Trial Outcomes & Findings for Comparative Effects Of Dialysate Flow Rate And Membrane Packing On The Performance Of Dialyzers Used For Hemodialysis (NCT NCT00636077)

NCT ID: NCT00636077

Last Updated: 2025-07-14

Results Overview

KoA is a constant that describes the efficiency of a dialyzer in removing urea. KoA is determined by surface area of the membrane, the thickness of the membrane and pore size.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

12 participants

Primary outcome timeframe

During the third treatment with each dialyzer (one time during each trial period week)

Results posted on

2025-07-14

Participant Flow

Participant milestones

Participant milestones
Measure
HD-C4 Big First
Patients in this Arm will receive 3 consecutive treatments with the HD-C4 Big dialyzer first. For the study each patient will be treated for three consecutive treatments with each of the 4 study dialyzers (total of 12 consecutive treatments).
HD-C4 Small First
Patients in this Arm will receive 3 consecutive treatments with the HD-C4 Small dialyzer first. For the study each patient will be treated for three consecutive treatments with each of the 4 study dialyzers (total of 12 consecutive treatments).
Optiflux F160NR First
Patients in this Arm will receive 3 consecutive treatments with the F160NR dialyzer first. For the study each patient will be treated for three consecutive treatments with each of the 4 study dialyzers (total of 12 consecutive treatments).
Optiflux F200NR First
Patients in this Arm will receive 3 consecutive treatments with the F200NR dialyzer first. For the study each patient will be treated for three consecutive treatments with each of the 4 study dialyzers (total of 12 consecutive treatments).
Week 1
STARTED
3
3
3
3
Week 1
COMPLETED
3
3
3
3
Week 1
NOT COMPLETED
0
0
0
0
Week 2
STARTED
3
3
3
3
Week 2
COMPLETED
3
3
3
3
Week 2
NOT COMPLETED
0
0
0
0
Week 3
STARTED
3
3
3
3
Week 3
COMPLETED
3
3
3
3
Week 3
NOT COMPLETED
0
0
0
0
Week 4
STARTED
3
3
3
3
Week 4
COMPLETED
3
3
3
3
Week 4
NOT COMPLETED
0
0
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Comparative Effects Of Dialysate Flow Rate And Membrane Packing On The Performance Of Dialyzers Used For Hemodialysis

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Study Participants
n=12 Participants
Age, Continuous
60 Years
n=5 Participants
Sex: Female, Male
Female
9 Participants
n=5 Participants
Sex: Female, Male
Male
3 Participants
n=5 Participants

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

KoA is a constant that describes the efficiency of a dialyzer in removing urea. KoA is determined by surface area of the membrane, the thickness of the membrane and pore size.

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
844 KoA for urea (mL/min)
Standard Error 30
763 KoA for urea (mL/min)
Standard Error 34
770 KoA for urea (mL/min)
Standard Error 36
798 KoA for urea (mL/min)
Standard Error 21
Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
898 KoA for urea (mL/min)
Standard Error 31
748 KoA for urea (mL/min)
Standard Error 34
738 KoA for urea (mL/min)
Standard Error 40
890 KoA for urea (mL/min)
Standard Error 33
Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
922 KoA for urea (mL/min)
Standard Error 31
761 KoA for urea (mL/min)
Standard Error 25
809 KoA for urea (mL/min)
Standard Error 24
817 KoA for urea (mL/min)
Standard Error 36

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
279 Whole blood urea clearance (mL/min)
Standard Error 3
270 Whole blood urea clearance (mL/min)
Standard Error 4
271 Whole blood urea clearance (mL/min)
Standard Error 4
274 Whole blood urea clearance (mL/min)
Standard Error 2
Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
315 Whole blood urea clearance (mL/min)
Standard Error 3
295 Whole blood urea clearance (mL/min)
Standard Error 5
293 Whole blood urea clearance (mL/min)
Standard Error 8
313 Whole blood urea clearance (mL/min)
Standard Error 4
Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
340 Whole blood urea clearance (mL/min)
Standard Error 3
320 Whole blood urea clearance (mL/min)
Standard Error 3
313 Whole blood urea clearance (mL/min)
Standard Error 4
327 Whole blood urea clearance (mL/min)
Standard Error 5

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
434 KoA for Phosphorus (mL/min)
Standard Error 48
334 KoA for Phosphorus (mL/min)
Standard Error 24
434 KoA for Phosphorus (mL/min)
Standard Error 34
504 KoA for Phosphorus (mL/min)
Standard Error 58
Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
435 KoA for Phosphorus (mL/min)
Standard Error 31
422 KoA for Phosphorus (mL/min)
Standard Error 48
450 KoA for Phosphorus (mL/min)
Standard Error 62
491 KoA for Phosphorus (mL/min)
Standard Error 29
Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
520 KoA for Phosphorus (mL/min)
Standard Error 39
415 KoA for Phosphorus (mL/min)
Standard Error 35
458 KoA for Phosphorus (mL/min)
Standard Error 33
560 KoA for Phosphorus (mL/min)
Standard Error 40

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
286 whole blood phosphorus clearance mL/min
Standard Error 7
273 whole blood phosphorus clearance mL/min
Standard Error 7
272 whole blood phosphorus clearance mL/min
Standard Error 7
281 whole blood phosphorus clearance mL/min
Standard Error 7
Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
255 whole blood phosphorus clearance mL/min
Standard Error 7
252 whole blood phosphorus clearance mL/min
Standard Error 9
247 whole blood phosphorus clearance mL/min
Standard Error 6
249 whole blood phosphorus clearance mL/min
Standard Error 5
Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
269 whole blood phosphorus clearance mL/min
Standard Error 10
261 whole blood phosphorus clearance mL/min
Standard Error 9
260 whole blood phosphorus clearance mL/min
Standard Error 7
265 whole blood phosphorus clearance mL/min
Standard Error 7

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
65 KoA for b2-microglobulin (mL/min)
Standard Error 4
61 KoA for b2-microglobulin (mL/min)
Standard Error 2
27 KoA for b2-microglobulin (mL/min)
Standard Error 2
27 KoA for b2-microglobulin (mL/min)
Standard Error 2
Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
62 KoA for b2-microglobulin (mL/min)
Standard Error 3
57 KoA for b2-microglobulin (mL/min)
Standard Error 2
28 KoA for b2-microglobulin (mL/min)
Standard Error 2
25 KoA for b2-microglobulin (mL/min)
Standard Error 2
Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
64 KoA for b2-microglobulin (mL/min)
Standard Error 3
56 KoA for b2-microglobulin (mL/min)
Standard Error 3
31 KoA for b2-microglobulin (mL/min)
Standard Error 5
21 KoA for b2-microglobulin (mL/min)
Standard Error 1

PRIMARY outcome

Timeframe: During the third treatment with each dialyzer (one time during each trial period week)

Outcome measures

Outcome measures
Measure
HD-C4 Big
n=12 Participants
3 consecutive treatments with the HD-C4 Big dialyzer.
HD-C4 Small
n=12 Participants
3 consecutive treatments with the HD-C4 Small dialyzer.
F160NR
n=12 Participants
3 consecutive treatments with the F160NR dialyzer.
F200NR
n=12 Participants
3 consecutive treatments with the F200NR dialyzer.
Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
350 mL/min Dialysate Flow Rate
106 whole blood clearance B2-microglobulin
Standard Error 9
119 whole blood clearance B2-microglobulin
Standard Error 6
87 whole blood clearance B2-microglobulin
Standard Error 6
72 whole blood clearance B2-microglobulin
Standard Error 9
Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
500 mL/min Dialysate Flow Rate
129 whole blood clearance B2-microglobulin
Standard Error 7
118 whole blood clearance B2-microglobulin
Standard Error 8
86 whole blood clearance B2-microglobulin
Standard Error 10
65 whole blood clearance B2-microglobulin
Standard Error 5
Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.
800 mL/min Dialysate Flow Rate
118 whole blood clearance B2-microglobulin
Standard Error 7
121 whole blood clearance B2-microglobulin
Standard Error 9
78 whole blood clearance B2-microglobulin
Standard Error 8
75 whole blood clearance B2-microglobulin
Standard Error 7

Adverse Events

HD-C4 Big

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

HD-C4 Small

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

F160NR

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

F200NR

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Global CORP Clinical Trials Disclosure

Vantive

Phone: +1 2249484283

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place