Trial Outcomes & Findings for Ease of Use and Blood Clotting for the Gambro Polyflux HD-C4 Big Dialyzer (NCT NCT00659724)
NCT ID: NCT00659724
Last Updated: 2025-07-24
Results Overview
For every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.
COMPLETED
33 participants
Priming at each treatment
2025-07-24
Participant Flow
Participant milestones
| Measure |
Hemodialysis Patients
Hemodialysis patients 18 years of age or older receiving routine dialysis treatment for chronic renal failure.
|
|---|---|
|
Weeks 1-2: Assessment of 180NR or 200NR
STARTED
|
33
|
|
Weeks 1-2: Assessment of 180NR or 200NR
COMPLETED
|
33
|
|
Weeks 1-2: Assessment of 180NR or 200NR
NOT COMPLETED
|
0
|
|
Weeks 3-4: Assessment of HD-C4 Big
STARTED
|
33
|
|
Weeks 3-4: Assessment of HD-C4 Big
COMPLETED
|
33
|
|
Weeks 3-4: Assessment of HD-C4 Big
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Ease of Use and Blood Clotting for the Gambro Polyflux HD-C4 Big Dialyzer
Baseline characteristics by cohort
| Measure |
Hemodialysis Patients
n=33 Participants
Hemodialysis patients 18 years of age or older receiving routine dialysis treatment for chronic renal failure.
|
|---|---|
|
Age, Continuous
|
66.0 years
STANDARD_DEVIATION 11.1 • n=5 Participants
|
|
Sex: Female, Male
Female
|
18 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
15 Participants
n=5 Participants
|
|
Height
|
65.6 inches
STANDARD_DEVIATION 4.3 • n=5 Participants
|
PRIMARY outcome
Timeframe: Priming at each treatmentFor every study treatment/dialyzer, the nursing staff rated the ease of priming blood side as follows: Very Easy:Air is easily removed without knocking or clamping procedures. Acceptable:Knocking and/or clamping required for efficient air removal. Difficult:Knocking and/or clamping and additional volume of saline or extended recirculation needed to remove air. Very Difficult:Air could not be removed.
Outcome measures
| Measure |
Revaclear MAX
n=219 Dialyzers
|
Optiflux F180NR
n=95 Dialyzers
|
Optiflux F200NR
n=97 Dialyzers
|
|---|---|---|---|
|
Ease of Use: Priming Blood Side
Acceptable
|
80 Number of Dialyzers
|
82 Number of Dialyzers
|
84 Number of Dialyzers
|
|
Ease of Use: Priming Blood Side
Very Easy
|
139 Number of Dialyzers
|
13 Number of Dialyzers
|
13 Number of Dialyzers
|
PRIMARY outcome
Timeframe: Priming at each treatmentFor every study treatment/dialyzer, the nursing staff rated the ease of priming dialysate side as follows: Dialysate Side at 500 ml Priming Volume (check one): Perfect: No air visible. Acceptable: Some air visible, but considered insignificant. Not Acceptable: Additional actions needed to sufficiently remove air.
Outcome measures
| Measure |
Revaclear MAX
n=220 Dialyzers
|
Optiflux F180NR
n=95 Dialyzers
|
Optiflux F200NR
n=97 Dialyzers
|
|---|---|---|---|
|
Ease of Use: Priming Dialysate Side
Perfect
|
131 Number of Dialyzers
|
4 Number of Dialyzers
|
2 Number of Dialyzers
|
|
Ease of Use: Priming Dialysate Side
Acceptable
|
89 Number of Dialyzers
|
91 Number of Dialyzers
|
95 Number of Dialyzers
|
PRIMARY outcome
Timeframe: Each treatment: assessment of the condition of the dialyzer fibers after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the dialyzer fibers after rinse-back as follows: Very Good:All fibers appear white. Good:Few fibers (less than 10) appear PINK / RED. (check one) Poor:Several fibers (more than 10) appear PINK / RED. (check one) Very Poor:Most fibers (more than 75%) appear PINK / RED. (check one)
Outcome measures
| Measure |
Revaclear MAX
n=218 Dialyzers
|
Optiflux F180NR
n=95 Dialyzers
|
Optiflux F200NR
n=97 Dialyzers
|
|---|---|---|---|
|
Dialyzer Assessment: Fiber Condition
Very Good
|
126 Number of Dialyzers
|
23 Number of Dialyzers
|
15 Number of Dialyzers
|
|
Dialyzer Assessment: Fiber Condition
Good
|
75 Number of Dialyzers
|
44 Number of Dialyzers
|
45 Number of Dialyzers
|
|
Dialyzer Assessment: Fiber Condition
Poor
|
12 Number of Dialyzers
|
17 Number of Dialyzers
|
30 Number of Dialyzers
|
|
Dialyzer Assessment: Fiber Condition
Very Poor
|
5 Number of Dialyzers
|
11 Number of Dialyzers
|
7 Number of Dialyzers
|
PRIMARY outcome
Timeframe: Each treatment: assessment of the condition of the arterial header after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the arterial header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.
Outcome measures
| Measure |
Revaclear MAX
n=218 Dialyzers
|
Optiflux F180NR
n=95 Dialyzers
|
Optiflux F200NR
n=96 Dialyzers
|
|---|---|---|---|
|
Dialyzer Assessment: Arterial Header Condition
Very Good
|
155 Number of Dialyzers
|
0 Number of Dialyzers
|
0 Number of Dialyzers
|
|
Dialyzer Assessment: Arterial Header Condition
Good
|
49 Number of Dialyzers
|
66 Number of Dialyzers
|
64 Number of Dialyzers
|
|
Dialyzer Assessment: Arterial Header Condition
Poor
|
13 Number of Dialyzers
|
26 Number of Dialyzers
|
30 Number of Dialyzers
|
|
Dialyzer Assessment: Arterial Header Condition
Very Poor
|
1 Number of Dialyzers
|
3 Number of Dialyzers
|
2 Number of Dialyzers
|
PRIMARY outcome
Timeframe: Each treatment: the assessment of the condition of the venous header after rinse-backFor every study treatment/dialyzer, the nursing staff rated the condition of the venous header after rinse-back as follows: Very Good:Surface as clean as expected. Good:Distinct red ring appears at the dialyzer wall. Poor:Significant appearance of clots, randomly distributed on the surface. Very Poor:Completely red and/or covered with clots.
Outcome measures
| Measure |
Revaclear MAX
n=217 Dialyzers
|
Optiflux F180NR
n=94 Dialyzers
|
Optiflux F200NR
n=97 Dialyzers
|
|---|---|---|---|
|
Dialyzer Assessment: Venous Header Condition
Very Good
|
65 Number of Dialyzers
|
0 Number of Dialyzers
|
0 Number of Dialyzers
|
|
Dialyzer Assessment: Venous Header Condition
Good
|
140 Number of Dialyzers
|
31 Number of Dialyzers
|
35 Number of Dialyzers
|
|
Dialyzer Assessment: Venous Header Condition
Poor
|
7 Number of Dialyzers
|
52 Number of Dialyzers
|
57 Number of Dialyzers
|
|
Dialyzer Assessment: Venous Header Condition
Very Poor
|
5 Number of Dialyzers
|
11 Number of Dialyzers
|
5 Number of Dialyzers
|
Adverse Events
Hemodialysis Patients
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place