Trial Outcomes & Findings for Product Performance of a Toric Contact Lens (NCT NCT00658996)

NCT ID: NCT00658996

Last Updated: 2011-12-12

Results Overview

1-100 Scale for each eye. Zero represented the least favorable rating and 100 represented the most favorable.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

220 participants

Primary outcome timeframe

Over-all follow-up visits for 2 week period

Results posted on

2011-12-12

Participant Flow

This 4-week study was conducted at 12 sites in Asia. First participant enrolled 4/16/08, last participant visit was 6/20/08

220 participants were enrolled in the study, 4 participants were ineligible at screening visit, 216 Subjects started the first intervention. 211 participants started the second intervention. 209 participants completed the study.

Participant milestones

Participant milestones
Measure
SofLens DD Toric First, Then Focus Dailies Toric
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens first then crossover to Ciba Vision Focus Dailies Toric Lens
Focus Dailies Toric First, Then SofLens DD Toric
Ciba Vision Focus Dailies Toric Lens first, then crossover to Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
First Intervention
STARTED
109
107
First Intervention
COMPLETED
108
106
First Intervention
NOT COMPLETED
1
1
Crossover Period
STARTED
108
106
Crossover Period
COMPLETED
106
105
Crossover Period
NOT COMPLETED
2
1
Second Intervention
STARTED
106
105
Second Intervention
COMPLETED
105
104
Second Intervention
NOT COMPLETED
1
1

Reasons for withdrawal

Reasons for withdrawal
Measure
SofLens DD Toric First, Then Focus Dailies Toric
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens first then crossover to Ciba Vision Focus Dailies Toric Lens
Focus Dailies Toric First, Then SofLens DD Toric
Ciba Vision Focus Dailies Toric Lens first, then crossover to Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
First Intervention
Positive slit lamp finding
1
0
First Intervention
Lack of lenses
0
1
Crossover Period
Physician Decision
1
0
Crossover Period
Ineligible
1
1
Second Intervention
Developed a stye on upper lid
1
0
Second Intervention
Positive slit lamp finding
0
1

Baseline Characteristics

Product Performance of a Toric Contact Lens

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Entire Study Population
n=216 Participants
Population enrolled at start of study
Age Continuous
28.7 Years
STANDARD_DEVIATION 7.19 • n=5 Participants
Sex: Female, Male
Female
150 Participants
n=5 Participants
Sex: Female, Male
Male
66 Participants
n=5 Participants
Race/Ethnicity, Customized
Asian
216 participants
n=5 Participants
Race/Ethnicity, Customized
White
0 participants
n=5 Participants
Race/Ethnicity, Customized
Black
0 participants
n=5 Participants
Race/Ethnicity, Customized
Other
0 participants
n=5 Participants

PRIMARY outcome

Timeframe: Over-all follow-up visits for 2 week period

Population: All eligible, dispensed eyes, Overall follow-up visits.

1-100 Scale for each eye. Zero represented the least favorable rating and 100 represented the most favorable.

Outcome measures

Outcome measures
Measure
SofLens DD Toric
n=213 Participants
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
Focus Dailies Toric
n=210 Participants
Ciba Vision Focus Dailies Toric contact lenses
Symptoms and Complaints
Lens Cleanliness
87.5 Scores on a Scale
Standard Deviation 12.7
86.5 Scores on a Scale
Standard Deviation 13.0
Symptoms and Complaints
Comfort
79.3 Scores on a Scale
Standard Deviation 17.1
77.7 Scores on a Scale
Standard Deviation 17.6
Symptoms and Complaints
End of Day Comfort
73.5 Scores on a Scale
Standard Deviation 19.6
71.5 Scores on a Scale
Standard Deviation 19.9
Symptoms and Complaints
Lens Awareness
78.8 Scores on a Scale
Standard Deviation 17.9
75.6 Scores on a Scale
Standard Deviation 20.4
Symptoms and Complaints
Irritation
83.0 Scores on a Scale
Standard Deviation 15.6
79.3 Scores on a Scale
Standard Deviation 19.6
Symptoms and Complaints
Itching
86.4 Scores on a Scale
Standard Deviation 14.9
83.9 Scores on a Scale
Standard Deviation 17.2
Symptoms and Complaints
Dryness
77.4 Scores on a Scale
Standard Deviation 18.0
73.7 Scores on a Scale
Standard Deviation 20.3
Symptoms and Complaints
Redness
86.3 Scores on a Scale
Standard Deviation 14.9
81.5 Scores on a Scale
Standard Deviation 19.5
Symptoms and Complaints
Vision
77.1 Scores on a Scale
Standard Deviation 19.1
84.3 Scores on a Scale
Standard Deviation 14.6
Symptoms and Complaints
Lens Handling
87.4 Scores on a Scale
Standard Deviation 13.0
76.1 Scores on a Scale
Standard Deviation 21.2
Symptoms and Complaints
Overall Impression
78.7 Scores on a Scale
Standard Deviation 16.2
75.9 Scores on a Scale
Standard Deviation 17.5

PRIMARY outcome

Timeframe: Over-all follow-up visits, 2 weeks

Population: All eligible, dispensed eyes

VA measures at each scheduled visit were averaged to obtain an overall measure for each eye. A non-inferiority upper bound of 0.06 (3 letters) were used to assess the difference (Test - Control) in overall logMAR VA.

Outcome measures

Outcome measures
Measure
SofLens DD Toric
n=426 eyes
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
Focus Dailies Toric
n=420 eyes
Ciba Vision Focus Dailies Toric contact lenses
Contact Lens High Contrast Visual Acuity
-0.007 LogMAR
Standard Deviation 0.0669
-0.014 LogMAR
Standard Deviation 0.0696

SECONDARY outcome

Timeframe: Over-all follow-up visits, 2 weeks

Population: All dispensed eyes, over all follow-up visits

Graded 0-4 where 0=none and 4=severe on measure of epithelial edema, epithelial microcysts, corneal staining, limbal injection, bulbar injection, superior tarsal conjunctival abnormalities, corneal neovascularization, and corneal infiltrates

Outcome measures

Outcome measures
Measure
SofLens DD Toric
n=430 eyes
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
Focus Dailies Toric
n=430 eyes
Ciba Vision Focus Dailies Toric contact lenses
Slit Lamp Findings
Slit Lamp Findings > Grade 2
0 Eyes
2 Eyes
Slit Lamp Findings
Slit Lamp Findings </= Grade 2
430 Eyes
428 Eyes

Adverse Events

SofLens DD Toric

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Ciba Vision Toric Lens

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Gerard Cairns, PhD, MCOptom

Bausch & Lomb Incorporated

Phone: (585) 338-5170

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: OTHER