Trial Outcomes & Findings for Study of Robotic Template Guidance for Needle Placement in Transperineal Prostate Brachytherapy (NCT NCT00381966)
NCT ID: NCT00381966
Last Updated: 2019-11-26
Results Overview
Confirmatory measurements of robotic movements was assessed using infrared tracking to calculate mean error in millimeters.
Recruitment status
COMPLETED
Study phase
NA
Target enrollment
5 participants
Primary outcome timeframe
at time of intervention
Results posted on
2019-11-26
Participant Flow
Participant milestones
| Measure |
Robotic Placement Device
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy.
robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
|
|---|---|
|
Overall Study
STARTED
|
5
|
|
Overall Study
COMPLETED
|
5
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Robotic Placement Device
n=5 Participants
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy.
robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
3 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
2 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
5 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: at time of interventionConfirmatory measurements of robotic movements was assessed using infrared tracking to calculate mean error in millimeters.
Outcome measures
| Measure |
Robotic Placement Device
n=5 Participants
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy.
robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
|
|---|---|
|
Positioning Effectiveness as Determined by Confirmatory Measurements of Robotic Movements
|
0.489 millimeters
Standard Deviation 0.328
|
PRIMARY outcome
Timeframe: at time of interventionUtility was demonstrated in terms of fine adjustments (≤ 2 mm) and adjustments \>2mm in needle positioning, being possible when tissue deflection was encountered.
Outcome measures
| Measure |
Robotic Placement Device
n=54 Number of Adjustments
The intervention involves use of a robotic template to assist in placement of needles for prostate brachytherapy.
robotic placement device: this device is a robotic arm that utilizes a template as a guidance system to precisely inplant radioactive brachytherapy seeds directly to the prostate.
|
|---|---|
|
Utility of the Needle Position Device: Number of Adjustments Possible Out of Adjustments Attempted
Adjustments ≤ 2 mm
|
18 Number of Adjustments
|
|
Utility of the Needle Position Device: Number of Adjustments Possible Out of Adjustments Attempted
Adjustments > 2 mm
|
36 Number of Adjustments
|
Adverse Events
Robotic Placement Device
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place