialyzers
30 Number of Dialyzers
Dialyzer Assessment: Arterial Header Condition
Very Poor
1 Number of Dialyzers
3 Number of Dialyzers
2 Number of Dialyzers

PRIMARY outcome

Timeframe: Each treatment: the assessment of the condition of the venous header after rinse-back

For every study treatment/dialyzer, the nursing staff rated the condition of the venous header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.

Outcome measures

Outcome measures
Measure
Revaclear MAX
n=217 Dialyzers
Optiflux F180NR
n=94 Dialyzers
Optiflux F200NR
n=97 Dialyzers
Dialyzer Assessment: Venous Header Condition
Very Good
65 Number of Dialyzers
0 Number of Dialyzers
0 Number of Dialyzers
Dialyzer Assessment: Venous Header Condition
Good
140 Number of Dialyzers
31 Number of Dialyzers
35 Number of Dialyzers
Dialyzer Assessment: Venous Header Condition
Poor
7 Number of Dialyzers
52 Number of Dialyzers
57 Number of Dialyzers
Dialyzer Assessment: Venous Header Condition
Very Poor
5 Number of Dialyzers
11 Number of Dialyzers
5 Number of Dialyzers

Adverse Events

Hemodialysis Patients

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported