Clinical Risk Assesment of CS-1 Decision Support System
NCT ID: NCT00502645
Last Updated: 2007-07-17
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
10 participants
INTERVENTIONAL
2007-04-30
2007-06-30
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
The target range for blood glucose levels will be 80-110 mg/dl \[4.4 - 6.1 mmol/l\]. The current blood glucose level and insulin infusion rate will continuously be documented by the CS-1 decision support system and supervised by the investigator. Glucose values will be provided to the system by manual entry. In case of hypoglycaemia, intravenous insulin infusion will be stopped and glucose will be administered via intravenous bolus.
The treatment visit, including all trial related activities as described above, will last for 72 hours. Adverse events and relevant medication will be continuously monitored and documented The Acute Physiology and Chronic Health Evaluation Score (Apache II) will be documented for each patient at the beginning of the trial. There is no physical connection between the algorithm suggesting the insulin infusion rate and the insulin infusion pump such as that the algorithm would be capable to independently steer the insulin infusion. There is also no physical connection between the glucose measurement device and the algorithm such as that the algorihthm would be able to independently get a glucose measurement. For safety reasons, both the glucose measurement and the insulin infusion rate are parameters that need to be manually entered by a specially trained nurse. For glucose measurement, a certified device for ICU application is being used (Accuchek® Inform; Roche Diagnostics), for intravenous insulin infusion, a certified insulin infusion pump (Perfusor® Space, BBraun Melsungen) as it is used under routine conditions in the ICU is being used. Follow up information will be acquired at hospital discharge or at the latest one week after visit 1. .
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NON_RANDOMIZED
SINGLE_GROUP
TREATMENT
NONE
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
CS-1 DECISION SUPPORT SYSTEM
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Stay in the ICU expected to be \> 72 h
* Blood glucose \> 110 mg/dl or patient on insulin treatment
Exclusion Criteria
* Known or suspected allergy to insulin
* Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the patient (i.e., liver failure, other fatal organ failures)
* Patients participating in another study
* Moribund patients likely to die within 24 hours
* Patients after organ transplantation within the last three months
* Patients under high dose cortisol treatment (cortisol \> 1000 mg/day or equivalent dosis of hydrocortisol)
19 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Medical University of Graz
OTHER
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Thomas R. Pieber, MD
Role: PRINCIPAL_INVESTIGATOR
Medical University of Graz
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Medical University Graz
Graz, , Austria
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Plank J, Blaha J, Cordingley J, Wilinska ME, Chassin LJ, Morgan C, Squire S, Haluzik M, Kremen J, Svacina S, Toller W, Plasnik A, Ellmerer M, Hovorka R, Pieber TR. Multicentric, randomized, controlled trial to evaluate blood glucose control by the model predictive control algorithm versus routine glucose management protocols in intensive care unit patients. Diabetes Care. 2006 Feb;29(2):271-6. doi: 10.2337/diacare.29.02.06.dc05-1689.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
CM10_CS-1
Identifier Type: -
Identifier Source: org_study_id