Trial Outcomes & Findings for Study of Robotic Template Guidance for Needle Placement in Transperineal Prostate Brachytherapy (NCT NCT00381966)

NCT ID: NCT00381966

Last Updated: 2019-11-26

Results Overview

Confirmatory measurements of robotic movements was assessed using infrared tracking to calculate mean error in millimeters.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

5 participants

Primary outcome timeframe

at time of intervention

Results posted on

2019-11-26

Participant Flow

Participant milestones

Participant milestones
Measure
Robotic Placement Device
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy. robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
Overall Study
STARTED
5
Overall Study
COMPLETED
5
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Robotic Placement Device
n=5 Participants
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy. robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
3 Participants
n=5 Participants
Age, Categorical
>=65 years
2 Participants
n=5 Participants
Sex: Female, Male
Female
0 Participants
n=5 Participants
Sex: Female, Male
Male
5 Participants
n=5 Participants

PRIMARY outcome

Timeframe: at time of intervention

Confirmatory measurements of robotic movements was assessed using infrared tracking to calculate mean error in millimeters.

Outcome measures

Outcome measures
Measure
Robotic Placement Device
n=5 Participants
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy. robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
Positioning Effectiveness as Determined by Confirmatory Measurements of Robotic Movements
0.489 millimeters
Standard Deviation 0.328

PRIMARY outcome

Timeframe: at time of intervention

Utility was demonstrated in terms of fine adjustments (≤ 2 mm) and adjustments \>2mm in needle positioning, being possible when tissue deflection was encountered.

Outcome measures

Outcome measures
Measure
Robotic Placement Device
n=54 Number of Adjustments
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy. robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
Utility of the Needle Position Device: Number of Adjustments Possible Out of Adjustments Attempted
Adjustments ≤ 2 mm
18 Number of Adjustments
Utility of the Needle Position Device: Number of Adjustments Possible Out of Adjustments Attempted
Adjustments > 2 mm
36 Number of Adjustments

Adverse Events

Robotic Placement Device

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Daniel Song

Johns Hopkins University

Phone: 401-502-5875

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place