Trial Outcomes & Findings for Product Performance of a Toric Contact Lens (NCT NCT00658996)
NCT ID: NCT00658996
Last Updated: 2011-12-12
Results Overview
1-100 Scale for each eye. Zero represented the least favorable rating and 100 represented the most favorable.
COMPLETED
NA
220 participants
Over-all follow-up visits for 2 week period
2011-12-12
Participant Flow
This 4-week study was conducted at 12 sites in Asia. First participant enrolled 4/16/08, last participant visit was 6/20/08
220 participants were enrolled in the study, 4 participants were ineligible at screening visit, 216 Subjects started the first intervention. 211 participants started the second intervention. 209 participants completed the study.
Participant milestones
| Measure |
SofLens DD Toric First, Then Focus Dailies Toric
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens first then crossover to Ciba Vision Focus Dailies Toric Lens
|
Focus Dailies Toric First, Then SofLens DD Toric
Ciba Vision Focus Dailies Toric Lens first, then crossover to Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
|
|---|---|---|
|
First Intervention
STARTED
|
109
|
107
|
|
First Intervention
COMPLETED
|
108
|
106
|
|
First Intervention
NOT COMPLETED
|
1
|
1
|
|
Crossover Period
STARTED
|
108
|
106
|
|
Crossover Period
COMPLETED
|
106
|
105
|
|
Crossover Period
NOT COMPLETED
|
2
|
1
|
|
Second Intervention
STARTED
|
106
|
105
|
|
Second Intervention
COMPLETED
|
105
|
104
|
|
Second Intervention
NOT COMPLETED
|
1
|
1
|
Reasons for withdrawal
| Measure |
SofLens DD Toric First, Then Focus Dailies Toric
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens first then crossover to Ciba Vision Focus Dailies Toric Lens
|
Focus Dailies Toric First, Then SofLens DD Toric
Ciba Vision Focus Dailies Toric Lens first, then crossover to Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
|
|---|---|---|
|
First Intervention
Positive slit lamp finding
|
1
|
0
|
|
First Intervention
Lack of lenses
|
0
|
1
|
|
Crossover Period
Physician Decision
|
1
|
0
|
|
Crossover Period
Ineligible
|
1
|
1
|
|
Second Intervention
Developed a stye on upper lid
|
1
|
0
|
|
Second Intervention
Positive slit lamp finding
|
0
|
1
|
Baseline Characteristics
Product Performance of a Toric Contact Lens
Baseline characteristics by cohort
| Measure |
Entire Study Population
n=216 Participants
Population enrolled at start of study
|
|---|---|
|
Age Continuous
|
28.7 Years
STANDARD_DEVIATION 7.19 • n=5 Participants
|
|
Sex: Female, Male
Female
|
150 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
66 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Asian
|
216 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
White
|
0 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Black
|
0 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Other
|
0 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: Over-all follow-up visits for 2 week periodPopulation: All eligible, dispensed eyes, Overall follow-up visits.
1-100 Scale for each eye. Zero represented the least favorable rating and 100 represented the most favorable.
Outcome measures
| Measure |
SofLens DD Toric
n=213 Participants
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
|
Focus Dailies Toric
n=210 Participants
Ciba Vision Focus Dailies Toric contact lenses
|
|---|---|---|
|
Symptoms and Complaints
Lens Cleanliness
|
87.5 Scores on a Scale
Standard Deviation 12.7
|
86.5 Scores on a Scale
Standard Deviation 13.0
|
|
Symptoms and Complaints
Comfort
|
79.3 Scores on a Scale
Standard Deviation 17.1
|
77.7 Scores on a Scale
Standard Deviation 17.6
|
|
Symptoms and Complaints
End of Day Comfort
|
73.5 Scores on a Scale
Standard Deviation 19.6
|
71.5 Scores on a Scale
Standard Deviation 19.9
|
|
Symptoms and Complaints
Lens Awareness
|
78.8 Scores on a Scale
Standard Deviation 17.9
|
75.6 Scores on a Scale
Standard Deviation 20.4
|
|
Symptoms and Complaints
Irritation
|
83.0 Scores on a Scale
Standard Deviation 15.6
|
79.3 Scores on a Scale
Standard Deviation 19.6
|
|
Symptoms and Complaints
Itching
|
86.4 Scores on a Scale
Standard Deviation 14.9
|
83.9 Scores on a Scale
Standard Deviation 17.2
|
|
Symptoms and Complaints
Dryness
|
77.4 Scores on a Scale
Standard Deviation 18.0
|
73.7 Scores on a Scale
Standard Deviation 20.3
|
|
Symptoms and Complaints
Redness
|
86.3 Scores on a Scale
Standard Deviation 14.9
|
81.5 Scores on a Scale
Standard Deviation 19.5
|
|
Symptoms and Complaints
Vision
|
77.1 Scores on a Scale
Standard Deviation 19.1
|
84.3 Scores on a Scale
Standard Deviation 14.6
|
|
Symptoms and Complaints
Lens Handling
|
87.4 Scores on a Scale
Standard Deviation 13.0
|
76.1 Scores on a Scale
Standard Deviation 21.2
|
|
Symptoms and Complaints
Overall Impression
|
78.7 Scores on a Scale
Standard Deviation 16.2
|
75.9 Scores on a Scale
Standard Deviation 17.5
|
PRIMARY outcome
Timeframe: Over-all follow-up visits, 2 weeksPopulation: All eligible, dispensed eyes
VA measures at each scheduled visit were averaged to obtain an overall measure for each eye. A non-inferiority upper bound of 0.06 (3 letters) were used to assess the difference (Test - Control) in overall logMAR VA.
Outcome measures
| Measure |
SofLens DD Toric
n=426 eyes
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
|
Focus Dailies Toric
n=420 eyes
Ciba Vision Focus Dailies Toric contact lenses
|
|---|---|---|
|
Contact Lens High Contrast Visual Acuity
|
-0.007 LogMAR
Standard Deviation 0.0669
|
-0.014 LogMAR
Standard Deviation 0.0696
|
SECONDARY outcome
Timeframe: Over-all follow-up visits, 2 weeksPopulation: All dispensed eyes, over all follow-up visits
Graded 0-4 where 0=none and 4=severe on measure of epithelial edema, epithelial microcysts, corneal staining, limbal injection, bulbar injection, superior tarsal conjunctival abnormalities, corneal neovascularization, and corneal infiltrates
Outcome measures
| Measure |
SofLens DD Toric
n=430 eyes
Bausch \& Lomb SofLens Daily Disposable Toric Contact Lens
|
Focus Dailies Toric
n=430 eyes
Ciba Vision Focus Dailies Toric contact lenses
|
|---|---|---|
|
Slit Lamp Findings
Slit Lamp Findings > Grade 2
|
0 Eyes
|
2 Eyes
|
|
Slit Lamp Findings
Slit Lamp Findings </= Grade 2
|
430 Eyes
|
428 Eyes
|
Adverse Events
SofLens DD Toric
Ciba Vision Toric Lens
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Gerard Cairns, PhD, MCOptom
Bausch & Lomb Incorporated
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: OTHER